Zilin Qin1,2, Jianying Zeng2, Guifeng Liu2, Xinan Long2, Gang Fang3, Zhonghai Li4, Kecheng Xu3, Lizhi Niu3.
Abstract
PURPOSE: To explore the safety and efficacy of irreversible electroporation ablation in unresectable fibrous sarcoma with 2 electrodes.Entities:
Keywords: 2-electrodes; irreversible electroporation; percutaneous
Year: 2017 PMID: 28558487 PMCID: PMC5762055 DOI: 10.1177/1533034617711530
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Preoperative contrast-enhanced computed tomography (CT) scan in a 74-year-old woman showing a 7.3 × 7.0 × 7.5 cm mass in retroperitoneal inferior vena cava near the portal triad (arrow), which was due to a malignant fibrous retroperitoneal sarcoma.
Figure 2.Computed tomography (CT) images of the irreversible electroporation (IRE) treatment being performed (in all images, the electrode distance was 2 cm): (A) the first ablation; (B) the 2 electrodes pulled back 1 cm; (C) the second target section, with the 2 electrodes moved 2 cm toward the right; (D) the electrodes pulled back 1.2 cm; (E) the third target section with the 2 electrodes moved 2 cm toward to the left and down; (F) the last target section with the 2 electrodes moved 2 cm toward the right and down; (G and H) the electrodes pulled back 2 cm in the last target section.
Physical and Pulse Parameters.a
| Pulse Application Set | Separation Between Electrodes (cm) | Voltage (V) | Repetition Rate (ms) |
|---|---|---|---|
| 1/1 | 2 | 3000 | 90 |
| 1/2 | 2 | 2600 | 90 |
| 2/1 | 2 | 2600 | 90 |
| 2/2 | 2 | 2600 | 90 |
| 3/1 | 2 | 2600 | 90 |
| 4/1 | 2 | 2600 | 90 |
| 4/2 | 2 | 2600 | 90 |
| 4/3 | 2 | 2600 | 90 |
aAll sets had pulse length of 90 μs.
Figure 3.Contrast-enhanced computed tomography (CT) images showing: (A) a hypodense ablation zone containing gas bubbles immediately postoperatively (arrow); (B) the tumor narrowed to 5.1 × 4.0 × 5.2 cm, with no obvious enhancement 2 months postoperatively.