The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: When are thick cement mantles detrimental?

E Takahashi1, A Kaneuji2, R Tsuda1, Y Numata1, T Ichiseki1, K Fukui1, N Kawahara1.   

Abstract

OBJECTIVES: Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep.
METHODS: We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.
RESULTS: A strong positive correlation was observed between mean cement thickness and stem subsidence in the CT slices on the balls. In the small stems, the balls moved downward to almost the same extent as the stem. There was a significant negative correlation between cement thickness and the horizontal:downward ratio of ball movement.
CONCLUSION: Collarless polished tapered stems with thicker cement mantles resulted in greater subsidence of both stem and cement. This suggests that excessive thickness of the cement mantle may interfere with effective radial cement creep.Cite this article: E. Takahashi, A. Kaneuji, R. Tsuda, Y. Numata, T. Ichiseki, K. Fukui, N. Kawahara. The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: When are thick cement mantles detrimental? Bone Joint Res 2017;6:-357. DOI: 10.1302/2046-3758.65.BJR-2017-0028.R1.
© 2017 Kaneuji et al.

Entities:  

Keywords:  Cement thickness; Polished tapered stem; Total hip arthroplasty

Year:  2017        PMID: 28566327      PMCID: PMC5457643          DOI: 10.1302/2046-3758.65.BJR-2017-0028.R1

Source DB:  PubMed          Journal:  Bone Joint Res        ISSN: 2046-3758            Impact factor:   5.853


Reliable information is currently unavailable on whether cement thickness affects radial cement creep for cemented polished tapered stems. Our study was designed to investigate the influence of cement thickness on stem subsidence and cement creep using micro-CT. In our biomechanical study, data from CT slices showed a strong positive correlation between subsidence of collarless polished tapered stems and mean cement thickness. There was a significant negative correlation between cement thickness and the horizontal:downward ratio of tantalum ball movement in the cement. Our findings suggest that excessively thick cement mantles are unlikely to provide ideal hoop stress or radial creep for stem subsidence. Strengths: We reproduced the environment within the human body in terms of moistness and temperature, and we included unloading “sleep” periods to provide stress relaxation for the cement. Precise micro-CT was used for measurements. Limitations: Cement movement in this study was limited to the tracking of tantalum ball migration within the cement, rather than the movement of the cement itself.

Introduction

Since the cement fixation system of femoral stems for total hip arthroplasty (THA) was established in the 1960s, a variety of different concepts have been applied to the development of the associated implants. Recent studies of cemented stems have reported different outcomes depending on the surface finish.[1,2] Bedard et al,[3] in a review of long-term clinical outcomes over more than 20 years suggested superior outcomes for polished surface stems. According to Lee et al,[4] the polished tapered stems slipped in the cement and created considerable compressive stress at the bone-cement interface, but with little shear stress. Kaneuji et al[5] verified this theory using a biomechanical model, and demonstrated that the polished tapered stems were associated with greater subsidence and compressive force at the bone-cement interface than was seen with rough surface stems. The authors also reported that thicker cement mantles led to greater subsidence of the stems and higher compressive force. However, excessive subsidence of the stem has been associated with stem loosening,[6] and it is still unclear whether thickening the cement mantle leads to better outcomes. Many studies have reported good outcomes for a cement mantle thickness of at least 2 mm,[7-13] but a favourable outcome with a thin cement mantle was also reported for polished French stems.[14,15] These findings are termed the “French paradox”, and the reason for the outcome remains unclear. Up to this point, no basic experiments have been performed to investigate the appropriate thickness of the cement mantle for polished stems. This study was conducted to investigate the subsidence of stem and cement, using polished tapered stems in varying thicknesses of cement mantle, and to measure three-dimensional (3D) cement creep in order to determine the optimal thickness of the cement mantle.

Materials and Methods

We used a biomechanical instrument that has already been reported by Kaneuji et al[5] for this study, and measured stem subsidence after a loading test. We implanted tantalum balls in the cement and measured their movement with a micro-CT scanner. Details of the methodology are described below.

