Continuous monitoring of CT dose indexes at Dubai Hospital.

Jamila S AlSuwaidi1, Laila G AlBalooshi, Hashim M AlAwadhi, Ali Rahanjam, Mohamad A ElHallag, Jassem S Ibrahim, Madan M Rehani.   

Abstract

OBJECTIVE: Experience of continuous monitoring and control of patient doses in CT in Dubai Hospital over a period of approximately 4 years (January 2008 through August 2011) is presented.
MATERIALS AND METHODS: Dose measurements-in particular, weighted and volumetric CT dose index, dose-length product (DLP), and estimated effective dose-were regularly monitored using head (16 cm diameter) and body (32 cm diameter) CT phantoms. Patient radiation dose indexes were manually recorded during 2008 for common CT examinations: head, chest, and abdomen and pelvis scans. In 2009-2011, these CT dose data were recorded within the radiology information system and the PACS. Dose reduction actions were taken while maintaining a watch on image quality. The effects of these factors were monitored through change in average DLP on a monthly basis and third quartile annually. Adapted diagnostic reference levels were used for comparison.
RESULTS: The reduction in adult dose indexes in 2010 as compared with 2008 was 52%, 16.4%, and 34.8% for head, chest, and abdomen and pelvis examinations, respectively. For the pediatric group, the reduction was 45.23%, 39.6%, and 43.34% for head, chest, and abdomen and pelvis examinations, respectively.
CONCLUSION: Substantial reduction in DLP for common examinations of adults and children is shown through a program of continuous monitoring and action. The results indicate the need to introduce local diagnostic reference levels to substitute for the adapted ones.

Entities:  

Mesh:

Year:  2013        PMID: 24059376     DOI: 10.2214/AJR.12.10233

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  5 in total

1.  Assessment of radiation dose from abdominal quantitative CT with short scan length.

Authors:  Xinhua Li; Kai Yang; Matthew C DeLorenzo; Bob Liu
Journal:  Br J Radiol       Date:  2017-05-23       Impact factor: 3.039

2.  Opportunities to Reduce CT Radiation Exposure, Experience Over 5 Years at the NIH Clinical Center.

Authors:  William C Kovacs; Jianhua Yao; David A Bluemke; Les R Folio
Journal:  Radiat Prot Dosimetry       Date:  2017-08-01       Impact factor: 0.972

Review 3.  Computed tomography dose optimisation in cystic fibrosis: A review.

Authors:  Helena Ferris; Maria Twomey; Fiachra Moloney; Siobhan B O'Neill; Kevin Murphy; Owen J O'Connor; Michael Maher
Journal:  World J Radiol       Date:  2016-04-28

Review 4.  Computed tomography in cystic fibrosis lung disease: a focus on radiation exposure.

Authors:  Stella Joyce; Brian W Carey; Niamh Moore; David Mullane; Michael Moore; Mark F McEntee; Barry J Plant; Michael M Maher; Owen J O'Connor
Journal:  Pediatr Radiol       Date:  2021-03-20

5.  Implementation of a patient dose monitoring system in conventional digital X-ray imaging: initial experiences.

Authors:  Christina Heilmaier; Niklaus Zuber; Dominik Weishaupt
Journal:  Eur Radiol       Date:  2016-06-23       Impact factor: 5.315

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.