Literature DB >> 28503349

Medial Malleolar Fractures: An Anatomic Survey Determining the Ideal Screw Length.

P J Labronici1,2, R E Pires3, M V Franco2, R Freitas2, G C Araújo1, R S Pires E Albuquerque1, V S Gameiro1, K Jeray4.   

Abstract

BACKGROUND: Medial malleolar fractures are frequent, and their treatment is familiar to the orthopedic surgeon. Lag screw fixation using partially threaded screws remains the standard treatment method for medial malleolar fractures. However, the literature lacks a defined method for selecting lag screw length, relying more so on the empiric choice of the surgeon. AIM: The aim of this study is to help define the ideal lag screw length for medial melleolar fracture fixation.
MATERIALS AND METHODS: One hundred and sixteen anatomic specimens were included in the study. A transverse cut was performed in the distal third of the tibia, roughly 1 and a half times the distal tibial plafond width from the ankle joint. A coronal cut was then performed using the center of the medial malleolus. Three observers measured the distance between the medial malleolus tip and beginning of the medullary canal in all anatomic specimens. Differences in measurements were statistically compared, level of (P ≤ 0.05). Interclass correlation coefficient (ICC) significance level was set at P < 0.05.
RESULTS: Measurement average was 55 mm between the medial malleolus tip and the medullary canal, with a standard deviation of 10 mm. High concordance (ICC: 0.819) was achieved among all pairs of observers (P < 0.01). The systematic difference among measurements was absent, and random distribution around general measurements was observed.
CONCLUSION: The authors recommend a screw length of no more than 45 mm to optimize the location of the screw threads in the best cancellous bone in an effort to obtain the most compression.

Entities:  

Keywords:  Ankle; ankle fractures; medial malleolus

Year:  2016        PMID: 28503349      PMCID: PMC5414444          DOI: 10.4103/amhsr.amhsr_41_16

Source DB:  PubMed          Journal:  Ann Med Health Sci Res        ISSN: 2141-9248


Introduction

Ankle fractures are one of the most common injuries treated by the orthopedic surgeon.[1] Although nondisplaced medial malleolar fractures are suitable for conservative treatment, in the vast majority of cases the fractures are displaced and require surgical treatment. Several treatment options are available for medial malleolar fracture fixation, including lag screws, tension band, small precontoured plates, and K-wires. However, partially threaded screws remain the most common method for fracture fixation.[2345] However, the literature lacks a defined method for selecting lag screw length, relying more so on the empiric choice of the surgeon. The present study aims to determine the ideal screw length for medial malleolar fractures based on an anatomic study of the distal tibia metaphyseal bone.

Materials and Methods

Settings and design

One hundred and sixteen skeletally mature anatomic specimens of the Latin American population were used in the study. Determining the age and gender was impossible due to only the limbs were available for measurements. A transverse cut was performed in the distal third of the tibia, roughly one and a half times the distal tibial plafond width from the ankle joint. Another coronal cut was then performed using the center of the medial malleolus. Using a caliper, three observers measured the shortest distance between the medial malleolus tip and the medullary canal in all anatomic specimens. Two measurements of the same location were performed by each observer to improve measurement accuracy [Figure 1].
Figure 1

Coronal cut. Red line: Distance between the medial malleolus tip and the medullary canal

Coronal cut. Red line: Distance between the medial malleolus tip and the medullary canal

Statistical analysis

According to Bland-Altman recommendation of 100 as an adequate sample size for a study of agreement among measurements (95% confidence interval [CI]), 116 anatomic specimens were obtained to ensure that this study was well powered (95% CI).[6] The interclass correlation coefficient (ICC) was applied according to Landis and Koch to assess interobserver reproducibility.[7] Bland-Altman plots were used to analyze measurement dispersion by each observer and verify if bias or another systematic discrepancy existed among these measurements. SPSS software (version 20.0, Chicago IL, USA) was used for statistical analysis, and significance level was 5% (P ≤ 0.05). Ethical approval was granted by the Local Ethics Committee.

Results

The measurements are described in their central dispersion tendency in Table 1. The narrow CIs and variations suggest high measurement precision [Table 1].
Table 1

Observers measurements

Observers measurements Observer 1 presented the highest agreement level (ICC = 0.979), almost perfect. His precision was followed by observers 2 (ICC = 0.962) and 3 (ICC = 0.936), expressing high measurement reproducibility (P < 0.01). Table 2 provides information about means and agreement limits according to Bland-Altman [Table 2].
Table 2

Interobserver reproducibility

Interobserver reproducibility Agreement limits for observers 1 and 2 were narrow (maximum 1 mm), showing high reproducibility. Observer 3 presented slightly wider limits, reaching 1.3 mm. The absence of systematic differences between repeated measurements from the same observer was found. This study showed an average distance of 55 mm between the medial malleolus tip and the start of the medullary canal in 116 skeletally mature anatomic specimens, with a 10 mm standard deviation (SD).

