| Literature DB >> 28515409 |
Kosuke Morita1, Seitaro Oda2, Daisuke Utsunomiya2, Takeshi Nakaura2, Takatoshi Matsubara1, Makoto Goto1, Tomoyuki Okuaki3, Hideaki Yuki2, Yasunori Nagayama2, Masafumi Kidoh2, Kenichiro Hirata2, Yuij Iyama2, Narumi Taguchi2, Masahiro Hatemura1, Masahiro Hashida1, Yasuyuki Yamashita2.
Abstract
PURPOSE: To evaluate the effect of a composite radiofrequency (RF) pulse on saturation recovery (SR) myocardial T1 mapping using a 3T MR system.Entities:
Keywords: 3T magnetic resonance; composite radio-frequency pulse; myocardial T1 mapping; reproducibility; saturation recovery
Mesh:
Year: 2017 PMID: 28515409 PMCID: PMC5760231 DOI: 10.2463/mrms.mp.2016-0092
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471
Fig. 1.Saturation recovery T1 mapping sequence with conventional and composite radiofrequency (RF) pulses. Short and long saturation time delay images using a 2D turbo field echo readout. A composite RF pulse applied a four-pulse train to saturate magnetization uniformly and yielded more accurate and reproducible T1 measurements on a high-field 3T MRI system. TD, time delay; TFE, turbo field echo.
Fig. 2.Pulse sequence diagrams for the saturation recovery T1 mapping sequence with composite radiofrequency (RF) pulse. The composite RF pulse consists of specified non-selective four hard pulses. The angles of these four pulses, α1, α2, α3, and α4 are used 72, 92, 126, and 193 degrees, respectively.
Fig. 3.For the in vivo study, we manually contoured the myocardium in each segment (anterior, septal, lateral, and inferior segments) on the mid-ventricular short-axis T1 map image.
The mean and standard deviation (SD) of the measured T1 values for conventional and composite radiofrequency (RF) pulses in the phantom study
| 1 | 290 | 320.6 | 325.2 | 0.27 | 12.2 | 7.3 | 0.17 |
| 2 | 570 | 561.3 | 579.3 | <0.01 | 11.6 | 5.3 | 0.04 |
| 3 | 630 | 623.9 | 645.1 | <0.01 | 16.8 | 6.7 | 0.02 |
| 4 | 810 | 768.9 | 863.4 | 0.03 | 99.7 | 9.7 | <0.01 |
| 5 | 910 | 840.3 | 923 | 0.03 | 86.6 | 9.2 | <0.01 |
| 6 | 1180 | 1095.2 | 1211 | 0.04 | 140.9 | 9.6 | <0.01 |
| 7 | 1333 | 1242.6 | 1396.2 | 0.04 | 179.9 | 4.4 | <0.01 |
| 8 | 1797 | 1543 | 1730.8 | <0.01 | 164.4 | 1.2 | <0.01 |
Fig. 4.Box plot showing the mean T1 values of eight cylindrical phantoms (reference T1 value = 290–1797 ms). There were significant differences in the mean T1 values of the vials except for vial no. 1 (reference T1 value = 290 ms). standard deviation (SD) of the measured T1 values for the composite radiofrequency (RF) pulse was significantly smaller than that for conventional RF pulse except for vial no. 1. SD of the measured T1 values for composite RF pulse was >10 ms.
Fig. 5.Box plot showing the inter-segment variability for the composite radiofrequency (RF) pulse of the in vivo study. The inter-segment variability for the composite radiofrequency (RF) pulse was significantly smaller than that for the conventional RF pulse (44.5 ± 21.4 vs. 72.8 ± 29.2 ms, P < 0.01).
Fig. 6.Bland–Altman analysis of the T1 measurements for the composite radiofrequency (RF) pulse methods. The lower Bland–Altman limit of agreement was with the composite RF pulse method. SD, standard deviation.