Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator.

Raul Weiss1, Bradley P Knight, Michael R Gold, Angel R Leon, John M Herre, Margaret Hood, Mayer Rashtian, Mark Kremers, Ian Crozier, Kerry L Lee, Warren Smith, Martin C Burke.   

Abstract

BACKGROUND: The most frequent complications associated with implantable cardioverter-defibrillators (ICDs) involve the transvenous leads. A subcutaneous implantable cardioverter-defibrillator (S-ICD) has been developed as an alternative system. This study evaluated the safety and effectiveness of the S-ICD System (Cameron Health/Boston Scientific) for the treatment of life-threatening ventricular arrhythmias (ventricular tachycardia/ventricular fibrillation). METHODS AND
RESULTS: This prospective, nonrandomized, multicenter trial included adult patients with a standard indication for an ICD, who neither required pacing nor had documented pace-terminable ventricular tachycardia. The primary safety end point was the 180-day S-ICD System complication-free rate compared with a prespecified performance goal of 79%. The primary effectiveness end point was the induced ventricular fibrillation conversion rate compared with a prespecified performance goal of 88%, with success defined as 2 consecutive ventricular fibrillation conversions of 4 attempts. Detection and conversion of spontaneous episodes were also evaluated. Device implantation was attempted in 321 of 330 enrolled patients, and 314 patients underwent successful implantation. The cohort was followed for a mean duration of 11 months. The study population was 74% male with a mean age of 52±16 years and mean left ventricular ejection fraction of 36±16%. A previous transvenous ICD had been implanted in 13%. Both primary end points were met: The 180-day system complication-free rate was 99%, and sensitivity analysis of the acute ventricular fibrillation conversion rate was >90% in the entire cohort. There were 38 discrete spontaneous episodes of ventricular tachycardia/ventricular fibrillation recorded in 21 patients (6.7%), all of which successfully converted. Forty-one patients (13.1%) received an inappropriate shock.
CONCLUSIONS: The findings support the efficacy and safety of the S-ICD System for the treatment of life-threatening ventricular arrhythmias.

Entities:  

Keywords:  defibrillators, implantable; heart arrest; tachycardia

Mesh:

Year:  2013        PMID: 23979626     DOI: 10.1161/CIRCULATIONAHA.113.003042

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  108 in total

Review 1.  Initial experience of subcutaneous implantable cardioverter defibrillators in Singapore: a case series and review of the literature.

Authors:  Tien Siang Eric Lim; Boon Yew Tan; Kah Leng Ho; Chuh Yih Paul Lim; Wee Siong Teo; Chi-Keong Ching
Journal:  Singapore Med J       Date:  2015-10       Impact factor: 1.858

2.  Limitations in S-ICD therapy: reasons for system explantation.

Authors:  Gerrit Frommeyer; Dirk G Dechering; Sven Zumhagen; Julia Köbe; Lars Eckardt; Florian Reinke
Journal:  Clin Res Cardiol       Date:  2015-07-07       Impact factor: 5.460

Review 3.  Current Review of Implantable Cardioverter Defibrillator Use in Patients With Left Ventricular Assist Device.

Authors:  Jacinthe Boulet; Emmanuelle Massie; Blandine Mondésert; Yoan Lamarche; Michel Carrier; Anique Ducharme
Journal:  Curr Heart Fail Rep       Date:  2019-12

Review 4.  Alzheimer's disease: the impact of age-related changes in reproductive hormones.

Authors:  C S Atwood
Journal:  Cell Mol Life Sci       Date:  2005-02       Impact factor: 9.261

Review 5.  Arrhythmias in structural heart disease.

Authors:  H Sawyer Gillespie; Charles C H Lin; Jordan M Prutkin
Journal:  Curr Cardiol Rep       Date:  2014-08       Impact factor: 2.931

6.  Subcutaneous implantable cardioverter-defibrillators: expanding the scope of technology.

Authors:  Mohammad Saeed
Journal:  Tex Heart Inst J       Date:  2015-04-01

Review 7.  Tricuspid valve incompetence following implantation of ventricular leads.

Authors:  Giselle A Baquero; Jerry Luck; Gerald V Naccarelli; Mario D Gonzalez; Javier E Banchs
Journal:  Curr Heart Fail Rep       Date:  2015-04

Review 8.  Subcutaneous implantable cardiac defibrillators: indications and limitations.

Authors:  Maria Grazia Bongiorni; Stefano Viani; Giulio Zucchelli; Andrea Di Cori; Luca Segreti; Luca Paperini; Dianora Levorato; Adriano Boem; Giulia Branchitta; Diana Andreini; Ezio Soldati
Journal:  Curr Heart Fail Rep       Date:  2015-02

Review 9.  The subcutaneous ICD-current evidence and challenges.

Authors:  Kiran Haresh Kumar Patel; Pier D Lambiase
Journal:  Cardiovasc Diagn Ther       Date:  2014-12

Review 10.  Comparing the safety of subcutaneous versus transvenous ICDs: a meta-analysis.

Authors:  Li Su; Jia Guo; Yingqun Hao; Hong Tan
Journal:  J Interv Card Electrophysiol       Date:  2021-01-11       Impact factor: 1.900

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.