Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function.

Sung-Min Park1, Rungkiet Kamondetdacha, John A Nyenhuis.   

Abstract

PURPOSE: To develop and demonstrate a method to calculate the temperature rise that is induced by the radio frequency (RF) field in MRI at the electrode of an implanted medical lead.
MATERIALS AND METHODS: The electric field near the electrode is calculated by integrating the product of the tangential electric field and a transfer function along the length of the lead. The transfer function is numerically calculated with the method of moments. Transfer functions were calculated at 64 MHz for different lengths of model implants in the form of bare wires and insulated wires with 1 cm of wire exposed at one or both ends.
RESULTS: Heating at the electrode depends on the magnitude and the phase distribution of the transfer function and the incident electric field along the length of the lead. For a uniform electric field, the electrode heating is maximized for a lead length of approximately one-half a wavelength when the lead is terminated open. The heating can be greater for a worst-case phase distribution of the incident field.
CONCLUSION: The transfer function is proposed as an efficient method to calculate MRI-induced heating at an electrode of a medical lead. Measured temperature rises of a model implant in a phantom were in good agreement with the rises predicted by the transfer function. The transfer function could be numerically or experimentally determined.

Mesh:

Year:  2007        PMID: 17969143     DOI: 10.1002/jmri.21159

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  40 in total

1.  RF induced energy for partially implanted catheters: a computational study.

Authors:  Elena Lucano; Micaela Liberti; Tom Lloyd; Francesca Apollonio; Steve Wedan; Wolfgang Kainz; Leonardo M Angelone
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2016-08

2.  The 'virtual DBS population': five realistic computational models of deep brain stimulation patients for electromagnetic MR safety studies.

Authors:  Bastien Guerin; Maria Ida Iacono; Mathias Davids; Darin Dougherty; Leonardo M Angelone; Lawrence L Wald
Journal:  Phys Med Biol       Date:  2019-02-04       Impact factor: 3.609

3.  Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heating.

Authors:  N Gudino; M Sonmez; Z Yao; T Baig; S Nielles-Vallespin; A Z Faranesh; R J Lederman; M Martens; R S Balaban; M S Hansen; M A Griswold
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

Review 4.  [MR safety assessment of active implanted medical devices. German version].

Authors:  Sarra Aissani; Elmar Laistler; Jacques Felblinger
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

Review 5.  MR safety assessment of active implantable medical devices.

Authors:  Sarra Aissani; Elmar Laistler; Jacques Felblinger
Journal:  Radiologe       Date:  2019-12       Impact factor: 0.635

6.  Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.

Authors:  Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald
Journal:  Neuroimage       Date:  2016-12-21       Impact factor: 6.556

Review 7.  A Review of Numerical Simulation and Analytical Modeling for Medical Devices Safety in MRI.

Authors:  J Kabil; L Belguerras; S Trattnig; C Pasquier; J Felblinger; A Missoffe
Journal:  Yearb Med Inform       Date:  2016-11-10

8.  A numerical investigation on the effect of RF coil feed variability on global and local electromagnetic field exposure in human body models at 64 MHz.

Authors:  Elena Lucano; Micaela Liberti; Tom Lloyd; Francesca Apollonio; Steve Wedan; Wolfgang Kainz; Leonardo M Angelone
Journal:  Magn Reson Med       Date:  2017-04-18       Impact factor: 4.668

9.  MRI Conditional Actively Tracked Metallic Electrophysiology Catheters and Guidewires With Miniature Tethered Radio-Frequency Traps: Theory, Design, and Validation.

Authors:  Akbar Alipour; Eric S Meyer; Charles L Dumoulin; Ronald D Watkins; Hassan Elahi; Wolfgang Loew; Jeffrey Schweitzer; Gregory Olson; Yue Chen; Susumu Tao; Michael Guttman; Aravindan Kolandaivelu; Henry R Halperin; Ehud J Schmidt
Journal:  IEEE Trans Biomed Eng       Date:  2019-09-13       Impact factor: 4.538

10.  Reducing RF-induced Heating near Implanted Leads through High-Dielectric Capacitive Bleeding of Current (CBLOC).

Authors:  Laleh Golestanirad; Leonardo M Angelone; John Kirsch; Sean Downs; Boris Keil; Giorgio Bonmassar; Lawrence L Wald
Journal:  IEEE Trans Microw Theory Tech       Date:  2019-01-01       Impact factor: 3.599

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.