| Literature DB >> 28297426 |
Amy Walker1,2,3, Peter Metcalfe1,2,3, Gary Liney1,2,3,4, Vikneswary Batumalai2,3,5, Kylie Dundas1,2,3, Carri Glide-Hurst6, Geoff P Delaney2,5,7,8, Miriam Boxer2,5, Mei Ling Yap2,3,5,7,8, Jason Dowling9, David Rivest-Henault9, Elise Pogson1,2,3, Lois Holloway1,2,3,5,4.
Abstract
The purpose of this study was to determine the impact of magnetic resonance imaging (MRI) geometric distortions when using MRI for target delineation and planning for whole-breast, intensity-modulated radiotherapy (IMRT). Residual system distortions and combined systematic and patient-induced distortions are considered. This retrospective study investigated 18 patients who underwent whole-breast external beam radiotherapy, where both CT and MRIs were acquired for treatment planning. Distortion phantoms were imaged on two MRI systems, dedicated to radiotherapy planning (a wide, closed-bore 3T and an open-bore 1T). Patient scans were acquired on the 3T system. To simulate MRI-based planning, distortion maps representing residual system distortions were generated via deformable registration between phantom CT and MRIs. Patient CT images and structures were altered to match the residual system distortion measured by the phantoms on each scanner. The patient CTs were also registered to the corresponding patient MRI scans, to assess patient and residual system effects. Tangential IMRT plans were generated and optimized on each resulting CT dataset, then propagated to the original patient CT space. The resulting dose distributions were then evaluated with respect to the standard clinically acceptable DVH and visual assessment criteria. Maximum residual systematic distortion was measured to be 7.9 mm (95%<4.7mm) and 11.9 mm (95%<4.6mm) for the 3T and 1T scanners, respectively, which did not result in clinically unacceptable plans. Eight of the plans accounting for patient and systematic distortions were deemed clinically unacceptable when assessed on the original CT. For these plans, the mean difference in PTV V95 (volume receiving 95% prescription dose) was 0.13±2.51% and -0.73±1.93% for right- and left-sided patients, respectively. Residual system distortions alone had minimal impact on the dosimetry for the two scanners investigated. The combination of MRI systematic and patient-related distortions can result in unacceptable dosimetry for whole-breast IMRT, a potential issue when considering MRI-only radiotherapy treatment planning. PACS number(s): 87.61.-c, 87.57.cp, 87.57.nj, 87.55.D.Entities:
Keywords: breast IMRT; geometric distortion; magnetic resonance imaging
Mesh:
Year: 2016 PMID: 28297426 PMCID: PMC5495026 DOI: 10.1120/jacmp.v17i5.6242
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Methodology schematic showing the image distortion processes for the phantom and patient datasets.
The DVH criteria for plan evaluation (based on RTOG guidelines and clinical experience).
|
|
|
|
|---|---|---|
| PTV |
|
|
|
|
| |
| Seroma |
|
|
|
|
| |
| Combined Lung |
|
|
|
|
| |
| Ipsilateral Lung |
|
|
| Heart |
|
|
| Contralateral Breast |
|
|
Variations in the contour volumes and overlap comparisons between original and distorted contour volumes. Phantom data consists of residual system distortions, while patient scans also include patient‐specific distortions and setup uncertainties.
|
|
|
|
|
| |
|---|---|---|---|---|---|
| 3T phantom | CTV | (R) |
|
|
|
| (L) |
|
|
| ||
| Seroma | (R) |
|
|
| |
| (L) |
|
|
| ||
| CB | (R) |
|
|
| |
| (L) |
|
|
| ||
| Combined Lung |
|
|
| ||
| Heart |
|
|
| ||
| 1T phantom | CTV | (R) |
|
| |
| (L) |
|
| |||
| Seroma | (R) |
|
| ||
| (L) |
|
| |||
| CB | (R) |
|
| ||
| (L) |
|
| |||
| Combined Lung |
|
| |||
| Heart |
|
| |||
| As above | |||||
| 3T patient | CTV | (R) |
|
| |
| (L) |
|
| |||
| Seroma | (R) |
|
| ||
| (L) |
|
| |||
| CB | (R) |
|
| ||
| (L) |
|
| |||
| Combined Lung |
|
| |||
| Heart |
|
| |||
; ; ; ; .
Figure 2Coronal comparison of the CTV (green), contralateral breast (pink), lung (blue) and heart (red) volumes relative to distortions for (a) the 3T phantom, (b) an example 3T patient dataset (incorporating patient and system related distortions as well as set‐up uncertainties), and (c) the 1T phantom. Scales set relative to maximum distortion in (b). Visualization by http://smili‐project.sourceforge.net/.
Quantitative assessment of the DVH parameters and the clinical plan acceptance. All numbers were computed based on dose optimized on distorted datasets less dose when recomputed on undistorted dataset.
|
|
|
|
| |
|---|---|---|---|---|
|
| ||||
| PTV |
|
|
|
|
| Max 1cc |
|
|
| |
| Seroma |
|
|
|
|
| Max 1cc |
|
|
| |
| Ipsilateral Lung |
|
|
|
|
| Heart | Mean |
|
|
|
| CB | Max 1 cc |
|
|
|
|
| ||||
| PTV |
|
|
|
|
| Max 1cc |
|
|
| |
| Seroma |
|
|
|
|
| Max 1cc |
|
|
| |
| Ipsilateral Lung |
|
|
|
|
| Heart | Mean |
|
|
|
| CB | Max 1 cc |
|
|
|
| Combined Lung |
|
|
|
|
| Plan pass/fail | Overall |
|
|
|
; ; .
Figure 3DVHs for two patients (left and right) showing the variation in dose coverage variation that occurs when incorporating different distortion factors. .