Reliability and Validity of the Belt-Stabilized Handheld Dynamometer in Hip- and Knee-Strength Tests.

Jaqueline Martins1, Janaina Rodrigues da Silva1, Marcelo Rodrigues Barbosa da Silva1, Débora Bevilaqua-Grossi1.   

Abstract

CONTEXT: The belt-stabilized handheld dynamometer (HHD) has been used to assess the strength of knee- and hip-muscle groups. However, few researchers have examined its reliability and validity for assessing the strength of these muscles.
OBJECTIVE: To evaluate the intra-examiner reliability of the belt-stabilized HHD and its validity and agreement with the isokinetic dynamometer for assessing the strength of knee- and hip-muscle groups.
DESIGN: Cross-sectional study.
SETTING: University laboratory. PATIENTS OR OTHER PARTICIPANTS: We evaluated 26 healthy participants (13 men, 13 women; age = 23.5 ± 2.8 years, height = 1.7 ± 0.1 m, mass = 68.6 ± 12.4 kg) in 2 sessions using the belt-stabilized HHD and an isokinetic dynamometer for maximum strength of the hip adductors, abductors, flexors, extensors, internal rotators, and external rotators and the knee flexors and extensors. MAIN OUTCOME MEASURE(S): We used reliability values provided by the intraclass correlation coefficient (2,3), standard error of measurement (SEM and percentage SEM), and minimal detectable change; correlation values comparing the belt-stabilized HHD and the isokinetic instrument using the Pearson correlation coefficient (r); and the mean difference in values comparing the 2 instruments using the Bland-Altman method.
RESULTS: The intrarater HHD reliability was excellent for most measurements (range = 0.80-0.96; SEM = 1.3-5.3 kilograms of force or 4.8-18.9 Nm, percentage SEM = 7.0%-22.0%, minimal detectable change = 3.6-18.8 kilograms of force or 13.2-52.4 Nm) and was moderate only for bilateral knee flexion and left hip internal rotation (intraclass correlation coefficient [2,3] = 0.62-0.66 and 0.70, respectively). Correlation with the isokinetic dynamometer was moderate to high (r = 0.60-0.90), but the absolute values did not demonstrate concordance between results using the Bland-Altman method.
CONCLUSIONS: The belt-stabilized HHD measurements were reliable, and although they did not agree with those from the isokinetic dynamometer, the values were correlated for the hip- and knee-muscle groups.

Entities:  

Keywords:  lower extremity; muscle strength; reproducibility of results

Mesh:

Year:  2017        PMID: 28787180      PMCID: PMC5634229          DOI: 10.4085/1062-6050-52.6.04

Source DB:  PubMed          Journal:  J Athl Train        ISSN: 1062-6050            Impact factor:   2.860


  31 in total

1.  Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements.

Authors:  Joshua M Drouin; Tamara C Valovich-mcLeod; Sandra J Shultz; Bruce M Gansneder; David H Perrin
Journal:  Eur J Appl Physiol       Date:  2003-09-24       Impact factor: 3.078

Review 2.  Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review.

Authors:  Timothy Stark; Bruce Walker; Jacqueline K Phillips; René Fejer; Randy Beck
Journal:  PM R       Date:  2011-05       Impact factor: 2.298

3.  Reliability and validity of isometric knee extensor strength test with hand-held dynamometer depending on its fixation: a pilot study.

Authors:  Won Kuel Kim; Don-Kyu Kim; Kyung Mook Seo; Si Hyun Kang
Journal:  Ann Rehabil Med       Date:  2014-02-25

4.  Is hand-held dynamometry useful for the measurement of quadriceps strength in older people? A comparison with the gold standard Bodex dynamometry.

Authors:  H J Martin; V Yule; H E Syddall; E M Dennison; C Cooper; A Aihie Sayer
Journal:  Gerontology       Date:  2006       Impact factor: 5.140

5.  Hip abductor strength in people with knee osteoarthritis: A cross-sectional study of reliability and association with function.

Authors:  Michael A Tevald; Amanda Murray; Brittney A Luc; Kafai Lai; David Sohn; Brian Pietrosimone
Journal:  Knee       Date:  2015-06-30       Impact factor: 2.199

6.  The development, validity, and reliability of a manual muscle testing device with integrated limb position sensors.

Authors:  Raymond C Li; Jan M Jasiewicz; James Middleton; Peter Condie; Andrew Barriskill; Heidi Hebnes; Brendan Purcell
Journal:  Arch Phys Med Rehabil       Date:  2006-03       Impact factor: 3.966

7.  The reliability and validity of handheld dynamometry for the measurement of lower-extremity muscle strength in older adults.

