Literature DB >> 28402125

Assessment of radiation dose from abdominal quantitative CT with short scan length.

Xinhua Li1, Kai Yang1, Matthew C DeLorenzo1, Bob Liu1.   

Abstract

OBJECTIVE: To assess radiation dose for patients who received abdominal quantitative CT and to compare the midpoint dose [DL(0)] at the centre of a 1-cm scan length with the volume CT dose index (CTDIvol). Although the size-specific dose estimate (SSDE) proposed in The American Association of Physicists in Medicine Report No. 204 is not applicable for short-length scans, commercial dose-monitoring software, such as Radimetrics™ Enterprise Platform (Bayer HealthCare, Whippany, NJ), reports SSDE for all scans. SSDE was herein compared with DL(0).
METHODS: Data were analyzed from 398 abdominal quantitative CT examinations in 165 males and 233 females. The CTDIvol was 4.66 mGy, and the scan length was 1 cm for all examinations. Radimetrics was used to extract patient diameter and SSDE. DL(0) was assessed using a previously reported method that takes into account both patient size and scan length.
RESULTS: The mean patient diameter was 28.5 ± 6.3 cm (range, 16.5-46.6 cm); the mean SSDE was 6.22 ± 1.36 mGy (range, 3.12-9.42 mGy); and the mean DL(0) was 2.97 ± 0.95 mGy (range, 1.18-5.77 mGy). As patient diameter increased, the DL(0) to CTDIvol ratio decreased, ranging from 1.24 to 0.25; the DL(0) to SSDE ratio also decreased, ranging from 0.61 to 0.38.
CONCLUSION: The dose to the patients from abdominal quantitative CT may be largely different from CTDIvol and SSDE. This study demonstrates the necessity of taking into account not only patient size but also scan length for evaluating the dose from short-length scans. Advances in knowledge: In CT examinations with 1-cm scan length, dose evaluation needs to take into account both patient size and scan length. An omission of either factor can result in an erroneous result.

Entities:  

Mesh:

Year:  2017        PMID: 28402125      PMCID: PMC5602182          DOI: 10.1259/bjr.20160931

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  10 in total

1.  A comprehensive approach to CT radiation dose reduction: one institution's experience.

Authors:  Elias J Antypas; Farnoosh Sokhandon; Michael Farah; Scott Emerson; Kostaki G Bis; Hai Tien; Duane Mezwa
Journal:  AJR Am J Roentgenol       Date:  2011-10       Impact factor: 3.959

2.  Updated estimates of typical effective doses for common CT examinations in the UK following the 2011 national review.

Authors:  Paul C Shrimpton; Jan T M Jansen; John D Harrison
Journal:  Br J Radiol       Date:  2015-11-06       Impact factor: 3.039

3.  Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI100.

Authors:  Robert L Dixon; Adam C Ballard
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

4.  CT Radiation Dose Optimization and Tracking Program at a Large Quaternary-Care Health Care System.

Authors:  Ajit H Goenka; Frank Dong; Bonnie Wildman; Katie Hulme; Paul Johnson; Brian R Herts
Journal:  J Am Coll Radiol       Date:  2015-05-21       Impact factor: 5.532

5.  A community hospital's experience with an effective, affordable, and easily implemented CT radiation dose reduction initiative.

Authors:  Ronnie C Chen; John O Johnson
Journal:  J Am Coll Radiol       Date:  2013-03-13       Impact factor: 5.532

6.  Continuous monitoring of CT dose indexes at Dubai Hospital.

Authors:  Jamila S AlSuwaidi; Laila G AlBalooshi; Hashim M AlAwadhi; Ali Rahanjam; Mohamad A ElHallag; Jassem S Ibrahim; Madan M Rehani
Journal:  AJR Am J Roentgenol       Date:  2013-10       Impact factor: 3.959

7.  Does body mass index outperform body weight as a surrogate parameter in the calculation of size-specific dose estimates in adult body CT?

Authors:  Johannes Boos; Rotem S Lanzman; Philipp Heusch; Joel Aissa; Christoph Schleich; Christoph Thomas; Lino M Sawicki; Gerald Antoch; Patric Kröpil
Journal:  Br J Radiol       Date:  2015-12-23       Impact factor: 3.039

8.  Simplifying size-specific radiation dose estimates in pediatric CT.

Authors:  Ranish Deedar Ali Khawaja; Sarabjeet Singh; Beth Vettiyil; Ruth Lim; Michael Gee; Sjirk Westra; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2015-01       Impact factor: 3.959

9.  A study of the midpoint dose to CTDIvol ratio: Implications for CT dose evaluation.

Authors:  Xinhua Li; Kai Yang; Bob Liu
Journal:  Med Phys       Date:  2016-11       Impact factor: 4.071

10.  CT imaging: radiation risk reduction--real-life experience in a metropolitan outpatient imaging network.

Authors:  John O Johnson; Jon M Robins
Journal:  J Am Coll Radiol       Date:  2012-11       Impact factor: 5.532

  10 in total
  1 in total

1.  National survey on dose data analysis in computed tomography.

Authors:  Christina Heilmaier; Reto Treier; Elmar Max Merkle; Hatem Alkadhi; Dominik Weishaupt; Sebastian Schindera
Journal:  Eur Radiol       Date:  2018-05-28       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.