| Literature DB >> 27602266 |
Lukas P Beyer1, Benedikt Pregler1, Christoph Nießen1, Andreas Schicho1, Michael Haimerl1, Ernst Michael Jung1, Christian Stroszczynski1, Philipp Wiggermann1.
Abstract
Purpose. The purpose of this study was to compare CT-navigated stereotactic IRE (SIRE) needle placement to non-navigated conventional IRE (CIRE) for percutaneous ablation of liver malignancies. Materials and Methods. A prospective trial including a total of 20 patients was conducted with 10 patients in each arm of the study. IRE procedures were guided using either CT fluoroscopy (CIRE) or a stereotactic planning and navigation system (SIRE). Primary endpoint was procedure time. Secondary endpoints were accuracy of needle placement, technical success rate, complication rate and dose-length product (DLP). Results. A total of 20 IRE procedures were performed to ablate hepatic malignancies (16 HCC, 4 liver metastases), 10 procedures in each arm. Mean time for placement of IRE electrodes in SIRE was significantly shorter with 27 ± 8 min compared to 87 ± 30 min for CIRE (p < 0.001). Accuracy of needle placement for SIRE was higher than CIRE (2.2 mm vs. 3.3 mm mean deviation, p < 0.001). The total DLP and the fluoroscopy DLP were significantly lower in SIRE compared to CIRE. Technical success rate and complication rates were equal in both arms. Conclusion. SIRE demonstrated a significant reduction of procedure length and higher accuracy compared to CIRE. Stereotactic navigation has the potential to reduce radiation dose for the patient and the radiologist without increasing the risk of complications or impaired technical success compared to CIRE.Entities:
Keywords: CT-guided; Interventional radiology; Irreversible electroporation; Liver tumor; Robotic assistance; Stereotactic navigation
Year: 2016 PMID: 27602266 PMCID: PMC4991851 DOI: 10.7717/peerj.2277
Source DB: PubMed Journal: PeerJ ISSN: 2167-8359 Impact factor: 2.984
Figure 1Flow of participants through each stage of the study.
Number of ablated lesions with and without stereotactic navigation.
| Conventional or stereotactic | SIRE | CIRE |
|---|---|---|
| Hepatocellular carcinoma | 10 | 6 |
| Metastasis of colorectal cancer | 0 | 4 |
| All | 10 | 10 |
Figure 2Planning the path of access on the navigation system.
A virtual entry point on the skin and tumor center is defined. Then the trajectories of the individual electrodes are calculated, but can still be adjusted individually.
Figure 3Positioning the needle guidance device for inserting the IRE electrodes.
Figure 4Evaluation of the lateral deviation of the IRE electrodes.
(A) Orthogonal plane (thickness 0.7 mm) at the tip of the reference electrode. (B) Orthogonal plane at a distance of 3 cm from the tip of the reference electrode. (C) Fusion of (A) and (B) with determination of lateral probe deviations.
Figure 5Colorectal liver metastasis in close relationship to the right portal vein.
(A) Late phase MRI shows complete ablation (B) after stereotactic placement of 3 IRE electrodes above and 2 below the portal vein (C).
Base patient and lesion aspects.
Lesion conspicuity denotes the difference in attenuation between tumor and liver parenchyma in Hounsfield units (HU − HU). Data are presented as means and standard deviations.
| Conventional or stereotactic | CIRE ( | SIRE ( |
|---|---|---|
| Sex (male) | 9 (90 %) | 9 (90 %) |
| Age—years | 63.2 ± 8.6 | 69.0 ± 11.9 |
| BMI—kg/m2 | 28.4 ± 4.1 | 29.0 ± 3.1 |
| Skin to tumor depth—mm | 79.8 ± 26.9 | 67.7 ± 25.2 |
| Tumor long axis—mm | 25.4 ± 8.1 | 25.0 ± 9.0 |
| Tumor conspicuity native—HU | 10.6 ± 7.8 | 12.8 ± 5.6 |
| Tumor conspicuity enhanced—HU | 35.6 ± 14.2 | 29.1 ± 7.6 |
Time required for sterile patient preparation; placement of IRE electrodes and total intervention.
Data are presented as means and standard deviations.
| Conventional or stereotactic | CIRE ( | SIRE ( | |
|---|---|---|---|
| Sterile patient preparation duration—min | 17.1 ± 2.8 | 16.1 ± 3.8 | |
| Electrode placement duration—min | 87.0 ± 29.9 | 26.8 ± 7.7 | |
| Placement time per electrode—min | 18.0 ± 4.2 | 5.9 ± 2.0 | |
| Procedure time until start of the ablation—min | 104.1 ± 28.2 | 55.2 ± 9.3 |
Dose length product (DLP) of the entire intervention, CT fluoroscopy and all verification scans.
Data are presented as means and standard deviations.
| Conventional or stereotactic | CIRE ( | SIRE ( | |
|---|---|---|---|
| Total DLP—mGy*cm | 4886 ± 1775 | 3510 ± 887 | |
| Verification DLP—mGy*cm | 1351 ± 479 | 725 ± 326 | |
| Fluoroscopy DLP—mGy*cm | 1705 ± 1583 | 136 ± 206 |