| Literature DB >> 28270382 |
Dong Wen1,2, Xingting Zhang1,2, Xingyu Liu3, Jianbo Lei2,4.
Abstract
BACKGROUND: Wearable devices are gaining increasing market attention; however, the monitoring accuracy and consistency of the devices remains unknown.Entities:
Keywords: activities of daily living; fitness trackers; health status; monitoring, physiologic; motor activity
Mesh:
Year: 2017 PMID: 28270382 PMCID: PMC5360892 DOI: 10.2196/jmir.6874
Source DB: PubMed Journal: J Med Internet Res ISSN: 1438-8871 Impact factor: 5.428
Figure 1Box plot of the measurement distribution of the number of steps by different devices.
Figure 6Box plot of the measurement distribution of deep sleep duration by different devices.
Figure 2Box plot of the measurement distribution of distance by different devices.
Figure 3Box plot of the measurement distribution of activity duration by different devices.
Figure 4Box plot of the measurement distribution of energy expenditure by different devices.
Figure 5Box plot of the measurement distribution of sleep duration by different devices.
Intraclass correlation coefficient (ICC) of the measurements by different wearable devices
| Items | Intraclass correlation coefficient | 95% CI |
| Number of steps | .89 | 0.83-0.94 |
| Distance | .84 | 0.75-0.91 |
| Activity duration | .59 | 0.44-0.75 |
| Total EEa | .59 | 0.37-0.77 |
| Activity EE | .57 | 0.41-0.74 |
| Sleep duration | .30 | 0.13-0.52 |
| Deep sleep duration | .27 | 0.08-0.50 |
aEE: energy expenditure.
Figure 7Coefficient of variation of daily measurements of the number of steps by different devices.
Figure 13Coefficient of variation of daily measurements of deep sleep duration by different devices.
Comparison with previous studies.
| Indicators | Steps | Distance | Activity duration | EEa | Sleep | Deep sleep duration |
| This study | Excellent consistency with CVb (2-38%) | Excellent consistency with CV (5-30%) | Fair consistency with CV | Fair consistency with CV for total EE (.1-17%) and activity EE CV (22-100%) | Poor consistency with CV | Poor consistency with CV (35-117%) |
| Previous studies | Consumer-grade wearable devices provided consistently similar step counts with research-grade devices for average daily activity ( | The inter-device reliability of wearable devices in measuring distance was excellent for all treadmill speeds (ICCc≥.90) [ | Consumer-level wearable devices showed moderate validity for measurement of moderate to vigorous physical activity in free-living conditions ( | Consumer-level wearable devices showed moderate validity for measurement of total daily EE in free-living conditions ( | Consumer-level wearable devices showed strong validity for measurement of sleep duration in free-living conditions ( | Consumer-grade wearable devices showed good agreements with PSGd for sleep efficiency, and they overestimated PSG sleep efficiency slightly [ |
| The inter-device reliability of wearable devices in measuring steps in free-living conditions is good (ICC≥.90) [ | Distance errors in wearable devices were within 5% in level walking, and they overestimated distance for stair walking by at least 45% [ | Consumer-grade wearable devices can’t accurately capture activity data across the entire 24-h day, error rates ranged from 51.8% to 92% for moderate to vigorous physical activity [ | Consumer-grade wearable devices reasonably and reliably estimate EE during walking and running (ICC≥.95) [ | Consumer-grade wearable devices performed consistently compared with each other (reliability=96.5-99.1%), and they overestimated sleep time by an average of 67.1 min compared with PSG [ | Consumer-grade wearable devices performed consistently compared with each other (reliability=96.5%-99.1%), and they overestimated sleep efficiency by an average of 14.5% compared with PSG [ |
aEE: energy expenditure.
bCV: coefficient of variation.
cICC: intraclass correlation coefficient.
dPSG: Polysomnography.
Figure 14Scatter plot for each device showing measurements of the number of steps and distance.
Figure 15Scatter plot for each device showing the measurements of activity duration and energy expenditure.
Correlation coefficient on measurements by mobile phones and smart watches of the same brand.
| Brand | Steps | Distance | EEa |
| Apple | .99 | .99 | .96 |
| Samsung | .96 | .91 | .88 |
aEE: energy expenditure.