The effect of screw tunnels on the biomechanical stability of vertebral body after pedicle screws removal: a finite element analysis.

Jia-Ming Liu1, Yu Zhang2, Yang Zhou1, Xuan-Yin Chen1, Shan-Hu Huang1, Zi-Kai Hua3, Zhi-Li Liu4.   

Abstract

PURPOSE: Posterior reduction and pedicle screw fixation is a widely used procedure for thoracic and lumbar vertebrae fractures. Usually, the pedicle screws would be removed after the fracture healing and screw tunnels would be left. The aim of this study is to evaluate the effect of screw tunnels on the biomechanical stability of the lumbar vertebral body after pedicle screws removal by finite element analysis (FEA).
METHODS: First, the CT values of the screw tunnels wall in the fractured vertebral bodies were measured in patients whose pedicle screws were removed, and they were then compared with the values of vertebral cortical bone. Second, an adult patient was included and the CT images of the lumbar spine were harvested. Three dimensional finite element models of the L1 vertebra with unilateral or bilateral screw tunnels were created based on the CT images. Different compressive loads were vertically acted on the models. The maximum loads which the models sustained and the distribution of the force in the different parts of the models were recorded and compared with each other.
RESULTS: The CT values of the tunnels wall and vertebral cortical bone were 387.126±62.342 and 399.204±53.612, which were not statistically different (P=0.149). The models of three dimensional tetrahedral mesh finite element of normal lumbar 1 vertebra were established with good geometric similarity and realistic appearance. After given the compressive loads, the cortical bone was the first one to reach its ultimate stress. The maximum loads which the bilateral screw tunnels model, unilateral screw tunnel model, and normal vertebral model can sustain were 3.97 Mpa, 3.83 Mpa, and 3.78 Mpa, respectively. For the diameter of the screw tunnels, the model with a diameter of 6.5 mm could sustain the largest load. In addition, the stress distributing on the outside of the cortical bone gradually decreased as the thickness of the tunnel wall increased.
CONCLUSIONS: Based on the FEA, pedicle screw tunnels would not decrease the biomechanical stability and strength of the vertebral body. A large diameter of screw tunnel and thick tunnel wall were helpful for the biomechanical stability of the vertebral body.

Entities:  

Keywords:  Biomechanics; Finite element analysis; Screw tunnel; Vertebral fracture

Mesh:

Year:  2017        PMID: 28353052     DOI: 10.1007/s00264-017-3453-y

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  16 in total

Review 1.  Application of finite element analysis in implant dentistry: a review of the literature.

Authors:  J P Geng; K B Tan; G R Liu
Journal:  J Prosthet Dent       Date:  2001-06       Impact factor: 3.426

2.  The biomechanics of human femurs in axial and torsional loading: comparison of finite element analysis, human cadaveric femurs, and synthetic femurs.

Authors:  M Papini; R Zdero; E H Schemitsch; P Zalzal
Journal:  J Biomech Eng       Date:  2007-02       Impact factor: 2.097

Review 3.  Finite element application in implant research for treatment of lumbar degenerative disc disease.

Authors:  Qing Hang Zhang; Ee Chon Teo
Journal:  Med Eng Phys       Date:  2008-09-18       Impact factor: 2.242

4.  Failure analysis of broken pedicle screws on spinal instrumentation.

Authors:  Chen-Sheng Chen; Wen-Jer Chen; Cheng-Kung Cheng; Shyh-Hua Eric Jao; Shan-Chang Chueh; Chang-Chih Wang
Journal:  Med Eng Phys       Date:  2005-07       Impact factor: 2.242

Review 5.  Wolff's Law and bone's structural adaptations to mechanical usage: an overview for clinicians.

Authors:  H M Frost
Journal:  Angle Orthod       Date:  1994       Impact factor: 2.079

6.  Assessment of stress patterns on a spinal motion segment in healthy versus osteoporotic bony models with or without disc degeneration: a finite element analysis.

Authors:  Alexander Tsouknidas; Stylianos Orestis Sarigiannidis; Kleovoulos Anagnostidis; Nikolaos Michailidis; Sashin Ahuja
Journal:  Spine J       Date:  2015-01-08       Impact factor: 4.166

7.  Internal fixation of the lumbar spine with pedicle screw plating.

Authors:  R Roy-Camille; G Saillant; C Mazel
Journal:  Clin Orthop Relat Res       Date:  1986-02       Impact factor: 4.176

8.  Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants.

Authors:  Ilser Turkyilmaz; Celal Tumer; Esma Nida Ozbek; Tolga F Tözüm
Journal:  J Clin Periodontol       Date:  2007-08       Impact factor: 8.728

9.  A comprehensive classification of thoracic and lumbar injuries.

Authors:  F Magerl; M Aebi; S D Gertzbein; J Harms; S Nazarian
Journal:  Eur Spine J       Date:  1994       Impact factor: 3.134

10.  Biomechanical comparison of anterior lumbar interbody fusion: stand-alone interbody cage versus interbody cage with pedicle screw fixation -- a finite element analysis.

Authors:  Kyung-Chul Choi; Kyeong-Sik Ryu; Sang-Ho Lee; Yeong Hyeon Kim; Sung Jae Lee; Chun-Kun Park
Journal:  BMC Musculoskelet Disord       Date:  2013-07-26       Impact factor: 2.362

View more
  3 in total

1.  Biomechanical study of oblique lumbar interbody fusion (OLIF) augmented with different types of instrumentation: a finite element analysis.

Authors:  Xin-Yi Cai; Han-Ming Bian; Chao Chen; Xin-Long Ma; Qiang Yang
Journal:  J Orthop Surg Res       Date:  2022-05-14       Impact factor: 2.677

2.  Biomechanical Effects of Pedicle Screw Positioning on the Surgical Segment in Models After Oblique Lumbar Interbody Fusion: An in-silico Study.

Authors:  Chen Xu; Chenyi Huang; Ping Cai; Zhongxin Fang; Zhangchao Wei; Fei Liu; Jingchi Li; Yang Liu
Journal:  Int J Gen Med       Date:  2022-02-02

3.  Poor bone mineral density aggravates adjacent segment's motility compensation in patients with oblique lumbar interbody fusion with and without pedicle screw fixation: An in silico study.

Authors:  Chen-Yi Huang; Zi-Fan Zhang; Xiao-Yu Zhang; Fei Liu; Zhong-Xin Fang; Zhi-Peng Xi; Jing-Chi Li
Journal:  Front Surg       Date:  2022-08-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.