Phantom scatter factors for small MV photon fields.

Carolyn McKerracher1, David I Thwaites.   

Abstract

BACKGROUND AND
PURPOSE: A table of small field (<4 cm width) S(p) values was created from measured S(cp) and S(c) data and compared with other authors' work. METHODS AND MATERIALS: S(cp) and S(c) values measured with three types of diode and a diamond were used to calculate S(p).
RESULTS: The results from measurements with different detectors were examined to determine the reasons for the differences and these were found to be due to detector and phantom-specific problems in the experimental determination of S(cp) and S(c). Small field S(p) were shown to be independent of beam defining system and linac design and dependent only on measurement depth and beam area irradiated. S(p) values obtained from S(cp) and S(c) measurements with a photon diode were found to be unreliable, due to problems with S(cp) measurements.
CONCLUSION: Although the electron and stereotactic diodes and the diamond are suitable detectors for S(cp) and S(c) measurements, the differences between S(p) values calculated from diamond measurements and those calculated from measurements with unshielded diodes still require clarification through Monte Carlo modelling. A final table of S(p) was produced, for fields >0.5 cm width.

Mesh:

Year:  2007        PMID: 18061694     DOI: 10.1016/j.radonc.2007.10.040

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  An Analytical Method to Calculate Phantom Scatter Factor for Photon Beam Accelerators.

Authors:  Mohammad Javad Tahmasebi Birgani; Nahid Chegeni; Mohammad Ali Behrooz; Marziyeh Bagheri; Amir Danyaei; Azin Shamsi
Journal:  Electron Physician       Date:  2017-01-25

2.  Detector-specific correction factors in radiosurgery beams and their impact on dose distribution calculations.

Authors:  Olivia A García-Garduño; Manuel A Rodríguez-Ávila; José M Lárraga-Gutiérrez
Journal:  PLoS One       Date:  2018-05-15       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.