Literature DB >> 28243402

An Analytical Method to Calculate Phantom Scatter Factor for Photon Beam Accelerators.

Mohammad Javad Tahmasebi Birgani1, Nahid Chegeni2, Mohammad Ali Behrooz3, Marziyeh Bagheri4, Amir Danyaei2, Azin Shamsi5.   

Abstract

INTRODUCTION: One of the important input factors in the commissioning of the radiotherapy treatment planning systems is the phantom scatter factor (Sp) which requires the same collimator opening for all radiation fields. In this study, we have proposed an analytical method to overcome this issue.
METHODS: The measurements were performed using Siemens Primus Plus with photon energy 6 MV for field sizes from 5×5cm2 to 40×40cm2. Phantom scatter factor was measured through the division of total scatter output factors (Scp), and collimator scatter factor (Sc).
RESULTS: The mean percent difference between the measured and calculated Sp was 1.00% and -3.11% for 5×5, 40×40 cm2 field size respectively.
CONCLUSION: This method is applicable especially for small fields used in IMRT which, measuring collimator scatter factor is not reliable due to the lateral electron disequilibrium.

Entities:  

Keywords:  Phantom scatter factor; Radiotherapy; Total scatter factor

Year:  2017        PMID: 28243402      PMCID: PMC5308490          DOI: 10.19082/3523

Source DB:  PubMed          Journal:  Electron Physician        ISSN: 2008-5842


1. Introduction

In radiation therapy, the calculation of the monitor unit (MU) is necessary to deliver a prescribed absorbed dose for the planning target volume (PTV) (1). One of the important factors to determine MU is the scatter factor. According to Kahn et al. the collimator scatter factor (Sc) is commonly called the output factor or head scatter factor, and may be defined as the ratio of the output in air for a given field to that for the reference field (e.g., 10 × 10 cm2) with a buildup cap of a size large enough to provide electronic equilibrium (2). To prevent contaminating electrons to the detector, the columnar mini phantom is used instead of buildup cap (3). The Sp is defined as the ratio of the absorbed dose at the depth of maximum dose (dmax) for a given field to that in the reference field at the same depth with the same collimator opening (4).The Sp becomes difficult to measure in practice (5); An indirect method of determining Sp, relates the total scatter factor (Scp) to the collimator scatter factor (Sc) (Figure 1, Eq.1); where Scp (r) is defined as the ratio of the absorbed dose at dmax in a water phantom for specific field to the absorbed dose in a reference field (6–10). Most treatment planning systems (TPS) need to enter Scp, Sc and Sp. Error in the measurement of Scp and Sc will cause problems to drive Sp (11).Therefore, it is necessary to calculate directly the phantom scatter factor using Monte Carlo simulation or analytical methods. According to the Sp definition, each measurement in the phantom has scattering effects from both phantom and collimator together. To eliminate the effect of collimator in this measurement, the field collimator should be tried to be constant during the measurement process. To solve this issue, Khan et al. produced the treatment field in phantom by collimator and then, without a change in collimator, they created the field references by lead blocks (12); Then, for the same condition, they defined the ratio of dosimeter readings for these two fields as Sp regardless of the block scattering involved in the measurement. An analytical model for calculating Sp is presented in this paper. We proposed a new setting to achieve the same collimator opening and all measuring done in phantom, therefore the lack of the lateral electron equilibrium for small field is not an issue.
Figure 1

