Transapical aortic valve implantation using a Symetis Acurate self-expandable bioprosthesis: initial outcomes of 10 patients.

Marcin Misterski1, Mateusz Puślecki1, Marek Grygier2, Anna Olasińska-Wiśniewska2, Maciej Lesiak2, Aleksander Araszkiewicz2, Barłomiej Perek1, Aneta Choręziak1, Jacek Lindner1, Anna Komosa2, Piotr Buczkowski1, Marcin Ligowski1, Sławomir Katarzyński1, Marek Jemielity1.   

Abstract

INTRODUCTION: Transapical aortic valve implantation (TA-AVI) has been widely introduced for treatment of patients with severe aortic stenosis in the last decade. Here we report our first clinical experience with 10 patients using the second-generation transapical Symetis Acurate TA aortic valve designed for transapical implantation. AIM: To evaluate the results of transapical access in transcatheter aortic valve implantation (TAVI) among patients with unsuitable vascular access.
MATERIAL AND METHODS: All patients had been assessed by a local heart team and were disqualified from surgical aortic valve replacement (AVR) and the transfemoral TAVI approach. Mean age was 75.4 ±3.9 years (range: 68-80), with 20% being female. Logistic EuroSCORE (European System for Cardiac Operative Risk Evaluation) and STS (Society of Thoracic Surgeons) were 15.4 ±8.9% and 20.5 ±4.5%, respectively.
RESULTS: All implantations were performed successfully in the intra-annular and subcoronary position. There were no conversions to surgical AVR. All patients survived 30-day follow-up. No strokes or transient ischemic attacks were reported. There was no need for pacemaker implantation and none of the patients demonstrated moderate or significant paravalvular leakage. The mean aortic gradients improved significantly from a baseline of 57.0 ±19.2 mm Hg to a 30-day value of 14.2 ±4.1 mm Hg.
CONCLUSIONS: Our initial clinical results indicate satisfactory functionality in patients after trans-apical implantation of the Symetis Acurate aortic valve. The procedure of implantation seems to be straightforward and may be considered in patients in whom a transfemoral approach is not a good option.

Entities:  

Keywords:  Symetis Acurate valve; aortic stenosis; transapical valve implantation; transcatheter aortic valve implantation

Year:  2017        PMID: 28694904      PMCID: PMC5502338          DOI: 10.5114/wiitm.2017.67481

Source DB:  PubMed          Journal:  Wideochir Inne Tech Maloinwazyjne        ISSN: 1895-4588            Impact factor:   1.195


Introduction

Trans-catheter aortic valve implantation (TAVI) has evolved rapidly in the last decade and has become a routine procedure in many centers for treatment of high-risk patients who are not suitable candidates for surgical aortic valve replacement (SAVR) [1]. The potential benefits are minimally invasive access and off-pump valve implantation. However, SAVR, even in elderly patients, is still a good alternative, presenting excellent results and 30-day mortality rates of about 2% [2]. The TAVI should be reserved for patients who are not surgical candidates and for extremely high-risk patients as far as the EACTS and ESC guidelines are concerned [3]. The TAVI can be performed by using a retrograde transfemoral (TF), transaortic/direct aortic, trans-subclavian, transcarotid or antegrade transapical (TA) approach. The transfemoral method is the most frequently chosen approach as it is thought to be the least invasive. However, [4] problems with vascular access or anatomical malformations still disqualify some patients – an alternative approach should be then selected. The first generation of TAVI valves demonstrated its safety and effectiveness in various clinical trials [5]. The next-generation of transcatheter valves improved the results by way of feasibility for anatomically correct positioning, repositioning, and retrievability [6]. Here we report our clinical experience with the first 10 patients using the second-generation transapical valve Symetis Acurate (Symetis Inc., Switzerland). All patients were initially disqualified from the femoral approach.

Aim

The aim of the study was to evaluate the results of transapical access in TAVI among the patients with unsuitable vascular access.

