Literature DB >> 28581411

Computerized System for Safety Verification of External Beam Radiation Therapy Planning.

Clay Holdsworth1, Jacek Kukluk2, Christina Molodowitch2, Maria Czerminska2, Cindy Hancox2, Robert A Cormack2, Kevin Beaudette2, Joseph H Killoran2.   

Abstract

PURPOSE: To report an assessment of in-house software, Verifier, developed to improve efficacy and efficiency of the radiation therapy (RT) treatment planning process and quality control review (QCR). METHODS AND MATERIALS: Radiation therapy plan parameters retrieved from our treatment planning database are used by automated tests to give 75 types of warnings, such as prescription and plan discrepancies. The software is continuously updated on the basis of new issues, ideas, and planning policies. Verifier was retrospectively assessed (2007-2015) by examining impact on treatment plan revisions, frequency of quality improvement incident reports of avoidable RT plan-related safety events, unaddressed issues, and staff efficiency.
RESULTS: Plan revisions for specific issues declined dramatically in response to implementation of corresponding Verifier tests. Between 2012 and 2015 our institution's total rate of plan revisions dropped from 18.0% to 11.2%. Between 2008 and 2015 specific tests were added to Verifier while the rate of corresponding avoidable safety events was reduced from 0.34% to 0.00% over the same period. Simulations suggest Verifier saves approximately 2 to 5 minutes per QCR.
CONCLUSIONS: The decrease in quantifiable metrics of plan revisions and incident reports suggests automatic RT plan-checking software enhances patient safety and clinical efficiency. Although only modest time savings may be gained using Verifier for the QCR itself, the greater impact on efficiency is through avoiding late-stage plan modifications and improving documentation via automation. We encourage other institutions to consider working toward adding similar technologies to enhance their RT quality assurance programs.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28581411     DOI: 10.1016/j.ijrobp.2017.03.001

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Cautiously optimistic: A survey of radiation oncology professionals' perceptions of automation in radiotherapy planning.

Authors:  Vikneswary Batumalai; Michael G Jameson; Odette King; Rhiannon Walker; Chelsea Slater; Kylie Dundas; Glen Dinsdale; Andrew Wallis; Cesar Ochoa; Rohan Gray; Phil Vial; Shalini K Vinod
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2020-11-17

2.  Early detection of potential errors during patient treatment planning.

Authors:  Danielle Lack; Jian Liang; Lisa Benedetti; Cory Knill; Di Yan
Journal:  J Appl Clin Med Phys       Date:  2018-07-05       Impact factor: 2.102

3.  Optimizing efficiency and safety in external beam radiotherapy using automated plan check (APC) tool and six sigma methodology.

Authors:  Shi Liu; Karl K Bush; Julian Bertini; Yabo Fu; Jonathan M Lewis; Daniel J Pham; Yong Yang; Thomas R Niedermayr; Lawrie Skinner; Lei Xing; Beth M Beadle; Annie Hsu; Nataliya Kovalchuk
Journal:  J Appl Clin Med Phys       Date:  2019-08       Impact factor: 2.102

4.  Toward automation of initial chart check for photon/electron EBRT: the clinical implementation of new AAPM task group reports and automation techniques.

Authors:  Huijun Xu; Baoshe Zhang; Mariana Guerrero; Sung-Woo Lee; Narottam Lamichhane; Shifeng Chen; Byongyong Yi
Journal:  J Appl Clin Med Phys       Date:  2021-03-11       Impact factor: 2.102

5.  Saudi Radiology Personnel's Perceptions of Artificial Intelligence Implementation: A Cross-Sectional Study.

Authors:  Abdulaziz A Qurashi; Rashed K Alanazi; Yasser M Alhazmi; Ahmed S Almohammadi; Walaa M Alsharif; Khalid M Alshamrani
Journal:  J Multidiscip Healthc       Date:  2021-11-23

6.  Automation of DVH Constraint Checks and Physics Quality Control Review Improves Patient Safety in Radiotherapy.

Authors:  Isak Wahlstedt; Nikolaj Jensen
Journal:  J Med Phys       Date:  2021-12-31

Review 7.  Barriers and facilitators to clinical implementation of radiotherapy treatment planning automation: A survey study of medical dosimetrists.

Authors:  Rachel Petragallo; Naomi Bardach; Ezequiel Ramirez; James M Lamb
Journal:  J Appl Clin Med Phys       Date:  2022-03-03       Impact factor: 2.243

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.