Estimate interaction timing in a large volume HgI2 detector using cathode pulse waveforms.

L J Meng1, Z He1.   

Abstract

This paper presents experimental results on the timing resolution achieved with a 5 mm thick HgI2 detector. The timing information was derived based on the cathode pre-amplifier pulse waveform, measured using a digital oscilloscope. The times of interaction were estimated by fitting the measured pulse waveforms to pre-defined waveform models. With this approach, problems related to the conventional leading edge or constant fraction triggering, such as slow charge carrier mobility, pulse shape variation and depth-dependent detector response can be greatly reduced. As a result, we showed a 13 ns timing resolution measured using the 5 mm thick HgI2 detector and a BaF2 coincidence detector with 511 keV full energy events. In this paper, we discuss several waveform models and the results achieved using these models.

Entities:  

Keywords:  Large volume HgI2 detector; Pulse waveform; Timing resolution

Year:  2005        PMID: 28260824      PMCID: PMC5333923          DOI: 10.1016/j.nima.2005.01.323

Source DB:  PubMed          Journal:  Nucl Instrum Methods Phys Res A        ISSN: 0168-9002            Impact factor:   1.455


  1 in total

1.  Charge transport in arrays of semiconductor gamma-ray detectors.

Authors: 
Journal:  Phys Rev Lett       Date:  1995-07-03       Impact factor: 9.161

  1 in total
  2 in total

1.  Simulation study of the second-generation MR-compatible SPECT system based on the inverted compound-eye gamma camera design.

Authors:  Xiaochun Lai; Ling-Jian Meng
Journal:  Phys Med Biol       Date:  2018-02-12       Impact factor: 3.609

2.  Hybrid Pixel-Waveform (HPWF) Enabled CdTe Detectors for Small Animal Gamma-Ray Imaging Applications.

Authors:  A Groll; K Kim; H Bhatia; J C Zhang; J H Wang; Z M Shen; L Cai; J Dutta; Q Li; L J Meng
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2016-11-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.