| Literature DB >> 28785480 |
Bo Mussmann1,2, Søren Overgaard2,3, Trine Torfing1,2, Morten Bøgehøj2,3, Oke Gerke4,5, Poul Erik Andersen1,2.
Abstract
BACKGROUND: Periprosthetic bone loss is considered to be a potentially contributing factor in aseptic loosening of acetabular hip components, but no studies have shown this association. The lack of association might be caused by insufficient image quality because of metal artifacts and challenges in measuring bone density (BMD) in complex anatomic structures which might be overcome using dual-energy computed tomography (DECT).Entities:
Keywords: Segmentation; bone loss; bone mineral density; computed tomography (CT); dual-energy CT; hip arthroplasty
Year: 2017 PMID: 28785480 PMCID: PMC5521355 DOI: 10.1177/2058460117719746
Source DB: PubMed Journal: Acta Radiol Open
Acquisition parameters for SECT and DECT.
| Parameter | Single-energy CT | Dual-energy CT |
|---|---|---|
| kVp | 120 | 80/140 dual |
| Tube current (mA) | 300 | 375 |
| Scan time (s) | 0.5 | 0.6 |
| Pitch | 0.984:1 | 0.984:1 |
| Collimation (mm) | 40 | 40 |
| Image acquisition (mm) | 64 × 0.625 | 64 × 0.625 |
| Scan field of view | Body large (50 cm) | Body large (50 cm) |
| Kernel | “Bone” | “Detail” |
| ASIR | 30% | Not available |
| CTDIvol (mGy) | 11.35 | 12.92 |
ASIR, adaptive statistical iterative reconstruction; CTDIvol, volumetric CT dose index.
Fig. 1.Example of an axial DECT slice with the hemispherical volume defined by the pixels included in the intersection between the circles (red) and the borders of the free-hand drawn area (blue). Underneath the specimen, five ROIs (yellow) are positioned in the calibration phantom rods.
Fig. 2.Example of a ROI divided into quadrants superimposed onto axial and coronal slices in SECT (left) and DECT (right). Beam hardening artifacts are present in both scan modes, but appear slightly more prominent in SECT (arrow).
Intra-observer analysis. Differences between repeated BMD measurements derived from SECT and DECT (units of mg K2HPO4/cm3).
| Scan mode | Volume of interest | Mean BMD measurement 1 | Mean BMD measurement 2 | Difference | 95% CI | |
|---|---|---|---|---|---|---|
| Single-energy CT | Cementless cups, n = 10 | |||||
| All quadrants | 1188 | 1180 | −8 | −40 to 24 | 0.64 | |
| Superior quadrant | 1351 | 1349 | −2 | −53 to 49 | 0.95 | |
| Inferior quadrant | 1187 | 1211 | 24 | −43 to 92 | 0.48 | |
| Anterior quadrant | 1270 | 1220 | −50 | −93 to −8 | 0.02 | |
| Posterior quadrant | 598 | 657 | 59 | 1−116 | 0.045 | |
| Cemented cups, n = 12 | ||||||
| All quadrants | 1282 | 1249 | −33 | −67 to 1 | 0.054 | |
| Superior quadrant | 1334 | 1304 | −30 | −66 to 6 | 0.11 | |
| Inferior quadrant | 1382 | 1300 | −82 | −160 to −5 | 0.04 | |
| Anterior quadrant | 1476 | 1442 | −34 | −88 to 21 | 0.22 | |
| Posterior quadrant | 962 | 921 | −41 | −106 to 24 | 0.21 | |
| Dual-energy CT | Cementless cups, n = 10 | |||||
| All quadrants | 295 | 297 | 2 | −6 to 10 | 0.60 | |
| Superior quadrant | 328 | 328 | 0 | −7 to 8 | 0.93 | |
| Inferior quadrant | 276 | 285 | 9 | −11 to 30 | 0.38 | |
| Anterior quadrant | 336 | 328 | −8 | −19 to 2 | 0.13 | |
| Posterior quadrant | 150 | 170 | 20 | −3 to 44 | 0.08 | |
| Cemented cups, n = 12 | ||||||
| All quadrants | 305 | 294 | −11 | −21 to −2 | 0.014 | |
| Superior quadrant | 309 | 310 | 1 | −16 to 18 | 0.89 | |
| Inferior quadrant | 312 | 281 | −31 | −60 to −2 | 0.037 | |
| Anterior quadrant | 360 | 349 | −11 | −31 to 9 | 0.28 | |
| Posterior quadrant | 251 | 232 | −19 | −42 to 4 | 0.10 | |
Fig. 3.Bland–Altman-plots of repeated acetabular density measurements in the cementless cup derived from SECT (left) and DECT (right). Horizontal lines indicate limits of agreement and the mean difference between the measurements. n = 10.
Fig. 4.Bland–Altman-plots of repeated acetabular density measurements in the cemented cup derived from SECT (left) and DECT (right). Horizontal lines indicate limits of agreement and the mean difference between the measurements. n = 12.
Repeatability coefficients and corresponding 95% CIs in units of mg K2PO4/cm3 for repeated BMD measurements performed with SECT and DECT. The absolute difference between the measurements would be expected to lie below RC with 95% certainty.
| Volume of interest | RCSECT | RCDECT |
|---|---|---|
| Cementless cups, n = 10 | 101 (64–160) | 25 (16–40) |
| Cemented cups, n = 12 | 117 (77–178) | 32 (21–48) |
Inter-observer analysis. Differences between average BMD measurements in cementless cups for observers 1 and 2 in SECT and DECT (units of mg K2HPO4/cm3). n = 10.
| Volume of interest | Mean BMD observer 1 | Mean BMD observer 2 | Difference | 95% CI | |
|---|---|---|---|---|---|
| Single-energy CT | |||||
| All quadrants | 1188 | 1146 | −42 | −70 to −13 | 0.005 |
| Superior quadrant | 1351 | 1339 | −12 | −56 to 32 | 0.59 |
| Inferior quadrant | 1187 | 1047 | −140 | −193 to −86 | <0.0001 |
| Anterior quadrant | 1270 | 1235 | −35 | −62 to −9 | 0.009 |
| Posterior quadrant | 598 | 620 | 22 | −66 to 110 | 0.63 |
| Dual-energy CT | |||||
| All quadrants | 295 | 285 | −10 | −14 to −5 | <0.0001 |
| Superior quadrant | 328 | 324 | −4 | −10 to 1 | 0.14 |
| Inferior quadrant | 276 | 244 | −32 | −41 to −22 | <0.0001 |
| Anterior quadrant | 336 | 331 | −5 | −11 to 1 | 0.09 |
| Posterior quadrant | 150 | 141 | −9 | −22 to 4 | 0.19 |
Fig. 5.Box plot with medians, quartiles, and ranges comparing BMD in units of mg K2PO4/cm3 for two observers (BM and KK) in the cementless concept. n = 10. Dots are outliers.
Fig. 6.Bland–Altman plots of density measurements from observers 1 and 2 derived from SECT (left) and DECT (right). Horizontal lines indicate limits of agreement and the mean difference between the observers. n = 10.