Prosthesis

We used clinically available collarless polished tapered (CPT) stems (Zimmer Biomet, Warsaw, Indiana) with less than 0.1 µm in surface roughness. The stems were tapered from proximal to distal on both the frontal and lateral sides. This study used a total of six CPT stems: two each of small, medium and large (Fig. 1). The most proximal transverse diameter increased by 2.5 mm, and the offset diameter increased by 2 mm, between the small and medium stems and between the medium and large stems, respectively. The stems were made of cobalt-chromium (CoCr) alloy.
Fig. 1

CPT stems and No. 3 rasp. These are CPT stem sizes small, medium and large, respectively, from the left (right, the No. 3 large rasp).

CPT stems and No. 3 rasp. These are CPT stem sizes small, medium and large, respectively, from the left (right, the No. 3 large rasp).

Composite femurs

We used composite femurs #3403 (Pacific Research Laboratories, Inc., Vashon, Washington) that resembled human femurs in form and in mechanical and material characteristics.[16] The composite femurs were cut at the femoral neck. All the femurs were reamed with a large No. 3 rasp (Zimmer Biomet), and CPT stems in different sizes were inserted to provide a variety of cement thicknesses. As a marker for 3D analysis by micro-CT, four tantalum balls 0.4 mm in diameter (Bal-tec, Micro Surface Engineering, Los Angeles, California) were embedded in the outer circumference of the composite femurs 1 cm proximal to just above the lesser trochanter (Fig. 2a and Fig. 2b). Similarly, tantalum balls 0.4 mm in diameter were embedded circumferentially near the tips of the stems (Fig. 2c) that were to be inserted into the femur models. Each ball-composite femur unit was then soaked with blended vegetable oil, made of rapeseed oil and soybean oil, (The Nissin Oillio group Ltd., Tokyo, Japan) for 24 hours (Fig. 2d) to mimic the wet conditions of the in vivo femoral environment.

a) Figure showing small drilled holes for inserting tantalum balls; b) radiograph film after fixing stem with cement. Tantalum markers are visible on the outside of the composite femur (white circles) and in the cement (black circles); c) Radiograph showing placement of tantalum balls near the tips of the stems and d) Figure showing a ball-composite femur unit soaked in vegetable oil.

a) Figure showing small drilled holes for inserting tantalum balls; b) radiograph film after fixing stem with cement. Tantalum markers are visible on the outside of the composite femur (white circles) and in the cement (black circles); c) Radiograph showing placement of tantalum balls near the tips of the stems and d) Figure showing a ball-composite femur unit soaked in vegetable oil.

Loading test equipment

The composite femurs were attached to the testing instrument, which was composed of S45C structural carbon steel and epoxy resin (Devcon B; ITW Industry Co. Ltd, Osaka, Japan). The epoxy resin was shaped in the same external form as the composite femurs. Bone cement (Osteobond; Zimmer Biomet) was mixed in a vacuum cement mixing system, and 80 g of the mixed cement was inserted into the composite femur using a cement injection gun. Tantalum balls (1 mm in diameter) (Bal-tec, Micro Surface Engineering) were attached to the tips of the stems using glue (Fig. 2c). A centraliser was also attached before inserting the stem into the composite femur. A bone plug made of wax was fixed into the femoral canal. Before the cement solidified, two or three 0.6 mm tantalum balls were placed in the cement at approximately the same height as the markers embedded in the proximal position of the composite femur (Fig. 2b). The hip prosthesis model was kept at 37°C to mimic the conditions of the in vivo environment. A temperature sensor (T-35, Takigen MFG Co., Ltd, Tokyo, Japan) was attached to the epoxy resin used to fix the composite femur, so that the temperature from a heater (G6A92 240V, 250W; Takigen) was maintained at 37°C.

Methods and duration of loading

Loading was applied at an internal angle of 15° on the coronal plane to the metal femoral head (Co-Cr, 26 mm in diameter), attached to the stem neck.[17] Sine wave load (1 Hz, 3000 N) was applied 500 000 times. This was estimated to be equivalent to the load of a six-month walk.[18,19] The test load of 3000 N was considered to be equivalent to the load to the hip joint of a 70 kg person in a standing position on one leg.[18] The loading tests were conducted using a fatigue strength testing system (EHF-UM 300KN-70L; Shimadzu Corp., Kyoto, Japan). In this experiment, an eight-hour unloading period was applied between the 16-hour loading periods to mimic an actual sleeping period in clinical settings. Unloading periods were considered to be important for stress relaxation of the cement.