Material and methods

Study group

From August 2014 until January 2016, a total of 10 patients were operated on using the Symetis Acurate valve. Mean age was 75.4 ±3.9 years (range: 68–80) with 20% being female. Logistic EuroSCORE (European System for Cardiac Operative Risk Evaluation) and STS (Society of Thoracic Surgeons) were 15.4 ±8.9% and 20.5 ±4.5%, respectively. Patients’ characteristics are presented in Table I.
Table I

Preoperative patients’ demographics (n = 10)

ParameterResult
Age [years]75.4 ±3.9 (range: 68–80)
Logistic EuroSCORE (%)15.4 ±8.9
STS score (%)20.5 ±4.5
Re-do procedure, n (%)4 (40)
Female, n (%)2 (20)
Left ventricular EF (%)57.6 ±10.8
Peripheral vascular disease, n (%)7 (70)
Carotid artery stenosis, n (%)5 (50)
Stroke, n (%)None
Chronic obstructive pulmonary disease, n (%)2 (20)
Mean aortic gradient before procedure [mm Hg]57.0 ±19.2
Maximal aortic gradient before procedure [mm Hg]91.6 ±23.2
Aortic valve opening area [cm2]0.71 ±0.07
NYHA class before procedure, n:
 I0
 II2
 III8
 IV0

STS – Society of Thoracic Surgeons, EF – ejection fraction, NYHA – New York Heart Association functional classification. Data are presented as mean ± standard deviation or number/percentage.

Preoperative patients’ demographics (n = 10) STS – Society of Thoracic Surgeons, EF – ejection fraction, NYHA – New York Heart Association functional classification. Data are presented as mean ± standard deviation or number/percentage. All patients preoperatively presented as class II or III in accordance with the New York Heart Association (NYHA) classification. The clinical decision to perform TAVI by the transapical route (TA-TAVI) was settled by a multidisciplinary heart team after discussing all available options. All patients were informed about the proposed method of treatment individually and signed informed consent before the procedure.

Methods (Trans-catheter Valve Implantation Technique)

All procedures were done under general anesthesia in a hybrid room. Implants were performed off-pump under fluoroscopic guidance with transesophageal echocardiography as an additional control. Cardiopulmonary bypass was prepared for emergency intervention if necessary. All the procedures were performed by the same heart team involving two cardiac surgeons, a cardiologist, cardiac anesthetists, a perfusionist, and a scrub nurse. A temporary endocardial electrode was placed into the right ventricle from the jugular vein. Apical access was obtained using a left anterior mini-thoracotomy that was usually performed at the fifth or sixth intercostal space (Photo 1). The skin incision was approximately 6 cm long. Soft tissue retractors were used for rib spreading. Once the left ventricular apex was visualized and exposed, two purse-string sutures of 3-0 polypropylene were placed and reinforced with Teflon pledgets. The optimal C-arm angulation was established according to the DynaCT preoperative measurements. In addition, a femoral arterial 6 Fr sheath was inserted and a pigtail catheter was introduced into the ascending aorta. The apex was punctured under direct visualization of the surgeon and a soft guide wire was introduced across the stenotic aortic valve under fluoroscopic control. The 6 Fr sheath was inserted into the left ventricle and a left Amplatz catheter was introduced over the soft wire. The soft wire was subsequently exchanged for a stiff wire (Amplatz Super stiff, 0.35’’, 260 cm, Boston Scientific). After guidewire placement a 14 Fr. sheath balloon valvuloplasty was performed during rapid ventricular pacing with an Osypka balloon (Osypka AG, Germany) sized 1 mm less than the measured diameter of the annulus. At the same time, a preselected valve was crimped and its delivery system was de-aired using carbon dioxide insufflation. Aortic valvuloplasty was performed under ventricular rapid pacing of 180/min once the systemic systolic pressure dropped below 50 mm Hg. The delivery system was bluntly introduced and positioned in the aortic annulus (Photo 2). By rotating radiopaque visibility markers, the anatomical position was reached. Step one had been performed – the stabilization arches and upper crown were released and a final angiography confirmed the proper valve position. The tactile feedback facilitated proper annular orientation. Finally, the “safety button” was removed and step two of the implantation was completed. Once the Symetis Acurate valve was fully deployed (Photo 3) and the delivery system removed, the apical purse-string sutures were tied. After introducing an 18F chest tube drain into the left pleural cavity, routine chest wall closure was performed and the patient was transferred to the postoperative intensive care unit for early extubation.
Photo 1

