Literature DB >> 28752197

The effect of different torque wrenches on rotational stiffness in compressive femoral nails: a biomechanical study.

A A Karaarslan1, N Acar2.   

Abstract

PURPOSE: Rotation instability and locking screws failure are common problems. We aimed to determine optimal torque wrench offering maximum rotational stiffness without locking screw failure.
METHODS: We used 10 conventional compression nails, 10 novel compression nails and 10 interlocking nails with 30 composite femurs. We examined rotation stiffness and fracture site compression value by load cell with 3, 6 and 8 Nm torque wrenches using torsion apparatus with a maximum torque moment of 5 Nm in both directions. Rotational stiffness of composite femur-nail constructs was calculated.
RESULTS: Rotational stiffness of composite femur-compression nail constructs compressed by 6 Nm torque wrench was 3.27 ± 1.81 Nm/angle (fracture site compression: 1588 N) and 60% more than that compressed with 3 Nm torque wrench (advised previously) with 2.04 ± 0.81 Nm/angle (inter fragmentary compression: 818 N) (P = 0.000). Rotational stiffness of composite-femur-compression nail constructs compressed by 3 Nm torque wrench was 2.04 ± 0.81 Nm/angle (fracture site compression: 818 N) and 277% more than that of interlocking nail with 0.54 ± 0.08 Nm/angle (fracture site compression: 0 N) (P = 0.000).
CONCLUSION: Rotational stiffness and fracture site compression value produced by 3 Nm torque wrench was not satisfactory. To obtain maximum rotational stiffness and fracture site compression value without locking screw failure, 6 Nm torque wrench in compression nails and 8 Nm torque wrench in novel compression nails should be used.

Entities:  

Keywords:  Femoral fracture; Interlocking femoral nail; Rotational stability; Torque wrench

Mesh:

Year:  2017        PMID: 28752197     DOI: 10.1007/s00068-017-0823-5

Source DB:  PubMed          Journal:  Eur J Trauma Emerg Surg        ISSN: 1863-9933            Impact factor:   3.693


  20 in total

1.  Setscrew distal locking for intramedullary nails: a biomechanical study.

Authors:  N Köse; I Günal; X Wang; K A Athanasiou; C M Agrawal; J D Mabrey
Journal:  J Orthop Trauma       Date:  2000-08       Impact factor: 2.512

2.  A new compression design that increases proximal locking screw bending resistance in femur compression nails.

Authors:  Ahmet Adnan Karaarslan; Ahmet Karakaşli; Tolga Karci; Hakan Aycan; Serhat Yildirim; Erhan Sesli
Journal:  Acta Orthop Belg       Date:  2015-06       Impact factor: 0.500

3.  Biomechanical comparision of femoral intramedullary nails for interfragmentary rotational stability.

Authors:  Ahmet Adnan Karaarslan; Hakan Aycan; Aslan Mayda; Fatih Ertem; Erhan Sesli
Journal:  Eklem Hastalik Cerrahisi       Date:  2015

4.  Fatigue performance of composite analogue femur constructs under high activity loading.

Authors:  Alexander C M Chong; Elizabeth A Friis; Gregory P Ballard; Peter J Czuwala; Francis W Cooke
Journal:  Ann Biomed Eng       Date:  2007-03-28       Impact factor: 3.934

5.  Rotational stability of femoral osteosynthesis in femoral fractures - navigated measurements.

Authors:  Mustafa Citak; Daniel Kendoff; Michael J Gardner; Markus Oszwald; Padhraig F O'Loughlin; Lucien C Olivier; Christian Krettek; Tobias Hüfner; Musa Citak
Journal:  Technol Health Care       Date:  2009       Impact factor: 1.285

6.  Interlocking compression nailing: a report on 402 applications.

Authors:  O Gonschorek; G O Hofmann; V Bühren
Journal:  Arch Orthop Trauma Surg       Date:  1998       Impact factor: 3.067

7.  Biomechanical study of nine different tibia locking nails.

Authors:  P Schandelmaier; C Krettek; H Tscherne
Journal:  J Orthop Trauma       Date:  1996       Impact factor: 2.512

8.  [Treatment of femoral shaft fractures and pseudoarthrosis with compressive and interlocking intramedullary nailing].

Authors:  Erhan Yilmaz; Lokman Karakurt; Mehmet Bulut; Oktay Belhan; Erhan Serin
Journal:  Acta Orthop Traumatol Turc       Date:  2005       Impact factor: 1.511

9.  Comparison of cyclic loading versus constant compression in the treatment of long-bone fractures in rabbits.

Authors:  J W Wolf; A A White; M M Panjabi; W O Southwick
Journal:  J Bone Joint Surg Am       Date:  1981-06       Impact factor: 5.284

10.  Mechanical influences on tibial fracture healing.

Authors:  J Kenwright; T Gardner
Journal:  Clin Orthop Relat Res       Date:  1998-10       Impact factor: 4.176

View more