Parameters and methods of testing: stem subsidence

Measurement data were obtained over time using a 5 mm digital displacement gauge (DTH-A5; Kyowa Electronic Instruments Co. Ltd, Tokyo, Japan) placed at the stem shoulder (Fig. 3). Data were sent automatically to a personal computer by means of data collection and analysis software (Sensor Interface PCD-300A; Kyowa). Each set of data, measurements collected for eight minutes every hour automatically, was sorted into a file to collect 10 000 data sets, and the data were entered into the computer. A total of 216 files were extracted per loading test. The measurement data were adjusted and converted to units of distance. Each of the 16-hour loading and eight-hour unloading periods of each day were divided into the early, middle and late phases, and the first two files from each phase (a total of 20 000 data sets) were used. Subsidence was defined as the mean of the maximum values of sine waves from the 20 000 data sets of each phase of the loading period. The mean values of the 20 000 data sets of the unloading period were also collected. Perpendicular migration of the tantalum balls at the end of the stems was measured by micro-CT, and this was used as the final amount of subsidence.
Fig. 3

The testing instrument. A composite femur with stem was inserted into the testing instrument. A metal head was attached to the stem neck and was placed in contact with a loading bar (2). A digital displacement gauge (1) was placed at the stem shoulder to measure stem motion continuously during loading and unloading periods.

The testing instrument. A composite femur with stem was inserted into the testing instrument. A metal head was attached to the stem neck and was placed in contact with a loading bar (2). A digital displacement gauge (1) was placed at the stem shoulder to measure stem motion continuously during loading and unloading periods.

Parameters and methods of testing: micro-CT analysis

Before and after loading, 3D images of the areas within a 2 cm radius of the markers were obtained using a micro-CT (Microfocus X-ray CT System) scanner (TOSCANER-32250 µhd; Toshiba IT & Control Systems Corp., Tokyo, Japan) to measure the amount of migration for the tantalum balls within the cement and for the balls that were attached to the stem tip. Images were taken with a slice spacing of 95 µm. The maximum spatial resolution of the micro-CT scanner was 5 µm. In the cross-sectional CT images in which the tantalum balls were detected, the shortest distance of the line segment that intersected a ball between the bone and the stem was defined as “cement width.” These images were also used to measure the shortest distance between the centre point of the anterior, posterior, medial and lateral surfaces of the stem and the canal bone. The mean value for these four measurements was defined as the “mean cement thickness” for each stem (Fig. 4).
Fig. 4

A micro-CT slice showing a tantalum ball as a black ball in the cement. The white arrow shows cement width. The mean value of the four black arrows was defined as “mean cement thickness ”.

A micro-CT slice showing a tantalum ball as a black ball in the cement. The white arrow shows cement width. The mean value of the four black arrows was defined as “mean cement thickness ”. The distance of the migration of a tantalum ball can represent the distance of cement movement in a specific location. With this measuring method, the embedded markers in the composite femur were used as a reference point. The position of a tantalum ball before applying load was defined as the original point, and the 3D coordinate values after applying load were calculated. The horizontal distance of the outward or inward migration related to the femoral medulla was expressed as “+” and “−”, respectively. For the amount of perpendicular migration, data were obtained on the distance parallel to the axis of the bone. Movement in the direction of subsidence was expressed as “+”.

Statistical analysis

The correlation between stem subsidence and mean cement thickness was investigated using Pearson’s correlation coefficient. The relationship between the behaviour of tantalum balls and cement thickness was investigated using Spearman’s rank-correlation coefficient. A comparison of the movement of balls in two different thicknesses of cement was investigated using the Mann-Whitney U test. A significance level of 0.05 was used for these tests. Statistical analysis was performed using StatView version 5.0 (SAS Institute Inc., Cary, North Carolina).