Apical access gained by a left anterior mini-thoracotomy in the fifth or sixth intercostal space

Photo 2

Positioning of a Symetis Acurate in the aortic annulus under fluoroscopic guidance

Photo 3

Final valve deployment – last step of implantation process

Apical access gained by a left anterior mini-thoracotomy in the fifth or sixth intercostal space Positioning of a Symetis Acurate in the aortic annulus under fluoroscopic guidance Final valve deployment – last step of implantation process

Statistical analysis

For statistical analysis all relevant data were recorded. Continuous variables are presented as the mean plus the standard deviation. Categorical data are expressed as proportions.

Results

All 10 patients survived the surgery and all valves were implanted successfully. There were no conversions to open heart surgery and all valve implantations were performed off-pump. In two cases a post-dilatation was performed to optimize the procedural result. The S, M and L valves were implanted in 1, 5 and 4 patients respectively. There was one intraoperative apical bleeding event requiring additional surgical suture placement which was successfully managed. One patient was re-operated on because of increased postoperative drainage due to intercostal small vessel bleeding, which was localized and coagulated. There were no neurological complications of the procedure. Postoperative kidney failure requiring temporary hemodialysis was observed in one patient. All patients were discharged between the 7th and 10th postoperative day. There was no need for permanent pacemaker implantations. A trivial paravalvular leak was noted in 4 patients. In the 30-day postoperative period no deaths were recorded, but one patient was urgently treated for an abdominal aortic aneurysm (AAA) rupture (Tables II and III).
Table II

Intra-operative data (n = 10)

VariableResult
Successful valve implantation, n (%)10 (100)
Acurate sizes, n (%):
 Small1 (10)
 Medium5 (50)
 Large4 (40)
Mean aortic gradient after procedure [mm Hg]14.2 ±4.1
Maximal aortic gradient after procedure [mm Hg]27.9 ±8.8
Bleeding, n (%):
 MajorNone
 Minor1 (10)

Data are presented as mean ± standard deviation or number/percentage.

Table III

NYHA class after TAVI

NYHA beforeNYHA afterPatients, n (%)
III2 (20)
IIIII8 (80)

NYHA – New York Heart Association functional classification. Data are presented as number/percentage.

Intra-operative data (n = 10) Data are presented as mean ± standard deviation or number/percentage. NYHA class after TAVI NYHA – New York Heart Association functional classification. Data are presented as number/percentage.