Discussion

Displaced medial malleolar fractures usually require surgical treatment, and partially threaded screws are the most commonly employed surgical option for fracture fixation.[8910111213] However, the literature lacks criteria for the adequate choice of lag screw length. Excessively long partially threaded screws can bypass the good cancellous bone in the metaphyseal area and lead to inefficient compression due to poor screw thread purchase inside the medullary canal. An excessively short screw can endanger fracture compression due to screw threads not completely bypassing the fracture site. The present study showed an average distance of 55 mm (SD: 10 mm) between the medial malleolus tip and the start of the medullary canal in 116 Latin American cadavers, with a 10 mm SD. Therefore, when using a typical partially threaded screw (roughly 16 mm of thread length), one can conclude that the ideal screw length is roughly 45 mm for placement of the screw threads in the optimal cancellous bone to generate effective fracture site compression. Parker et al.,[3] in a biomechanical study of human cadavers, compared the compression at the fracture site treated with either 30 or 45 mm long partially threaded screws or 45 mm fully threaded screws. The authors hypothesized that the partially threaded screws provide inadequate compression as a result of bypassing the physeal scar, and proposed that better purchase may be achieved with fully threaded screws or shorter partially threaded screws. The median compression at the fracture site achieved with 45 mm fully threaded screws and 30 mm partially threaded screws were significantly higher than that achieved with 45 mm partially threaded screws. Therefore, the authors recommend the use of 45 mm fully threaded or 30 mm partially threaded screws when fixing fractures of the medial malleolus. This information strongly confirms the relevance of the present study. The current literature lacks consensus concerning the precise screw length to provide adequate fracture compression in medial malleolar fractures. Our study demonstrated that screws longer that 55 mm (SD: 10 mm) will probably bypass the metaphyseal area, thereby leading to poor fracture compression. Weaknesses of the present study include the lack of specimen gender, age, and most importantly bone quality. Although this information could potentially change the measurements (for example with severe osteopenia the canal may be wider and the distance from the tip of the medial mallelous shorter), the authors believe the sample size helps minimize this weakness.

Conclusion

The authors recommend a screw length of no more than 45 mm to optimize the location of the screw threads in the best cancellous bone in an effort to obtain the most compression. The authors believe stable and effective fracture fixation is achieved when using the appropriate screw length, thereby improving clinical outcomes.

Compliance with ethical standards

Authors declare that they have no conflict of interest related to the present study. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. This study was performed in accordance with the ethical standards of the Institutional Research Committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  10 in total

1.  Force measurements in screw fixation.

Authors:  S M Perren
Journal:  J Biomech       Date:  1976       Impact factor: 2.712

Review 2.  Characteristics and operative management of supination external rotation ankle fractures.

Authors:  R P Reinherz; S R Granoff; K E Henning; B S Ross
Journal:  J Foot Surg       Date:  1991 Jul-Aug

3.  Lag screw fixation of medial malleolar fractures: a biomechanical, radiographic, and clinical comparison of unicortical partially threaded lag screws and bicortical fully threaded lag screws.

Authors:  William M Ricci; Paul Tornetta; Joseph Borrelli
Journal:  J Orthop Trauma       Date:  2012-10       Impact factor: 2.512

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The measurement of observer agreement for categorical data.

Authors:  J R Landis; G G Koch
Journal:  Biometrics       Date:  1977-03       Impact factor: 2.571

6.  A clinical evaluation of alternative fixation techniques for medial malleolus fractures.

Authors:  Hayley Barnes; Lisa K Cannada; J Tracy Watson
Journal:  Injury       Date:  2014-07-03       Impact factor: 2.586

7.  Results of operative fixation of unstable ankle fractures in geriatric patients.

Authors:  A J Pagliaro; J D Michelson; M S Mizel
Journal:  Foot Ankle Int       Date:  2001-05       Impact factor: 2.827

8.  Screw fixation of medial malleolar fractures: a cadaveric biomechanical study challenging the current AO philosophy.

Authors:  L Parker; N Garlick; I McCarthy; S Grechenig; W Grechenig; P Smitham
Journal:  Bone Joint J       Date:  2013-12       Impact factor: 5.082

9.  Preventing surgical complications: A survey on surgeons' perception of intra-articular malleolar screw misplacement in a cadaveric study.

Authors:  Vincenzo Giordano; Arthur Fs Gomes; Ney P Amaral; Rodrigo P Albuquerque; Robinson Es Pires
Journal:  Patient Saf Surg       Date:  2011-10-04

10.  Ottawa ankle rules and subjective surgeon perception to evaluate radiograph necessity following foot and ankle sprain.

Authors:  Res Pires; Aa Pereira; Gm Abreu-E-Silva; Pj Labronici; Lb Figueiredo; Al Godoy-Santos; M Kfuri
Journal:  Ann Med Health Sci Res       Date:  2014-05
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.