Literature DB >> 10676868

A retrospective study on perfusion incidents and safety devices.

B L Mejak1, A Stammers, E Rauch, S Vang, T Viessman.   

Abstract

Despite the acceptance of extracorporeal circulation as an effective modality to facilitate cardiac surgery, patient outcomes can be negatively influenced by the occurrence of perfusion incidents. A perfusion survey was conducted to identify safety techniques and incidents related to cardiopulmonary bypass (CPB). An 80-question survey was mailed to chief perfusionists of all 1030 USA cardiac surgical centers using CPB. The survey was designed to examine practices and incidents that occurred during a 2-year period (July 1996 to July 1998). Five-hundred-and-fifty-two (54% response rate) surveys were returned, which accounted for 797 hospitals (79% of all cardiac centers) and 653,621 surgical procedures. Of the 27 identified CPB safety devices, the highest utilization was arterial line filters (98.5%) and the lowest arterial line bubble traps (3.4%). Of the reported cases, a CPB incident occurred once every 138 cases. The most common occurring incidents were protamine reactions (1:783), coagulation problems (1:771), and heater/cooler failures 11:1809). The rate of occurrence of an incident resulting in a serious injury or death was one for every 1453 procedures. Although techniques and safety devices create a relatively secure environment for CPB, lower incident rates may be achieved with further improvements in coagulation monitoring and incident reporting.

Entities:  

Mesh:

Year:  2000        PMID: 10676868     DOI: 10.1177/026765910001500108

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  20 in total

1.  D'où venons-nous/que somes nous/où allons nous? Accidents are inevitable.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2012-03

2.  A Novel Method to Detect an Oxygenator Defect Prior to Cardiopulmonary Bypass Initiation.

Authors:  Armindo Fernandes; Eric Laliberte; Karine Toledano; Philippe Demers
Journal:  J Extra Corpor Technol       Date:  2015-09

3.  Vacuum-Assisted Venous Drainage: A 2014 Safety Survey.

Authors:  Rachel Gambino; Bruce Searles; Edward M Darling
Journal:  J Extra Corpor Technol       Date:  2015-09

4.  Should Air Bubble Detectors Be Used to Quantify Microbubble Activity during Cardiopulmonary Bypass?

Authors:  Richard F Newland; Robert A Baker; Annette L Mazzone; Vijaykumar N Valiyapurayil
Journal:  J Extra Corpor Technol       Date:  2015-09

5.  A 2013 Survey on Pressure Monitoring in Adult Cardiopulmonary Bypass Circuits: Modes and Applications.

Authors:  Laura Rigg; Bruce Searles; Edward Morse Darling
Journal:  J Extra Corpor Technol       Date:  2014-12

Review 6.  From trash to leucocytes: what are we filtering and why?

Authors:  Simon J Mitchell
Journal:  J Extra Corpor Technol       Date:  2006-03

7.  Getting it right: optimizing the patient and technique for the procedure.

Authors:  Alfred H Stammers
Journal:  J Extra Corpor Technol       Date:  2009-12

8.  Incident Reporting in Perfusion: Current Perceptions on PIRS-2.

Authors:  Timothy W Willcox; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2020-03

9.  Isoflurane cracks the polycarbonate connector of extra-corporeal circuit -A case report-.

Authors:  Hyung Sun Lim; Sung Hun Cho; Deok Kyu Kim; Sang Kyi Lee; Seong Hoon Ko
Journal:  Korean J Anesthesiol       Date:  2010-03-29

10.  Calculating mixed venous saturation during veno-venous extracorporeal membrane oxygenation.

Authors:  Joshua L Walker; Jonathan Gelfond; Lee Ann Zarzabal; Edward Darling
Journal:  Perfusion       Date:  2009-11-30       Impact factor: 1.972

View more