| Literature DB >> 28729372 |
Justis P Ehlers1, Atsuro Uchida1, Sunil K Srivastava1.
Abstract
Intraoperative optical coherence tomography (OCT) is a potentially transformational technology that is now commercially available to ophthalmic surgeons. Currently, the use of the technology is primarily limited to a 'stop and image' approach due to the lack of OCT compatibility with surgical instrumentation. In this report, we describe multiple OCT-compatible surgical instruments that were developed for various surgical needs, based on previous evaluation of potential surgical materials for optical features and physical properties. OCT-compatible instrumentation included two membrane scrapers, a surgical pick and vitreoretinal forceps. Imaging during in vitro and ex vivo-simulated surgical procedures demonstrated excellent visualisation of the instrument tip and of the tissue-instrument interaction. These OCT-compatible instruments may be a key component to the feasibility of real-time image-guided surgery with intraoperative OCT. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.Entities:
Keywords: imaging; retina; treatment surgery
Mesh:
Year: 2017 PMID: 28729372 PMCID: PMC5675732 DOI: 10.1136/bjophthalmol-2017-310530
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638