Literature DB >> 28298668

Unexplained Obstruction of an Integrated Cardiotomy Filter During Cardiopulmonary Bypass.

Cory M Alwardt, Donald S Wilson, Octavio E Pajaro.   

Abstract

Cardiopulmonary bypass (CPB) is considered relatively safe in most cases, yet is not complication free. We present a case of an integrated cardiotomy filter obstruction during CPB, requiring circuit reconfiguration. Approximately an hour after uneventful initiation of CPB the integrated cardiotomy filter became obstructed over several minutes, requiring circuit reconfiguration using an external cardiotomy filter to maintain functionality. Following reconfiguration, CPB was maintained with a fully functional circuit allowing safe patient support throughout the remainder of CPB. Postoperatively, there was no sign of thrombus or mechanical obstruction of the filter, which was sent to the manufacturer for analysis. The cause of the obstruction was unclear even after chemical analysis, visual inspection, and a review of all techniques and products to which the patient was exposed. The patient had a generally routine hospital stay, with no signs or symptoms related to the incident. To our knowledge, this is the first report describing an obstructed integrated cardiotomy filter. An appropriate readiness plan for such an incident includes proper venting of the filter chamber, a method for detecting an obstruction, and a plan for circuit reconfiguration. This case illustrates the need for a formal reporting structure for incidents or "near miss" incidents during CPB.

Entities:  

Keywords:  CPB; cardiopulmonary bypass; cardiotomy; filter; obstruction

Mesh:

Year:  2017        PMID: 28298668      PMCID: PMC5347222     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  9 in total

1.  A retrospective study on perfusion incidents and safety devices.

Authors:  B L Mejak; A Stammers; E Rauch; S Vang; T Viessman
Journal:  Perfusion       Date:  2000-01       Impact factor: 1.972

2.  An update on perfusion safety: does the type of perfusion practice affect the rate of incidents related to cardiopulmonary bypass?

Authors:  A H Stammers; B L Mejak
Journal:  Perfusion       Date:  2001-05       Impact factor: 1.972

3.  D'où venons-nous/que somes nous/où allons nous? Accidents are inevitable.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2012-03

4.  Dutch perfusion incident survey.

Authors:  Ingrid Groenenberg; Patrick W Weerwind; Peter A M Everts; Jos G Maessen
Journal:  Perfusion       Date:  2010-07-14       Impact factor: 1.972

5.  A 2013 Survey on Pressure Monitoring in Adult Cardiopulmonary Bypass Circuits: Modes and Applications.

Authors:  Laura Rigg; Bruce Searles; Edward Morse Darling
Journal:  J Extra Corpor Technol       Date:  2014-12

6.  Australasian perfusion incident survey.

Authors:  O F Jenkins; R Morris; J M Simpson
Journal:  Perfusion       Date:  1997-09       Impact factor: 1.972

7.  FOCUS: the Society of Cardiovascular Anesthesiologists' initiative to improve quality and safety in the cardiovascular operating room.

Authors:  Atilio Barbeito; William Travis Lau; Nathaen Weitzel; James H Abernathy; Joyce Wahr; Jonathan B Mark
Journal:  Anesth Analg       Date:  2014-10       Impact factor: 5.108

8.  Trends in healthcare incident reporting and relationship to safety and quality data in acute hospitals: results from the National Reporting and Learning System.

Authors:  A Hutchinson; T A Young; K L Cooper; A McIntosh; J D Karnon; S Scobie; R G Thomson
Journal:  Qual Saf Health Care       Date:  2009-02

9.  Report from AmSECT's International Consortium for Evidence-Based Perfusion: American Society of Extracorporeal Technology Standards and Guidelines for Perfusion Practice: 2013.

Authors:  Robert A Baker; Shahna L Bronson; Timothy A Dickinson; David C Fitzgerald; Donald S Likosky; Nicholas B Mellas; Kenneth G Shann
Journal:  J Extra Corpor Technol       Date:  2013-09
  9 in total