Discussion

This paper presents our first experience with transapical trans-catheter aortic valve implantations (TA-TAVI). Our center established the TAVI program in 2012. Following initial experience with CoreValve (Medtronic) and Lotus Valve (Boston Scientific), using the retrograde femoral and direct aortic access with encouraging results, we began with the transapical (TA) procedure in September 2014. All patients who were operated on using the transapical approach were selected only when there was unsuitable vascular access. Although we continue to advocate transfemoral access as the first-line approach because of its minimally invasive nature, the transapical anterograde access has its advantages and benefits [7]. The superiority of one method over the other is not proven, and a randomized trial will probably never be performed. The potential advantages associated with an anterograde approach are many, and include the following: minimal aortic manipulation reflected in the low complication rates for strokes; less limitation regarding the sheath diameter, which avoids excessively tight valve crimping that can impact the long-term valve durability; and finally, as a result of shorter “wire” distance, the valve positioning seems to be easier and may lead to better procedural results [8, 9]. The Symetis Acurate valve is a self-positioning porcine bioprosthesis. The self-expanding nitinol stent is composed of three stabilization arms providing valve stabilization in the ascending aorta during deployment. The porcine tissue valve is sewn into a nitinol stent and is covered by a PET skirt. The valve is designed to obtain an intra-annular, sub-coronary position. Tactile feedback is achieved with a distal part of the stent, called the upper crown, which when optimally positioned produces a slight feeling of tension on the delivery system to securely lock it in place. The delivery catheter is designed for an easy two-step sheath-less implantation which facilitates the release of stabilization arms and the upper crown in step one [10], while step two deploys the body of the valve. The Symetis Acurate valve is available in three different sizes which are described as S, M, and L and which cover an aortic annulus diameter from 21 to 27 mm. The Symetis Acurate represents a significant improvement in almost every field of TAVI research over the first generation transapical TAVI valves [10]. The important issue to address are paravalvular leaks (PVL) and their influence on patient survival, which are already well known. A general consensus is that mild leakage is well tolerated, but moderate and severe PVL can have a direct impact on the 30-day hospital mortality and long-term survival of the patient [11]. The data from the German TAVI registry confirm this statement and highlight the necessity of avoiding residual leaks after TAVI procedures [12]. As already mentioned, the observed rate of PVL among our patients is very low due in part to the specific design of stents with a polyester skirt that cultivate self-sealing to the native valve and hinder paravalvular leakage. Current drawbacks for implantation of first generation TAVI devices, especially CoreValve, are the frequent occurrences of new AV blocks that necessitate pacemaker implantation [11, 13]. Based on our clinical data and many other referenced publications, the need for new pacemaker implantation in patients following Accurate implantation is infrequent. The final effect of a small 6 cm long thoracotomy for the placement of an Accurate valve is a better alternative to a pacemaker implantation that was observed in 7.5% of patients in Kempfert’s study [14] in comparison with the majority of transfemoral studies using different valves. The greatest potential advantage of the Acurate is the feasibility of achieving commissural alignment and precise anatomical positioning. This could alleviate the need for future coronary PCI procedures. Additional interventional procedures are potentially dangerous, especially when a valve strut is in front of a coronary ostium with concomitant severe calcification, further increasing the odds for a coronary obstruction. The commissural alignment may be beneficial, although there is no scientific evidence to support this [10, 14]. Significant simplification is associated with a C-arm position. While achieving the annulus plane the operators leave the C-arm in one position throughout the whole procedure. This is especially useful for physicians who have limited C-arm experience. Last but not least, the preparation of the valve seems to be easy. The crimping phase is simple and introduction of a three-step procedure can be managed by a physician or a nurse in a short time [15]. The short-term results of the first registry of Acurate TA proved a safety and efficacy profile comparable, if not improved in some aspects, with previously approved transcatheter aortic valve implantation devices [16]. Comparison of two valve systems – Acurate TA vs Sapien XT – in a propensity score matched analysis demonstrated comparable hemodynamic performance and clinical outcome. The self-expendable valve required more frequent post-ballooning without affecting the safety profile [17]. Meanwhile more than 3000 patients have been treated worldwide with transapical and transfemoral Symetis valves, and additional registries are currently under investigation [18].

Conclusions

This report documents the first experience in Wielkopolska with the anterograde Symetis Acurate second generation valve. The functional hemodynamic outcomes are very good, with low residual gradients and very low rates of serious paravalvular leaks. With its promising initial results and relatively low rate of complications, the Symetis Acurate seems to be a good option for patients with poor femoral access.

Conflict of interest

The authors declare no conflict of interest.
  18 in total

1.  Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery.

