| Literature DB >> 28491598 |
Christian Steinberg1, Santabhanu Chakrabarti2, Andrew D Krahn2, Jamil Bashir2.
Abstract
Entities:
Keywords: DFT, defibrillation threshold; Device interaction; Epicardial pacemaker; ICD; ICD, implantable cardioverter-defibrillator; Limited venous access; MRSA, methicillin-resistant Staphylococcus aureus; Recurrent device infection; S-ICD; S-ICD, subcutaneous ICD; Sensing screening; VF, ventricular fibrillation; bpm, beats per minute
Year: 2015 PMID: 28491598 PMCID: PMC5419715 DOI: 10.1016/j.hrcr.2015.04.001
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Chest radiographs before and after implantation of subcutaneous implantable cardioverter-defibrillator (S-ICD) and permanent epicardial pacemaker. A: Lateral chest radiograph of the transvenous dual-chamber ICD system prior to extraction. B: Posterior-anterior chest radiograph post implantation of S-ICD and permanent epicardial pacemaker. The 2 epicardial atrial electrodes are fixed to the free wall of the right atrium (bold white arrows). The 2 ventricular electrodes of the epicardial pacemaker are fixed to the inferior and anterior surface of the right ventricle, respectively (bold black arrows). The can of the S-ICD is positioned in a subcutaneous pocket at the lower left-lateral thorax (dashed white arrow). The S-ICD lead is placed in a right parasternal position (black arrowheads). C: Shown are the 3 bipolar sensing vectors of an S-ICD. The primary vector senses between the proximal lead electrode and the can. The secondary vector senses between the distal lead electrode and the can. The alternate vector senses between the 2 lead electrodes.
Figure 2Sensing findings at implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD). A: Higher atrial pacing output (3.5 V/0.4 ms) by the epicardial pacemaker resulted in detection of atrial pacing and double counting through the S-ICD in the secondary sensing vector, but not in the primary sensing vector (C). B: Lowering of the atrial pacing output to 1.5 V/0.4 ms eliminated double counting through the S-ICD in the secondary sensing vector.
Figure 3Defibrillation threshold testing. A: Ventricular fibrillation (VF) was induced with a 50 Hz train through the subcutaneous implantable cardioverter-defibrillator (S-ICD). The VF induction was not detected by the epicardial dual-chamber pacemaker and continued pacing is recorded during the VF induction. B: Subsequently appropriate VF detection by the epicardial pacemaker. Despite several attempts, the inducible VF was self-terminating before shock delivery through the S-ICD. RA = right atrial.
KEY TEACHING POINTS
Combination of epicardial pacemaker system and subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative and safe device option in patients with limited or absent venous access. Complex interaction between both devices exhibits potential for double counting and inappropriate shocks through S-ICD and requires a very careful sensing screening and programming. Close follow-up and frequent check for oversensing through the S-ICD is mandatory. |