Literature DB >> 28768777

Ceramic-on-ceramic bearing fractures in total hip arthroplasty: an analysis of data from the National Joint Registry.

D P Howard1, P D H Wall2, M A Fernandez2, H Parsons2, P W Howard3.   

Abstract

AIMS: Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA) are commonly used, but concerns exist regarding ceramic fracture. This study aims to report the risk of revision for fracture of modern CoC bearings and identify factors that might influence this risk, using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man. PATIENTS AND METHODS: We analysed data on 223 362 bearings from 111 681 primary CoC THAs and 182 linked revisions for bearing fracture recorded in the NJR. We used implant codes to identify ceramic bearing composition and generated Kaplan-Meier estimates for implant survivorship. Logistic regression analyses were performed for implant size and patient specific variables to determine any associated risks for revision.
RESULTS: A total of 222 852 bearings (99.8%) were CeramTec Biolox products. Revisions for fracture were linked to seven of 79 442 (0.009%) Biolox Delta heads, 38 of 31 982 (0.119%) Biolox Forte heads, 101 of 80 170 (0.126%) Biolox Delta liners and 35 of 31 258 (0.112%) Biolox Forte liners. Regression analysis of implant size revealed smaller heads had significantly higher odds of fracture (chi-squared 68.0, p < 0.001). The highest fracture risk was observed in the 28 mm Biolox Forte subgroup (0.382%). There were no fractures in the 40 mm head group for either ceramic type. Liner thickness was not predictive of fracture (p = 0.67). Body mass index (BMI) was independently associated with revision for both head fractures (odds ratio (OR) 1.09 per unit increase, p = 0.031) and liner fractures (OR 1.06 per unit increase, p = 0.006).
CONCLUSIONS: We report the largest independent study of CoC bearing fractures to date. The risk of revision for CoC bearing fracture is very low but previous studies have underestimated this risk. There is good evidence that the latest generation of ceramic has greatly reduced the odds of head fracture but not of liner fracture. Small head size and high patient BMI are associated with an increased risk of ceramic bearing fracture. Cite this article: Bone Joint J 2017;99-B:1012-19. ©2017 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Arthroplasty; Ceramic-on-ceramic; Fracture; Hip; National Joint Registry; Tribology

Mesh:

Year:  2017        PMID: 28768777     DOI: 10.1302/0301-620X.99B8.BJJ-2017-0019.R1

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  24 in total

1.  Catastrophic ceramic liner failure-The subtle signs of a non-engaged ceramic liner.

Authors:  Andrew Walls; Adam Tucker; David S Warnock; David Beverland
Journal:  J Orthop       Date:  2018-02-20

2.  Non-traumatic Ceramic Head Fracture in Total Hip Arthroplasty with Ceramic-on-Ceramic Articulation at Postoperative 16th Years.

Authors:  Onur Kocadal; Turhan Ozler; Alp Er Tunga Bolukbasi; Faik Altintas
Journal:  Hip Pelvis       Date:  2019-05-30

3.  Third generation delta ceramic-on-ceramic bearing for total hip arthroplasty at mid-term follow-up.

Authors:  Alessandro Bistolfi; Riccardo Ferracini; Alessandro Aprato; Alessandro Massè; Walter Daghino; Sara Lea; Stefano Artiaco; Gwo-Chin Lee
Journal:  J Orthop       Date:  2020-08-28

Review 4.  [Postoperative outcomes and survival rates after aseptic revision total hip arthroplasty : What can patients expect from revision surgery?]

Authors:  Manuela Hoffmann; Johannes C Reichert; Anastasia Rakow; Janosch Schoon; Georgi I Wassilew
Journal:  Orthopadie (Heidelb)       Date:  2022-06-23

5.  Third and fourth generation ceramic-on-ceramic total hip arthroplasty at a minimum of ten years.

Authors:  Cathal J McCarthy; John Mahon; Gerard A Sheridan; Adanna Welch-Phillips; John M O'Byrne; Paddy J Kenny
Journal:  J Clin Orthop Trauma       Date:  2022-07-06

6.  A case-driven hypothesis for multi-stage crack growth mechanism in fourth-generation ceramic head fracture.

Authors:  Stefano Lucchini; Massimiliano Baleani; Federico Giardina; Andrea Martelli; Francesco Castagnini; Barbara Bordini; Francesco Traina
Journal:  J Orthop Surg Res       Date:  2022-06-03       Impact factor: 2.677

7.  Effect of Bearing Surface on Survival of Cementless and Hybrid Total Hip Arthroplasty: Study of Data in the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man.

Authors:  Edward T Davis; Joseph Pagkalos; Branko Kopjar
Journal:  JB JS Open Access       Date:  2020-05-15

8.  What Is the Frequency of Fracture of Ceramic Components in THA? Results from the Norwegian Arthroplasty Register from 1997 to 2017.

Authors:  Geir Hallan; Anne Marie Fenstad; Ove Furnes
Journal:  Clin Orthop Relat Res       Date:  2020-06       Impact factor: 4.755

9.  Third-generation pure alumina and alumina matrix composites in total hip arthroplasty: What is the evidence?

Authors:  Didier Hannouche; Matthieu Zingg; Hermes Miozzari; Remy Nizard; Anne Lübbeke
Journal:  EFORT Open Rev       Date:  2018-01-10

10.  Alumina ceramic-on-ceramic hybrid total hip arthroplasty. A median of 15 years follow-up.

Authors:  Omer M H Farhan-Alanie; Alex Hrycaiczuk; Craig Tinning; Bryn Jones; Andrew Stark; Kevin Bryceland
Journal:  Eur J Orthop Surg Traumatol       Date:  2021-08-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.