Cardiovascular mortality is grossly elevated in patients with chronic kidney disease (CKD), and is associated with a wide variety of structural and functional abnormalities. These issues have driven additional attempts to further characterise these abnormalities to elucidate the pathophysiology involved, assess individual risk and/or target and monitor therapies specifically directed at the cardiovascular (CV) system. This review aims to assess the techniques that are currently available for the study of the CV system. This includes conventional assessments of the whole CV system from heart to peripheral microcirculation (although not deal with VC assessment), as well as the key functional consequences relating to stress induced cardiovascular reserve, perfusion and vasoregulation. In addition this review will introduce a variety of techniques aiming to expand the envelope of conventional measurements.Entities:
Keywords: cardiovascular disease; chronic kidney disease; imaging; metabolic stress; monitoring
Year: 2008 PMID: 28657024 PMCID: PMC5477876 DOI: 10.1093/ndtplus/sfn146
Source DB: PubMed Journal: NDT Plus ISSN: 1753-0784
Devices available for the assessment of arterial stiffness
| Device | Manufacturer | Technique | Assesses | Benefits | Issues |
|---|---|---|---|---|---|
| SphygmoCor | AtCor Medical | Applanation tonometry + ECG | PWVPWA (AIx)Central BP | Can assess any arterial segment Calculation of central BP from peripheral site | Operator dependent, training and validation required |
| Vicorder | Skidmore Medical | Simultaneous BP cuff | PWV | Rapid assessment | New—validation in progress No PWA |
| PulseTrace PCA2 | Cardinal Heath (MicroMedical) | Digital photoplethysmography | PWV via PWA | Rapid assessment Portable | Reproducibility and correlation to PWV |
| PulseTrace PWV | Cardinal Heath (MicroMedical) | Doppler + ECG | PWV | Can assess any arterial segment | Outcome data |
| Complior | Artech Medical | Simultaneous pulse sensors | PWVCentral BP | Can assess any arterial segment | — |
| DynaPulse 200 | PulseMetric | Brachial artery distensibility via BP cuff | PWACentral BP + haemodynamics | Rapid assessment Can be used in patient's home | No direct assessment of PWV |
| CVProfilor DO-2020 CVProfilor MD-3000 HDI/Pulsewave CR-2000 | HDI Diagnostics | Radial automated applanation Tonometry + brachial BP cuff | PWACentral haemodynamics (CR-2000) | Differentiates large + small artery stiffness | No direct assessment of PWV |
| HEM-9000 AI | Omron | Automated radial applanation tonometry | PWA (AIx) | Automated assessment | No PWV or central BP |