Abstract
INTRODUCTION: The aim of this study was to assess the analytical performances of the newly developed Access2 25-hydroxyvitamin D (25(OH)D) total immunoassay on two analysers, DxI800 and Access2 (Beckman Coulter, Brea, CA, USA), and compare these two and a recalibrated Modular E 170 25(OH)D assay (Roche Diagnostics, Penzberg, Germany) with reference liquid chromatography tandem-mass spectrometry (LC-MS/MS) with special emphasis on clinical diagnosis.Entities:
Keywords: 25-Hydroxyvitamin D; method comparison; standardization; vitamin D
Mesh:
Substances:
Year: 2017 PMID: 28694724 PMCID: PMC5493169 DOI: 10.11613/BM.2017.036
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Performance characteristics of methods for 25(OH)D measurements
| Low | 8.9 | 8.1 | 5.2 | 3.9 | |
| High | 8.2 | 7.7 | 2.2 | 1.7 | |
| RIQAS 1 | - 23.82 | - 23.49 | 1.46 | 30.12 | |
| RIQAS 2 | - 1.08 | - 7.13 | 5.88 | 62.82 | |
| RIQAS 3 | - 16.30 | 0.11 | 7.78 | 29.60 | |
| - | 2.48 | 1.92 | 4.99† | Unknown | |
| 4.64 | 5.67 | 7.49† | 1.60 | ||
| 8.61 | 8.84 | 12.48† | 5.37 | ||
| - | 23.5–525 | 17.5–412 | 7.5–175† | 5.3–349 | |
| Hemoglobin (g/L) | > 0.50 | > 0.50 | 0.20† | * | |
| Triglyceride (mmol/L) | > 7.67‡ | > 7.67‡ | > 4.52† | * | |
| Bilirubin (umol/L) | > 513 | > 513 | > 1129† | * | |
| - | 0.276 | 0.001 | Not identified† | - | |
| *No interference. †Values claimed by the manufacturer. ‡Manufacturer claims were not met (Manufacturer claim was 37.06 mmol/L). RIQAS – Randox External Quality Assessment (EQA)/ Proficiency Testing (PT) scheme. LoB – Limit of blank. LoD – limit of detection. LoQ – limit of quantification. | |||||
Figure 1Box and whisker plots of 4 methods (Access2, Dxl 800, E170 and LC-MS/MS) for 25(OH)D measurements. Median values for Access2, DXL 800, E170 and LC-MS/MS were 50.44, 56.36, 59.77 and 67.39 nmol/L, respectively, P = 0.04. Each box shows 25 to 75 percentile range. Horizontal line in each box represents group median.
Figure 2Bland-Altman plots of method comparison for 25(OH)D measurements. Mean (thick solid line) - percentage bias (means of paired differences). Dashed lines demonstrate the 95% limits of agreement (bias ± 1.96 standard deviation). Thin dotted line - line of equality.
Figure 3Passing-Bablok regression analysis of three 25(OH)D immunoassays compared to LC-MS/MS. The regression equation is presented as y = a + bx. a - regression line’s intercept. b - regression line’s slope.
Method comparison results of immunoassays for 25(OH)D measurements with LC-MS/MS
| Slope (CI) | Intercept (CI) | CCC (CI) | P | Cb | Bias (%) | Κappa (CI) | |
| 0.878 | 0.420 | 0.941 | 0.955 | 0.986 | - 8.6 | 0.71 | |
| 0.748 | 0.697 | 0.854 | 0.950 | 0.899 | - 19.2 | 0.57 | |
| 0.960 | - 2.797 | 0.901 | 0.917 | 0.983 | - 12.2 | 0.77 | |
| CI – confidence interval. *Slope CI’s don’t include 1.0 meaning proportional error. †intercept CI’s don’t include 0 meaning constant error. Cb - Bias correction factor (accuracy). CCC - Concordance correlation coefficient. Ρ - Pearson correlation coefficient (precision). | |||||||