Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Is It Safe?

Alan R Shapiro1.   

Abstract

With the increasing accuracy of continuous glucose monitors (CGM) have come calls for the Food and Drug Administration (FDA) to label these devices as safe for nonadjunctive dosing of insulin. However, there is evidence that these devices are subject to sporadic, unpredictable, large errors. A text analysis of reports to the FDA MAUDE database since 2015 reveals over 25 000 complaints of CGM sensor inaccuracy, with instances directly leading to serious outcomes. These new data were not considered at a recent FDA Advisory Panel meeting that voted to approve Dexcom G5 relabeling for nonadjunctive use. Social media is another source of surveillance data providing evidence of large CGM inaccuracies in real-world use. We need to improve safety procedures, not remove them. CGMs offer unique information and alerts for managing diabetes, but the issue is not whether they are better than other approaches to monitoring glucose, but how they can be best used in conjunction with devices that offer the confirmatory readings needed for patient safety.

Entities:  

Keywords:  CGM accuracy; MAUDE; SMBG; continuous glucose monitor; nonadjunctive claim; text mining

Mesh:

Substances:

Year:  2017        PMID: 28540756      PMCID: PMC5588827          DOI: 10.1177/1932296816688303

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  8 in total

Review 1.  Utilizing social media data for pharmacovigilance: A review.

Authors:  Abeed Sarker; Rachel Ginn; Azadeh Nikfarjam; Karen O'Connor; Karen Smith; Swetha Jayaraman; Tejaswi Upadhaya; Graciela Gonzalez
Journal:  J Biomed Inform       Date:  2015-02-23       Impact factor: 6.317

2.  Medical device postapproval safety monitoring: where does the United States stand?

Authors:  Prashant V Rajan; Daniel B Kramer; Aaron S Kesselheim
Journal:  Circ Cardiovasc Qual Outcomes       Date:  2015-01-06

3.  Regulation Catches Up to Reality.

Authors:  Steven V Edelman
Journal:  J Diabetes Sci Technol       Date:  2016-09-25

4.  Electrochemical Glucose Sensing: Is There Still Room for Improvement?

Authors:  Emilia Witkowska Nery; Magdalena Kundys; Paulina S Jeleń; Martin Jönsson-Niedziółka
Journal:  Anal Chem       Date:  2016-11-04       Impact factor: 6.986

5.  Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm.

Authors:  Timothy S Bailey; Anna Chang; Mark Christiansen
Journal:  J Diabetes Sci Technol       Date:  2014-11-03

6.  Redundancy in Glucose Sensing: Enhanced Accuracy and Reliability of an Electrochemical Redundant Sensor for Continuous Glucose Monitoring.

Authors:  Amin Sharifi; Andrea Varsavsky; Johanna Ulloa; Jodie C Horsburgh; Sybil A McAuley; Balasubramanian Krishnamurthy; Alicia J Jenkins; Peter G Colman; Glenn M Ward; Richard J MacIsaac; Rajiv Shah; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-05-03

Review 7.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

8.  The accuracy benefit of multiple amperometric glucose sensors in people with type 1 diabetes.

Authors:  Jessica R Castle; Amy Pitts; Kathryn Hanavan; Rhonda Muhly; Joseph El Youssef; Colleen Hughes-Karvetski; Boris Kovatchev; W Kenneth Ward
Journal:  Diabetes Care       Date:  2012-02-22       Impact factor: 19.112

  8 in total
  5 in total

1.  The Safety of Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Still Not Resolved.

Authors:  Alan R Shapiro
Journal:  J Diabetes Sci Technol       Date:  2017-04-18

2.  Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison.

Authors:  Guido Freckmann; Manuela Link; Ulrike Kamecke; Cornelia Haug; Bernhard Baumgartner; Raimund Weitgasser
Journal:  J Diabetes Sci Technol       Date:  2019-02-07

3.  Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System.

Authors:  Manuela Link; Ulrike Kamecke; Delia Waldenmaier; Stefan Pleus; Arturo Garcia; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2019-12-17

4.  Commentary Regarding Shapiro, "Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Is It Safe?"

Authors:  David Price
Journal:  J Diabetes Sci Technol       Date:  2017-03-01

5.  Differences in Glucose Readings Between Right Arm and Left Arm Using a Continuous Glucose Monitor.

Authors:  Sonoko Kawakatsu; Xiaohan Liu; Brandon Tran; Brittany P Tran; Lucy Manzanero; Eric Shih; Allen Shek; Jeremy J Lim
Journal:  J Diabetes Sci Technol       Date:  2021-05-06
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.