Results

Stem subsidence

A digital displacement gauge, placed at the stem shoulder during testing, showed that all stems subsided after cyclic loading (Fig. 5). This subsidence was partially reversed during the unloading periods, but increased gradually over time. The smaller stems tended to be associated with greater subsidence; each of the small stems subsided further than either of the large stems. The mean values and standard deviations (sd) for final subsidence, as measured at the balls attached to the tips of the stems, were 0.179 mm (sd 0.084) for the large stems, 0.375 mm (sd 0.091) for the medium stems, and 0.495 mm (sd 0.126) for the small stems.
Fig. 5

Graph showing stem subsidence. All stems subsided rapidly on the first day, after which subsidence continued gradually. All stems subsided down during loading periods and rose up during unloading periods. This stem motion was observed repeatedly on a daily basis during the testing of all stems.

Graph showing stem subsidence. All stems subsided rapidly on the first day, after which subsidence continued gradually. All stems subsided down during loading periods and rose up during unloading periods. This stem motion was observed repeatedly on a daily basis during the testing of all stems.

Cement thickness

Mean cement thickness decreased as stem size increased: 1.52 mm (sd 0.45) for the large stems, 2.19 mm (sd 0.96) for the medium stems, and 2.57 mm (sd 1.61) for the small stems. A strong positive correlation was observed between mean cement thickness and final stem subsidence measured by the ball at the end of stem, with less subsidence associated with a thinner cement mantle. (Fig. 6).
Fig. 6

Graph showing the correlation between stem subsidence and cement thickness. This correlation diagram shows mean cement thickness and stem subsidence for each of six femurs. Pearson’s correlation coefficient was 0.94 (Y = 0.2469 x − 0.167, p ≤ 0.05).

Graph showing the correlation between stem subsidence and cement thickness. This correlation diagram shows mean cement thickness and stem subsidence for each of six femurs. Pearson’s correlation coefficient was 0.94 (Y = 0.2469 x − 0.167, p ≤ 0.05).

Investigation of the amount and direction of cement creep

A total of 13 tantalum balls (three for one of the two small stems, and two each for the other stems) were visualised in the micro-CT images. All of the visualised tantalum balls were found to have moved in the direction of stem subsidence. The mean distance of ball subsidence was 0.301 mm (0.015 to 0.607, sd 0.231). Table I shows measured and calculated values for different ball sizes.
Table I.

Measured and calculated values

Large 1
Large 2
Medium 1
Medium 2
Small 1
Small 2
Ball 1Ball 2Ball 1Ball 2Ball 1Ball 2Ball 1Ball 2Ball 1Ball 2Ball 1Ball 2Ball 3
Horizontal movement of ball (A) (mm)0.1610.1220.0290.0430.014–0.027–0.0330.019–0.0210.29–0.0820.059–0.075
Downward movement of ball (B) (mm)0.110.0550.0190.0150.0610.1890.4270.4220.4610.410.5710.6070.575
Stem subsidence (C) (mm)0.1190.2380.3110.4390.4060.584
A/B (%)146.4221.8152.6286.723.0–14.3–7.74.5–4.670.7–14.49.7–13.0
B/C (%)92.446.28.06.319.660.897.396.1113.5101.097.8103.998.5
Cement width (mm)2.281.331.021.521.92.663.992.281.521.525.324.181.9
Mean cement thickness (mm)1.521.521.762.612.093.04
Measured and calculated values The mean ratio of ball subsidence:stem subsidence was 102.9% (97.8% to 113.5%, sd 6.4%) for small stems, 68.4% (19.6% to 97.3%, sd 36.7%) for medium stems, and 38.2% (8.0% to 92.4%, sd 40.6%) for large stems. For small stems, the balls subsided downwards to almost the same extent as the stems. The mean horizontal movement was 0.038 mm (−0.082 to 0.290, sd 0.103). Eight tantalum balls moved outwards, and five moved inwards. Inward movement was noted in three of five tantalum balls for the small stems, two of four for the medium stems, and zero of four for the large stems. The mean horizontal:downward ratio of ball movement was 66.3% (−14.4% to 286.7%, sd 102.2%) in all balls. This ratio for the large stems was 202% (146% to 287%), indicating effective radial cement creep. The mean ratio for the small and medium stems was −1.1% (−14.4% to 23.0%), indicating that the thicker cement tended not to move in a horizontal direction by stem subsidence. Spearman’s rank-correlation coefficient showed a negative correlation between the horizontal:downward ratio of ball movement and cement width (Spearman’s rank correlation coefficient ρ = - 0.70, p < 0.05); thin cement mantle was associated with a higher percentage of outward movement. When compared to the movement of balls in two different thicknesses of cement (≥ 2.5 mm and < 2.5 mm), we found a significant difference in the horizontal:outward ratio of movement between the two groups (p = 0.03); outward movement of the balls was observed significantly more frequently for the cement width of < 2.5 mm (Fig. 7).
Fig. 7

A diagram of the horizontal:downward ratio of ball movement and cement width. This diagram demonstrated that effective horizontal cement movement in relation to cement subsidence was seen for cement widths of less than 2.5 mm (dotted line).