Formula in the study

2. Material and Methods

2.1. Theory

According to the definition, the in-air output factor (Sc) can be written as Figure 1, Eq.2; which numerator and denominator are chamber readings (R) for the same number of monitor units (M) in the presence of a build-up cap for a field with dimensions L and L0 (as a reference field e.g. 10×10cm2) at SSD=100cm and collimator opening CL, C0 respectively. For the total scatter factor, measuring at the dmax in water phantom for a specific field (L) to that reference field (L0) for the same number of monitor units (M) (13), it would be written as Figure 1, Eq.3. To prevent electron contamination of the photon beam, it is recommended that all readings are performed at a depth of d=10 cm and then they are transmitted to the maximum depth using PDD (d) for that field size (14). So, the formula in Figure 1, Eq.4. According to Figure 2., L, L0 (field size), CL, and C0 (collimator opening) are shown as AB, EF, AÔB, and EÔF respectively. As can be seen in Figure 2, the total scatter factor is obtained by dividing the dosimeter reading which is caused by primary and secondary radiation (S′) from two volume VABCD and reference volume VEFGH and then correct for depth d to dmax. The Sp is defined as the ratio of the absorbed dose at the dmax in phantom (water or solid water) for a given field to that in the reference field with the same collimator opening (C0) to eliminate the collimator effect (4). Therefore, the formula in Figure 1, Eq.5. In this study, to maintain the collimator opening for all fields, the phantom surface can be pulled up (for LL0) (Figure 2). For example, for a field smaller than L0, L=AB, it is necessary to pull up phantom surface to J=(L/L0) ×SSD to achieve the same collimator opening. However, care must be which the dosimeter reading at J′ needs to be corrected for SSD changes using Mayneord Factor. Therefore; we have to perform dosimetry in several SSDs and the main problem is phantom surface collision with Linac head for small fields. On the other hand, according to Eq. 3 and 5, the difference between the Sp and the total scatter factor is the amount of the collimator opening. Therefore, by considering a correction factor (for the effect of collimator) in the total scatter factor (Eq. 4), Sp can be achieved (Eq.5) as Figure 1, Eq.6. In Figure 2 with regards to the Pythagorean theorem, the trapezoid ABC″D″ and A′B′C′D′ are equal; therefore, the Sp is obtained by dividing the dosimeter readings at S′ form two volume VA′B′C″D″ and reference volume VEFGH. In this method, we propose to apply a factor (α) to correct scatter portion reached to the dosimeter (S′) from VABCD to VABC″D″. As mentioned in previous studies (15, 16), suppose that there is a parallel beam striking the surface of the phantom and consider a reference plane normal to the central axis, placed at depth d=10cm which the origin is on the central axis. Hence, each element (dx.dy) on this plane acts as a source of scattered radiation. The amount of scatter radiation reaching the central axis (dosimeter location, S′) is inversely proportional to the square of the distance between surface elements and the origin. Therefore, with taking into account the photon attenuation μ, α is defined as Figure 1, Eq.7; Where μ is energy attenuation coefficient which depends on energy, field size and depth (17, 18). Therefore, the formula in Figure 1, Eq.8. α will be greater or lesser than 1 while phantom surface was pulled up or down respectively. Let us eliminate the small area at the origin of the coordinate with radius ɛ (refers to the ion chamber radius =3mm) to overcome singularity in nominator and denominator in Eq.7 and use polar system (Figure 3). Therefore, the scatter contribution to dosimeter location (S′) for a rectangular field with dimensions (2Y × 2X) (Figure 3) is calculated as Figure 1, Eq.9. Finally, a general computer program using MATLAB 7.12.0 software was written to calculate α using the numerical method to solve Eq. 9.
Figure 2

(a) Diagram of under-radiation treatment volume in the standard mode of SSD=100 for reference field (VEFGH) and a small field (VABCD). In order to maintain collimator opening for smaller fields than the reference field, the phantom surface can be pulled up to the collimator location (A′B′). (b) DMCN, D″M″C″N″ are the projection the field size L at distance OS′, OJ′.

Figure 3

Polar system to calculate the scatter contribution to dosimeter location (S′) for a rectangular field with dimensions (2Y × 2X).

2.2. Dosimetry and measurements

The linear accelerator used in this study was a Siemens Primus Plus with photon energy 6 MV. A Scanditronix blue phantom (Wellhofer, Germany) (50 cm ×50 cm ×50 cm) and the Farmer ionization chamber 0.6cc were applied for measuring. Total scatter output factors, Scp, were measured at 10-cm depth water (S′ in Figure 2) and source-surface distance (SSD) 100cm and for field sizes from 5×5cm2 to 40×40cm2. For Sc measurements, a polystyrene build-up cap was employed at maximum depth and source-surface distance (SSD) 100-cm. The Sp was measured indirectly using Eq. 1 as a routine method and then compared with Sp which was calculated by Eq.6, 7.

