Characteristics and performance of the first commercial multileaf collimator for a robotic radiosurgery system.

Christoph Fürweger1, Paulette Prins2, Harun Coskan2, Ben J M Heijmen2.   

Abstract

PURPOSE: The "InCise™ multileaf-collimator (MLC)" is the first commercial MLC to be mounted on a robotic SRS/SBRT platform (CyberKnife). The authors assessed characteristics and performance of this novel device in a preclinical five months test period.
METHODS: Commissioning beam data were acquired with unshielded diodes. EBT3 radiochromic films were employed for measurement of transmission, leaf/bank position accuracy (garden fence) before and after exercising the MLC, for end-to-end testing and further characterization of the beam. The robot workspace with MLC was assessed analytically by transformation to an Euler geometry ("plane," "gantry," and "collimator" angles) and by measuring pointing accuracy at each node. Stability over time was evaluated in picket fence and adapted Winston-Lutz tests (AQA).
RESULTS: Beam penumbrae (80%-20%, with 100% = 2 × dose at inflection point for field sizes ≥ 50 × 50 mm(2)) were 2.2-3.7 mm for square fields in reference condition (source-axis-distance 800 mm, depth 15 mm) and depended on field size and off-axis position. Transmission and leakage did not exceed 0.5%. Accessible clinical workspace with MLC covered non-coplanar gantry angles of [-113°; +112°] and collimator angles of [-100°; +107°], with an average robot pointing accuracy of 0.12 ± 0.09 mm. For vertical beams, garden fence tests exhibited an average leaf positioning error of ≤0.2 mm, which increased by 0.25 and 0.30 mm (banks X1 and X2) with leaves traveling parallel to gravity. After execution of a leaf motion stress routine, garden fence tests showed slightly increased jaggedness and allowed to identify one malfunctioning leaf motor. Total system accuracy with MLC was 0.38 ± 0.05 mm in nine end-to-end tests. Picket fence and AQA tests displayed stable results over the test period.
CONCLUSIONS: The InCise™ MLC for CyberKnife showed high accuracy and adequate characteristics for SRS/SBRT applications. MLC performance after exercise demands specific quality assurance measures.

Entities:  

Mesh:

Year:  2016        PMID: 27147318     DOI: 10.1118/1.4944740

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Treatment planning for spinal radiosurgery : A competitive multiplatform benchmark challenge.

Authors:  Christos Moustakis; Mark K H Chan; Jinkoo Kim; Joakim Nilsson; Alanah Bergman; Tewfik J Bichay; Isabel Palazon Cano; Savino Cilla; Francesco Deodato; Raffaela Doro; Jürgen Dunst; Hans Theodor Eich; Pierre Fau; Ming Fong; Uwe Haverkamp; Simon Heinze; Guido Hildebrandt; Detlef Imhoff; Erik de Klerck; Janett Köhn; Ulrike Lambrecht; Britta Loutfi-Krauss; Fatemeh Ebrahimi; Laura Masi; Alan H Mayville; Ante Mestrovic; Maaike Milder; Alessio G Morganti; Dirk Rades; Ulla Ramm; Claus Rödel; Frank-Andre Siebert; Wilhelm den Toom; Lei Wang; Stefan Wurster; Achim Schweikard; Scott G Soltys; Samuel Ryu; Oliver Blanck
Journal:  Strahlenther Onkol       Date:  2018-05-25       Impact factor: 3.621

Review 2.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

3.  Tumor-dose-rate variations during robotic radiosurgery of oligo and multiple brain metastases.

Authors:  Maria-Lisa Wilhelm; Mark K H Chan; Benedikt Abel; Florian Cremers; Frank-Andre Siebert; Stefan Wurster; David Krug; Robert Wolff; Jürgen Dunst; Guido Hildebrandt; Achim Schweikard; Dirk Rades; Floris Ernst; Oliver Blanck
Journal:  Strahlenther Onkol       Date:  2020-06-25       Impact factor: 3.621

4.  Radiotoxicity in robotic radiosurgery: proposing a new quality index for optimizing the treatment planning of brain metastases.

Authors:  Alexandra Hellerbach; Klaus Luyken; Mauritius Hoevels; Andreas Gierich; Daniel Rueß; Wolfgang W Baus; Martin Kocher; Maximilian I Ruge; Harald Treuer
Journal:  Radiat Oncol       Date:  2017-08-17       Impact factor: 3.481

5.  Clinical usefulness of MLCs in robotic radiosurgery systems for prostate SBRT.

Authors:  Masashi Tomida; Takeshi Kamomae; Junji Suzuki; Yoichi Ohashi; Yoshiyuki Itoh; Hiroshi Oguchi; Takahito Okuda
Journal:  J Appl Clin Med Phys       Date:  2017-07-10       Impact factor: 2.102

6.  Largely reduced OAR doses, and planning and delivery times for challenging robotic SBRT cases, obtained with a novel optimizer.

Authors:  Marta K Giżyńska; Linda Rossi; Wilhelm den Toom; Maaike T W Milder; Kim C de Vries; Joost Nuyttens; Ben J M Heijmen
Journal:  J Appl Clin Med Phys       Date:  2021-01-21       Impact factor: 2.102

7.  Stereotactic Arrhythmia Radioablation (STAR) of Ventricular Tachycardia: A Treatment Planning Study.

Authors:  Lei Wang; Benjamin Fahimian; Scott G Soltys; Paul Zei; Anthony Lo; Edward A Gardner; Patrick J Maguire; Billy W Loo
Journal:  Cureus       Date:  2016-07-15

8.  Commissioning and clinical implementation of the first commercial independent Monte Carlo 3D dose calculation to replace CyberKnife M6™ patient-specific QA measurements.

Authors:  Maaike T W Milder; Markus Alber; Matthias Söhn; Mischa S Hoogeman
Journal:  J Appl Clin Med Phys       Date:  2020-10-25       Impact factor: 2.243

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.