A diagram of the horizontal:downward ratio of ball movement and cement width. This diagram demonstrated that effective horizontal cement movement in relation to cement subsidence was seen for cement widths of less than 2.5 mm (dotted line).

Discussion

Currently, at least 2 mm of cement width is recommended for cemented stem fixation, and this recommendation is widely accepted.[7-13] A thin cement mantle is considered to be less capable of absorbing energy and to be associated with poor outcomes, with potential for cracking and breakage.[20] Localised cement defects have reportedly been associated with osteolysis and/or stem loosening, because the defect allows a passage for wear debris particles to travel from the stem-cement interface to the cement-bone interface.[21] These findings support the recommendation for a cement mantle thickness of ≥ 2 mm. However, many reports have been based on rough-surface stems, and it has been unclear whether the same findings would be observed for polished-surface stems. El Masri et al[15] reported that polished collared tapered stems, inserted using a line-to-line cementing technique to create a thin mantle, were associated with favourable long-term results. This phenomenon is called the “French paradox”. However, increased loosening has been reported for French rough-surface stems.[14] This implies that it may be important for rough-surface stems to be used with a cement layer at least 2 mm in thickness. To date, very few studies have reported on the question of whether the cement thickness for polished tapered stems should indeed be ≥ 2 mm. Scheerlinck et al[22] conducted CT analysis of femoral hip implants that had been inserted line-to-line versus one-size undersized in paired cadaver femurs, and found that the line-to-line stems resulted in better alignment. Moreover, the mean thickness of the cement mantle for the stems inserted in a line-to-line fashion was 3.08 mm, while cement defects defined as ≤ 1 mm in thickness were found in only 6.2% of the cement.[22] This suggests that actual cement defects may not be significant even if cement layers appear thin in radiographs. Janssen et al[23] found that, based on finite element analysis, stems inserted in a line-to-line fashion were associated with more favourable rotation stability and less cement damage than the one-size undersized stems. In polished stems, Lee et al[4] reported that a strong compressive force was exerted at the bone-cement interface as the polished stem slipped within the cement. The present study on collarless polished tapered stems supports the theory that effective radial cement creep (i.e. outward movement of balls) is more likely to occur if the cement mantle thickness is ≤ 2.5 mm. Our findings also suggest that, if the cement mantle were too thick no effective radial compression would occur and the cement would subside along with the stem. This leads to the conclusion that undersized cemented stems are not desirable. Physically, cement is considered to be strong in compression but weak in shear and tensile stress.[4,24-26] The use of undersized polished tapered stems could therefore be associated with cement breakage or stem loosening because shear stress is exerted on the cement as the stem subsides. Stem subsidence and cement creep may be affected by the different surface finish of stems. A recent basic experiment of similar composite femurs showed that a cement mantle thickness of ≥ 5 mm could lead to cracking at the cement-bone interface for Charnley and Lubinus SP II stems.[27] This suggests that the use of undersized stems for the purpose of attaining greater cement thickness could also increase the risk of stem failure in the non-polished stems. The present study has the following potential limitations: this is a biomechanical study, not a human body; load was applied in one direction only. Actual hip motion is not in a single constant direction; observations were limited to the tracking of tantalum ball migration in the cement rather than the actual movement of the cement itself; different alignment of the inserted stems may yield different results; no loosening of the stems was observed after testing, which could mean no difference in durability based on cement thickness; and this study used CPT stems; the use of other types of stem or cement might not yield the same results. In addition, the study did not address the behaviour of stems and cement in instances of cement defects or a very thin cement mantle. Nevertheless, the study results may provide important basic data for polished tapered stems. Our findings suggest that excessively thick cement mantles due to the use of undersized stems are unlikely to provide ideal hoop stress or radial creep associated with subsidence of the stems.
  23 in total

Review 1.  The 'French paradox.'.