3. Results

As has been shown in Table 1, the Sp for the photon beam was indirectly measured for all field sizes using measuring Scp and Sc according to Eq.1. The correction factor in Eq. 6 (α) was calculated using numerical calculation then by multiplying to Scp, Sp-cal was obtained. As can be seen, variations regarded to the energy attenuation coefficient (μ) changes Sp-cal results. While, elimination of the photon attenuation shows better results (fourth column). The measured and calculated phantom scatter were compared by percentage of difference which is mentioned the last three columns. According to Figure 4. for fields, which are far from the reference field (e. g. 5 and 40 cm), the differences between calculation and measurement are further than fields next to the reference field. This difference for fields larger than the reference field is becoming more evident.
Table 1

Phantom scatter factor (Sp) measured and calculated based on routine method (Sp-meas= Scp/Sc) and analytical method (Sp-cal= α.Scp).

L (cm) (Field size)ScScpSp-meas (Scp/Sc)Sp-cal=α.ScpDiff% (Spmeas-Spcal)/ Spmeas
μ=0μ (r)+μ (d,L)*μ=0μ (r)+μ (d,L)*
5×50.9570.9480.9900.9830.9800.9780.721.051.23
6×60.9710.9630.9920.9860.9830.9820.620.871.01
7×70.9820.9760.9940.9900.9880.9870.390.570.67
8×80.9910.9931.0021.0011.0000.9990.100.220.28
9×90.9980.9991.0011.0021.0021.002−0.14−0.09−0.06
10×101.0001.0001.0001.0001.0001.0000.000.000.00
12×121.0111.0181.0071.0131.0141.015−0.62−0.72−0.77
15×151.0211.0381.0171.0291.0311.032−1.18−1.43−1.54
20×201.0291.0581.0281.0451.0491.051−1.61−2.07−2.24
25×251.0351.0811.0441.0661.0721.074−2.02−2.67−2.87
30×301.0371.0891.0501.0721.0811.083−2.13−2.94−3.14
35×301.0401.0911.0491.0741.0841.086−2.37−3.32−3.51
40×401.0401.1041.0621.0861.0981.100−2.33−3.42−3.59

μ has been extracted from Bjangard’s study;

μ has been extracted from Tahmasebi’s study

Figure 4

Phantom scatter factor (Sp) measured and calculated using various energy attenuation formula.

4. Discussion

The phantom scatter correction factor (Sp) accounts for the contribution of the radiation scattered in the phantom material to the dose at the point at depth dref on the central axis that in the reference irradiation set-up. The amount of scattered radiation reaching to the dosimeter (S′) depends on the volume of the phantom irradiated material, therefore, Sp is as a function of the field size at the surface of the phantom. There is little literature which has discussed analytical expression to calculate Sp. Blais and Karabrahimi in 2006 applied an equivalent field and Clarkson’s method to calculate collimator scatter factor and Sp for kilovoltage x-ray radiation fields. They reported that the percentage difference between calculated and measured Sp was 6.3% and 3.5% for the equivalent field and Clarkson’s method respectively(19). Chung et al. provided an analytical formalism to calculate phantom scatter factors for flattening filter free (FFF) mode photon beam by taking the product of the Sp in the presence of the flattening filter (FF) and the ratio of the off-axis ratio between the FFF mode and FF mode. They concluded that the percent difference between calculated and measured Sp ranged from −0.34% to 0.73% for Varian 6MV (20). This study has attempted to present an analytical approach to calculate the Sp using Scp values. It is most important to consider the percent difference range between the indirect measured and calculated Sp with 0.72%, 1.05% and 1.23% for 5×5 cm2 field size and −2.33%, −3.42% and −3.59% for 40×40 cm2 field size for μ=0, μ (r), and μ (d, L) respectively. As can be seen in the last two columns Table. 1, the Sp calculation shows approximately the same results for two expressions for μ. McKerracher showed indirect Sp values for fields >4 cm width are independent of beam defining system and dependent only on measurement depth, beam quality and beam area irradiated, and Sp in fields <4 cm width has also been shown to be independent of beam defining system and Linac design, and dependent only on measurement depth and beam area irradiated (6). One of the problems of indirect measurement method is measuring Sc for small fields due to the lack of the electron equilibrium. For small fields, this method is also insufficient due to the correction factor (α) will be close to 1, therefore, some other correction factors most be considered.
  13 in total

