Abstract

Revised spinal surgery is needed when there is a broken pedicle screw in the patient. This study investigated the pedicle screw breakage by conducting retrieval analyses of broken pedicle screws from 16 patients clinically and by performing stress analyses in the posterolateral fusion computationally using finite element (FE) models. Fracture surface of screws was studied by scanning electron microscope (SEM). The FE model of the posterolateral fusion with the screw showed that screws on the caudal side had larger axial stress than those on the cephalic side, supporting the clinical findings that 75% of the patients had the screw breakage on the caudal side. SEM fractography showed that all broken screws exhibited beach marks or striations on the fractured surface, indicating fatigue failure. Screws of patients with spinal fracture showed fatigue striations and final ductile fracture around the edge. Among the 16 patients who had broken pedicle screws 69% of them achieved bone union in the bone graft, showing that bone union in the bone graft did not warrant the prevention of screw breakage.

Entities:  

Mesh:

Year:  2005        PMID: 15990065     DOI: 10.1016/j.medengphy.2004.12.007

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  21 in total

1.  Evaluation of iliac screw, S2 alar-iliac screw and laterally placed triangular titanium implants for sacropelvic fixation in combination with posterior lumbar instrumentation: a finite element study.

Authors:  Gloria Casaroli; Fabio Galbusera; Ruchi Chande; Derek Lindsey; Ali Mesiwala; Scott Yerby; Marco Brayda-Bruno
Journal:  Eur Spine J       Date:  2019-05-15       Impact factor: 3.134

2.  Preclinical evaluation of posterior spine stabilization devices: can we compare in vitro and in vivo loads on the instrumentation?

Authors:  Luigi La Barbera; Fabio Galbusera; Hans-Joachim Wilke; Tomaso Villa
Journal:  Eur Spine J       Date:  2016-09-16       Impact factor: 3.134

3.  The effect of screw tunnels on the biomechanical stability of vertebral body after pedicle screws removal: a finite element analysis.

Authors:  Jia-Ming Liu; Yu Zhang; Yang Zhou; Xuan-Yin Chen; Shan-Hu Huang; Zi-Kai Hua; Zhi-Li Liu
Journal:  Int Orthop       Date:  2017-03-28       Impact factor: 3.075

4.  Replacement of Destructive Pull-out Test with Modal Analysis in Primary Fixation Stability Assessment of Spinal Pedicle Screw.

Authors:  Mohammadjavad Einafshar; Ata Hashemi; Gerrit Harry van Lenthe
Journal:  Arch Bone Jt Surg       Date:  2022-02

5.  Biomechanical effects of an oblique lumbar interbody fusion combined with posterior augmentation: a finite element analysis.

Authors:  Shengjia Huang; Shaoxiong Min; Suwei Wang; Anmin Jin
Journal:  BMC Musculoskelet Disord       Date:  2022-06-27       Impact factor: 2.562

6.  Standalone titanium/polyetheretherketone interbody cage for anterior lumbar interbody fusion: Clinical and radiological results at 24 months.

Authors:  Ralph J Mobbs; Tajrian Amin; Kevin Phan; Darweesh Al Khawaja; Wen Jie Choy; William C H Parr; Vedran Lovric; William R Walsh
Journal:  J Craniovertebr Junction Spine       Date:  2022-03-09

7.  Effect of bone material properties on effective region in screw-bone model: an experimental and finite element study.

Authors:  Shuai Liu; Wei Qi; Yang Zhang; Zi-Xiang Wu; Ya-Bo Yan; Wei Lei
Journal:  Biomed Eng Online       Date:  2014-06-21       Impact factor: 2.819

8.  Stress distribution of different lumbar posterior pedicle screw insertion techniques: a combination study of finite element analysis and biomechanical test.

Authors:  Mingzhi Song; Kebin Sun; Zhonghai Li; Junwei Zong; Xiliang Tian; Kai Ma; Shouyu Wang
Journal:  Sci Rep       Date:  2021-06-21       Impact factor: 4.379

9.  Transpedicular screw fixation in the thoracic and lumbar spine with a novel cannulated polyaxial screw system.

Authors:  Lutz Weise; Olaf Suess; Thomas Picht; Theodoros Kombos
Journal:  Med Devices (Auckl)       Date:  2008-10-13

10.  Treatment of unstable thoracolumbar fractures through short segment pedicle screw fixation techniques using pedicle fixation at the level of the fracture: a finite element analysis.

Authors:  Changqing Li; Yue Zhou; Hongwei Wang; Jun Liu; Liangbi Xiang
Journal:  PLoS One       Date:  2014-06-10       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.