Authors:  F Langlais; M Kerboull; L Sedel; R S M Ling
Journal:  J Bone Joint Surg Br       Date:  2003-01

2.  Analysis of long-term cemented total hip arthroplasty retrievals.

Authors:  William J Maloney; Thomas Schmalzried; William H Harris
Journal:  Clin Orthop Relat Res       Date:  2002-12       Impact factor: 4.176

3.  Cement microcracks in thin-mantle regions after in vitro fatigue loading.

Authors:  Kenneth A Mann; Sameer Gupta; Amos Race; Mark A Miller; Richard J Cleary; David C Ayers
Journal:  J Arthroplasty       Date:  2004-08       Impact factor: 4.757

4.  Importance of a thin cement mantle. Autopsy studies of eight hips.

Authors:  K Kawate; W J Maloney; C R Bragdon; S A Biggs; M Jasty; W H Harris
Journal:  Clin Orthop Relat Res       Date:  1998-10       Impact factor: 4.176

5.  Clinical outcomes of cemented double-tapered titanium femoral stems: a minimum 5-year follow-up.

Authors:  Haruhiko Akiyama; Keiichi Kawanabe; Koji Yamamoto; Kazutaka So; Yutaka Kuroda; Takashi Nakamura
Journal:  J Orthop Sci       Date:  2011-09-16       Impact factor: 1.601

6.  The cement mantle in total hip arthroplasty. Analysis of long-term radiographic results.

Authors:  E Ebramzadeh; A Sarmiento; H A McKellop; A Llinas; W Gogan
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

7.  The effect of femoral stem cross-sectional geometry on cement stresses in total hip reconstruction.

Authors:  R D Crowninshield; R A Brand; R C Johnston; J C Milroy
Journal:  Clin Orthop Relat Res       Date:  1980 Jan-Feb       Impact factor: 4.176

Review 8.  Systematic review of literature of cemented femoral components: what is the durability at minimum 20 years followup?

Authors:  Nicholas A Bedard; John J Callaghan; Michael D Stefl; Steve S Liu
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

9.  Assessing activity in joint replacement patients.

Authors:  C A Zahiri; T P Schmalzried; E S Szuszczewicz; H C Amstutz
Journal:  J Arthroplasty       Date:  1998-12       Impact factor: 4.757

10.  Finite element analysis of the effect of cementing concepts on implant stability and cement fatigue failure.

Authors:  Dennis Janssen; Jantien van Aken; Thierry Scheerlinck; Nico Verdonschot
Journal:  Acta Orthop       Date:  2009-06       Impact factor: 3.717

View more
  4 in total

1.  Inversion of the acetabular labrum causes increased localized contact pressure on the femoral head: a biomechanical study.

Authors:  Xipeng Wang; Kiyokazu Fukui; Ayumi Kaneuji; Kenichi Hirosaki; Hiroyasu Miyakawa; Norio Kawahara
Journal:  Int Orthop       Date:  2018-12-11       Impact factor: 3.075

2.  Biomechanical behaviour of a French femoral component with thin cement mantle: The 'French paradox' may not be a paradox after all.

Authors:  Y Numata; A Kaneuji; L Kerboull; E Takahashi; T Ichiseki; K Fukui; J Tsujioka; N Kawahara
Journal:  Bone Joint Res       Date:  2018-08-04       Impact factor: 5.853

3.  Short stems have lower load at failure than double-wedged stems in a cadaveric cementless fracture model.

Authors:  Antonio Klasan; Martin Bäumlein; Philipp Dworschak; Christopher Bliemel; Thomas Neri; Markus D Schofer; Thomas J Heyse
Journal:  Bone Joint Res       Date:  2019-11-02       Impact factor: 5.853

4.  Parathyroid hormone (1-34) can reverse the negative effect of valproic acid on the osseointegration of titanium rods in ovariectomized rats.

Authors:  Zhou-Shan Tao; Wan-Shu Zhou; Hong-Guang Xu; Min Yang
Journal:  J Orthop Translat       Date:  2020-12-28       Impact factor: 5.191

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.