Authors:  Martin B Leon; Craig R Smith; Michael Mack; D Craig Miller; Jeffrey W Moses; Lars G Svensson; E Murat Tuzcu; John G Webb; Gregory P Fontana; Raj R Makkar; David L Brown; Peter C Block; Robert A Guyton; Augusto D Pichard; Joseph E Bavaria; Howard C Herrmann; Pamela S Douglas; John L Petersen; Jodi J Akin; William N Anderson; Duolao Wang; Stuart Pocock
Journal:  N Engl J Med       Date:  2010-09-22       Impact factor: 91.245

2.  Comparison of two valve systems for transapical aortic valve implantation: a propensity score-matched analysis.

Authors:  Jörg Kempfert; Alexander Meyer; Won-Keun Kim; Arnaud Van Linden; Mani Arsalan; Johannes Blumenstein; Helge Möllmann; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2015-02-10       Impact factor: 4.191

3.  High-risk aortic valve replacement: are the outcomes as bad as predicted?

Authors:  Eugene A Grossi; Charles F Schwartz; Pey-Jen Yu; Ulrich P Jorde; Gregory A Crooke; Juan B Grau; Greg H Ribakove; F Gregory Baumann; Patricia Ursumanno; Alfred T Culliford; Stephen B Colvin; Aubrey C Galloway
Journal:  Ann Thorac Surg       Date:  2008-01       Impact factor: 4.330

4.  Dyna-CT during minimally invasive off-pump transapical aortic valve implantation.

Authors:  Jörg Kempfert; Volkmar Falk; Gerhard Schuler; Axel Linke; Denis Merk; Friedrich W Mohr; Thomas Walther
Journal:  Ann Thorac Surg       Date:  2009-12       Impact factor: 4.330

5.  First registry results from the newly approved ACURATE TA™ TAVI system†.

Authors:  Jörg Kempfert; David Holzhey; Steffen Hofmann; Evaldas Girdauskas; Hendrik Treede; Holger Schröfel; Matthias Thielmann; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2014-11-25       Impact factor: 4.191

6.  Symetis Acurate Transapical Aortic Valve: the initial experience with a second generation of transcatheter aortic valve replacement device.

Authors:  Michał Oskar Zembala; Jacek Piegza; Jacek Wacławski; Michał Hawranek; Michael Hilker; Tomasz Niklewski; Jan Głowacki; Monika Parys; Paweł Nadziakiewicz; Piotr Chodór; Krzysztof Wilczek; Roman Przybylski; Mariusz Gąsior; Marian Zembala
Journal:  Kardiol Pol       Date:  2014       Impact factor: 3.108

7.  Trans-apical aortic valve implantation using a new self-expandable bioprosthesis: initial outcomes.

Authors:  Jörg Kempfert; Ardawan J Rastan; Friedhelm Beyersdorf; Markus Schönburg; Gerhard Schuler; Stefan Sorg; Friedrich-W Mohr; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2011-09-15       Impact factor: 4.191

8.  Thirty-day results of the SAPIEN aortic Bioprosthesis European Outcome (SOURCE) Registry: A European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve.

Authors:  Martyn Thomas; Gerhard Schymik; Thomas Walther; Dominique Himbert; Thierry Lefèvre; Hendrik Treede; Holger Eggebrecht; Paolo Rubino; Iassen Michev; Rüdiger Lange; William N Anderson; Olaf Wendler
Journal:  Circulation       Date:  2010-06-21       Impact factor: 29.690

9.  Aortic regurgitation after transcatheter aortic valve implantation: incidence and early outcome. Results from the German transcatheter aortic valve interventions registry.

Authors:  Mohamed Abdel-Wahab; Ralf Zahn; Martin Horack; Ulrich Gerckens; Gerhard Schuler; Horst Sievert; Holger Eggebrecht; Jochen Senges; Gert Richardt
Journal:  Heart       Date:  2011-03-12       Impact factor: 5.994

10.  Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome.

Authors:  Eberhard Grube; Gerhard Schuler; Lutz Buellesfeld; Ulrich Gerckens; Axel Linke; Peter Wenaweser; Barthel Sauren; Friedrich-Wilhelm Mohr; Thomas Walther; Bernfried Zickmann; Stein Iversen; Thomas Felderhoff; Raymond Cartier; Raoul Bonan
Journal:  J Am Coll Cardiol       Date:  2007-06-06       Impact factor: 24.094

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.