Authors:  Cathy M Arnold; Kathryn D Warkentin; Philip D Chilibeck; Charlene R A Magnus
Journal:  J Strength Cond Res       Date:  2010-03       Impact factor: 3.775

8.  Preoperative quadriceps strength is a significant predictor of knee function two years after anterior cruciate ligament reconstruction.

Authors:  I Eitzen; I Holm; M A Risberg
Journal:  Br J Sports Med       Date:  2009-02-17       Impact factor: 13.800

9.  Prospective Predictors of Patellofemoral Pain Syndrome: A Systematic Review With Meta-analysis.

Authors:  Evangelos Pappas; Wing M Wong-Tom
Journal:  Sports Health       Date:  2012-03       Impact factor: 3.843

10.  The intra- and inter-rater reliabilities of lower extremity muscle strength assessment of healthy adults using a hand held dynamometer.

Authors:  Seong-Gil Kim; Yun-Seob Lee
Journal:  J Phys Ther Sci       Date:  2015-06-30
View more
  16 in total

1.  Reliability of hand-held dynamometer in measuring gluteus medius isometric muscle strength in healthy population.

Authors:  Rabail Rani Soomro; Hossein Karimi; Syed Amir Gillani
Journal:  Pak J Med Sci       Date:  2022 May-Jun       Impact factor: 2.340

2.  Influence of a periodized circuit training protocol on intermuscular adipose tissue of patients with knee osteoarthritis: protocol for a randomized controlled trial.

Authors:  Aline Castilho de Almeida; Maria Gabriela Pedroso; Jessica Bianca Aily; Glaucia Helena Gonçalves; Carlos Marcelo Pastre; Stela Marcia Mattiello
Journal:  BMC Musculoskelet Disord       Date:  2018-11-30       Impact factor: 2.362

3.  The effects of gluteal squeezes compared to bilateral bridges on gluteal strength, power, endurance, and girth.

Authors:  Bryan J Lehecka; Jessica Turley; Aaron Stapleton; Kyle Waits; John Zirkle
Journal:  PeerJ       Date:  2019-07-08       Impact factor: 2.984

4.  Tensiomyographical responsiveness to peripheral fatigue in quadriceps femoris.

Authors:  Rodrigo Martín-San Agustín; Francesc Medina-Mirapeix; José Casaña-Granell; José A García-Vidal; Carmen Lillo-Navarro; Josep C Benítez-Martínez
Journal:  PeerJ       Date:  2020-02-28       Impact factor: 2.984

5.  Comparison of Common Methodologies for the Determination of Knee Flexor Muscle Strength.

Authors:  Dan I Ogborn; Alix Bellemare; Brittany Bruinooge; Holly Brown; Sheila McRae; Jeff Leiter
Journal:  Int J Sports Phys Ther       Date:  2021-04-01

6.  Reliability of hip muscle strength measured in principal and intermediate planes of movement.

Authors:  Basilio A M Goncalves; David J Saxby; Adam Kositsky; Rod S Barrett; Laura E Diamond
Journal:  PeerJ       Date:  2021-06-02       Impact factor: 2.984

7.  Comparison of gluteus medius strength between individuals with obesity and normal-weight individuals: a cross-sectional study.

Authors:  Rafael Ratti Fenato; Allan Cezar Faria Araujo; Ana Tereza Bittencourt Guimarães
Journal:  BMC Musculoskelet Disord       Date:  2021-06-25       Impact factor: 2.362

8.  Trunk movement compensation identified by inertial measurement units is associated with deficits in physical performance, muscle strength and functional capacity in people with hip osteoarthritis.

Authors:  Jesse C Christensen; David L Quammen; Justin H Rigby; Cory L Christiansen; Jennifer E Stevens-Lapsley
Journal:  Clin Biomech (Bristol, Avon)       Date:  2021-07-21       Impact factor: 2.034

9.  Reliability, responsiveness, and validity of handheld dynamometry for assessing quadriceps strength in total knee arthroplasty.

Authors:  Andrew J Kittelson; Jesse C Christensen; Brian J Loyd; Kristine L Burrows; Johna Iannitto; Jennifer E Stevens-Lapsley
Journal:  Disabil Rehabil       Date:  2020-03-06       Impact factor: 2.439

10.  Validity and reliability of isometric knee extension muscle strength measurements using a belt-stabilized hand-held dynamometer: a comparison with the measurement using an isokinetic dynamometer in a sitting posture.

Authors:  Masahiro Hirano; Munenori Katoh; Masahiro Gomi; Saori Arai
Journal:  J Phys Ther Sci       Date:  2020-02-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.