1.  A consistent formalism for the application of phantom and collimator scatter factors.

Authors:  J L Venselaar; J J van Gasteren; S Heukelom; H N Jager; B J Mijnheer; R van der Laarse; H J Kleffens; C F Westermann
Journal:  Phys Med Biol       Date:  1999-02       Impact factor: 3.609

2.  [Estimation of the phantom scatter factor (Sp) of rectangular fields].

Authors:  Takahide Hayakawa; Takumi Yamada; Hironori Sakai; Toshifumi Kasahara; Tomio Inoue; Michio Miyakawa
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2012

3.  Phantom scatter factors for small MV photon fields.

Authors:  Carolyn McKerracher; David I Thwaites
Journal:  Radiother Oncol       Date:  2007-12-03       Impact factor: 6.280

4.  Report of AAPM Therapy Physics Committee Task Group 74: in-air output ratio, Sc, for megavoltage photon beams.

Authors:  Timothy C Zhu; Anders Ahnesjö; Kwok Leung Lam; X Allen Li; Chang-Ming Charlie Ma; Jatinder R Palta; Michael B Sharpe; Bruce Thomadsen; Ramesh C Tailor
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

5.  Is there a need for a revised table of equivalent square fields for the determination of phantom scatter correction factors?

Authors:  J L Venselaar; S Heukelom; H N Jager; B J Mijnheer; J J van Gasteren; H J van Kleffens; R van der Laarse; C F Westermann
Journal:  Phys Med Biol       Date:  1997-12       Impact factor: 3.609

6.  A comparison of phantom scatter from flattened and flattening filter free high-energy photon beams.

Authors:  Neil Richmond; Vince Allen; Jim Daniel; Rob Dacey; Chris Walker
Journal:  Med Dosim       Date:  2014-11-11       Impact factor: 1.482

7.  An analytical formalism to calculate phantom scatter factors for flattening filter free (FFF) mode photon beams.

Authors:  Heeteak Chung; Karl L Prado; Byong Yong Yi
Journal:  Phys Med Biol       Date:  2014-02-07       Impact factor: 3.609

8.  An analytical method to calculate equivalent fields to irregular symmetric and asymmetric photon fields.

Authors:  Mohamad J Tahmasebi Birgani; Nahid Chegeni; Mansoor Zabihzadeh; Nima Hamzian
Journal:  Med Dosim       Date:  2014       Impact factor: 1.482

9.  Revision of tissue-maximum ratio and scatter-maximum ratio concepts for cobalt 60 and higher energy x-ray beams.

Authors:  F M Khan; W Sewchand; J Lee; J F Williamson
Journal:  Med Phys       Date:  1980 May-Jun       Impact factor: 4.071

10.  Determination of square equivalent field for rectangular field in electron therapy.

Authors:  Mohammad J Tahmasebi Birgani; Mohammad A Behrouz; Saeedeh Aliakbari; Seyed M Hosseini; Davood Khezerloo
Journal:  J Med Phys       Date:  2013-04
View more
  1 in total

1.  Interpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions.

Authors:  Parinaz Mehnati; Farideh Biglari; Ali Jomehzadeh
Journal:  J Med Signals Sens       Date:  2019 Apr-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.