A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry.

Paul R Roberts1, Nicolas Clementy2, Faisal Al Samadi3, Christophe Garweg4, Jose Luis Martinez-Sande5, Saverio Iacopino6, Jens Brock Johansen7, Xavier Vinolas Prat8, Robert C Kowal9, Didier Klug10, Lluis Mont11, Jan Steffel12, Shelby Li13, Dirk Van Osch14, Mikhael F El-Chami15.   

Abstract

BACKGROUND: First-in-man studies of leadless pacemakers have demonstrated high rates of implant success, and safety and efficacy objectives were achieved. Outside of the investigational setting, there are concerns, particularly over cardiac effusion and perforation, device dislodgement, infection, telemetry, and battery issues.
OBJECTIVE: The acute performance of the Micra transcatheter pacemaker from a worldwide Post-Approval Registry is reported.
METHODS: The registry is an ongoing prospective single-arm observational study designed to assess the safety and effectiveness of Micra in the post-approval setting. The safety end point was system- or procedure-related major complications at 30 days post implant. We compared the major complication rate with that of the 726 patients from the investigational study. Electrical performance was also characterized.
RESULTS: The device was successfully implanted in 792 of 795 registry patients (99.6%) by 149 implanters at 96 centers in 20 countries. Through 30 days post implant, a total of 13 major complications occurred in 12 patients, for a major complication rate of 1.51% (95% confidence interval, 0.78%-2.62%). Major complications included cardiac effusion/perforation (1, 0.13%), device dislodgement (1, 0.13%), and sepsis (1, 0.13%). After adjusting for baseline differences, the rate of major complications in the registry trended lower than the investigational trial (odds ratio, 0.58, 95% confidence interval, 0.27-1.25; P = .16). Early pacing capture thresholds were low and stable.
CONCLUSION: Performance of the Micra transcatheter pacemaker in a real-world setting demonstrates a high rate (99.6%) of implant success and low rate (1.51%) of major complications through 30 days post implant. In particular, the rates of pericardial effusion, device dislodgement, and infection were low, reinforcing the positive results seen in the investigational study.
Copyright © 2017 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Acute performance; Leadless pacemaker; Real-world performance

Mesh:

Year:  2017        PMID: 28502871     DOI: 10.1016/j.hrthm.2017.05.017

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  48 in total

Review 1.  Transvenous Lead Extractions: Current Approaches and Future Trends.

Authors:  Adryan A Perez; Frank W Woo; Darren C Tsang; Roger G Carrillo
Journal:  Arrhythm Electrophysiol Rev       Date:  2018-08

Review 2.  [Micra™ leadless pacemaker : Clinical experience and perspectives].

Authors:  Clemens Steinwender; Hermann Blessberger; Daniel Kiblböck; Karim Saleh; Jürgen Kammler
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-11-19

3.  Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: is this the direction to go?

Authors:  Tomas Holubec; Julius Beckers; Zdenka Holubcova; Veronika Walter; Thomas Walther
Journal:  Cardiovasc Diagn Ther       Date:  2018-08

Review 4.  Leadless pacing.

Authors:  J Sperzel; C Hamm; A Hain
Journal:  Herz       Date:  2018-11       Impact factor: 1.443

Review 5.  Leadless Pacemakers: Recent and Future Developments.

Authors:  Anne Kroman; Basil Saour; Jordan M Prutkin
Journal:  Curr Treat Options Cardiovasc Med       Date:  2019-09-05

Review 6.  Keeping up appearances: the radiographic evolution of cardiovascular implantable electronic devices.

Authors:  Benedict M Wiles; Caroline A Illingworth; John R Paisey; Paul R Roberts; Stephen P Harden
Journal:  Br J Radiol       Date:  2017-10-26       Impact factor: 3.039

7.  Successful Denovo Implantation And Explanation Of An Old Malfunctioning Micratm Leadless Pacemaker.

Authors:  Muhammad R Afzal; Tawseef Dar; Mahmoud Houmsse; Ralph Augostini; Emile G Daoud; John Hummel
Journal:  J Atr Fibrillation       Date:  2017-12-31

8.  Comparison of the safety and efficacy of Nanostim and Micra transcatheter leadless pacemaker (LP) extractions: a multicenter experience.

Authors:  Tawseef Dar; Krishna Akella; Ghulam Murtaza; Sharan Sharma; Muhammad R Afzal; Rakesh Gopinathannair; Ralph Augostini; John Hummel; Dhanunjaya Lakkireddy
Journal:  J Interv Card Electrophysiol       Date:  2020-01-06       Impact factor: 1.900

9.  Implantation of the Micra transcatheter pacing system: Single Polish center experience with the real costs of hospitalization analysis.

Authors:  Marcin Grabowski; Marcin Michalak; Monika Gawałko; Sylwia Gajda; Andrzej Cacko; Łukasz Januszkiewicz; Agnieszka Kołodzińska; Przmysław P Mitkowski; Gabor Z Duray; Grzegorz Opolski
Journal:  Cardiol J       Date:  2018-08-29       Impact factor: 2.737

10.  VVI pacing with normal QRS duration and ventricular function: MOST trial findings relevant to leadless pacemakers.

Authors:  Zak Loring; Rebecca North; Anne S Hellkamp; Brett D Atwater; Camille G Frazier-Mills; Kevin P Jackson; Sean D Pokorney; Gervasio A Lamas; Jonathan P Piccini
Journal:  Pacing Clin Electrophysiol       Date:  2020-11-05       Impact factor: 1.976

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.