| Literature DB >> 28462690 |
Jun Li1, Wenyin Shi1, David Andrews2, Maria Werner-Wasik1, Bo Lu1, Yan Yu1, Adam Dicker1, Haisong Liu1.
Abstract
PURPOSE: The study was aimed to compare online 6 degree-of-freedom image registrations of TrueBeam cone-beam computed tomography and BrainLab ExacTrac X-ray imaging systems for intracranial radiosurgery.Entities:
Keywords: 6 degrees of freedom; ExacTrac X-ray imaging; TrueBeam; cone-beam computed tomography; image registration; intracranial; radiosurgery
Mesh:
Year: 2016 PMID: 28462690 PMCID: PMC5616049 DOI: 10.1177/1533034616683069
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Picture of the TrueBeam STx linear accelerator, which is equipped with CBCT and BrainLab X-ray imaging systems. Rando head phantom was immobilized with a BrainLab mask on the treatment couch. The coordinate system used in the study is shown. CBCT indicates cone-beam computed tomography.
Isocenter Locations in the Head Phantom.
| Isocenters | Isocenter Location |
|---|---|
| 1 | Brain stem |
| 2 | Left cerebellopontine angle |
| 3 | Right cerebellopontine angle |
| 4 | Left cerebellum |
| 5 | Right cerebellum |
| 6 | Left temporal lobe |
| 7 | Right temporal lobe |
| 8 | Left frontal lobe |
| 9 | Right frontal lobe |
| 10 | Thalamus |
Figure 2.Results of the phantom study of the 10 isocenters: absolute differences in calculated residual errors between ExacTrac X-ray imaging registration and TrueBeam CBCT imaging registration (difference = ExacTrac − CBCT) in (A) translational (vertical, longitudinal, and lateral) and (B) rotational (rotation, roll, and pitch) directions, respectively. CBCT indicates cone-beam computed tomography.
Results of the Phantom Study.a
| Difference | Vertical (mm) | Longitudinal (mm) | Lateral (mm) | Rotation (°) | Roll (°) | Pitch (°) |
|---|---|---|---|---|---|---|
| Minimum | 0.04 | 0.10 | 0.02 | 0.23 | 0.00 | 0.00 |
| Maximum | 0.32 | 0.64 | 0.36 | 0.46 | 0.34 | 0.32 |
| Mean | 0.17 | 0.36 | 0.25 | 0.34 | 0.13 | 0.12 |
| Standard deviation | 0.11 | 0.20 | 0.11 | 0.08 | 0.09 | 0.10 |
Abbreviation: CBCT, cone-beam computed tomography.
aResidual error differences (absolute values) in translational and rotational directions, between ExacTrac X-ray imaging registration and TrueBeam CBCT registration among 10 isocenter studies.
Figure 3.Results of the patient study: absolute differences in calculated residual errors between ExacTrac X-ray imaging registration and TrueBeam CBCT imaging registration (difference = ExacTrac − CBCT) in (A) translational (vertical, longitudinal, and lateral) and (B) rotational (rotation, roll, and pitch) directions, respectively. CBCT indicates cone-beam computed tomography.
Results of the Patient Study.a
| Difference | Vertical (mm) | Longitudinal (mm) | Lateral (mm) | Rotation (°) | Roll (°) | Pitch (°) |
|---|---|---|---|---|---|---|
| Minimum | 0.02 | 0.03 | 0.01 | 0.15 | 0 | 0 |
| Maximum | 0.64 | 0.79 | 0.62 | 0.72 | 0.48 | 0.47 |
| Mean | 0.20 | 0.30 | 0.21 | 0.40 | 0.17 | 0.20 |
| Standard deviation | 0.16 | 0.18 | 0.18 | 0.16 | 0.13 | 0.14 |
Abbreviation: CBCT, cone-beam computed tomography.
aResidual error differences (absolute values) in translational and rotational directions, between ExacTrac X-ray imaging registration and TrueBeam CBCT registration among 34 patient studies.