Yaejin Moon1, Ryan S McGinnis2, Kirsten Seagers3, Robert W Motl4, Nirav Sheth3, John A Wright3, Roozbeh Ghaffari3, Jacob J Sosnoff1.
Abstract
BACKGROUND: Mobility impairment is common in people with multiple sclerosis (PwMS) and there is a need to assess mobility in remote settings. Here, we apply a novel wireless, skin-mounted, and conformal inertial sensor (BioStampRC, MC10 Inc.) to examine gait characteristics of PwMS under controlled conditions. We determine the accuracy and precision of BioStampRC in measuring gait kinematics by comparing to contemporary research-grade measurement devices.Entities:
Mesh:
Year: 2017 PMID: 28178288 PMCID: PMC5298289 DOI: 10.1371/journal.pone.0171346
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1BioStampRC a novel wireless, skin-mounted conformal inertial sensor.
Fig 2Location of BioStampRC, MTx, and GT3X sensors on the body.
Fig 3Schematic of shank angular velocity and gait events.
Demographics and clinical characteristics.
| Control | Mild MS | Moderate MS | Severe MS | |
|---|---|---|---|---|
| N | 15 (10F/ 5M) | 15 (12F/ 3M) | 15 (10F/ 5M) | 15 (10F/ 5M) |
| Age (yrs) | 57.9±12.9 | 53.7±12.3 | 59.7±8.3 | 61.1±8.3 |
| Height (cm) | 169.0±6.7 | 165.9±7.7 | 171.7±8.6 | 168.0±7.0 |
| Weight (kg) | 77.4±15.0 | 75.1±15.9 | 81.5±16.3 | 86.4±28.9 |
| MS duration (yrs) | -- | 14±8 | 18±6 | 21±10 |
| Assistive device (none/cane/ walker) | 15/0/0 | 15/0/0 | 7/7/1 | 4/8/3 |
| EDSSSR (Median(IQR)) | -- | 1.5 (0–2.5) | 6.0 (3.5–6) | 6.0 (5.5–6) |
| MSWS (Median(IQR)) | -- | 17 (14–20) | 43 (26–49) | 48 (43–55) |
| PDDS (Median(IQR)) | -- | 0 (0–1) | 4(3–5) | 5 (4–5) |
| Over-ground comfortable walking speed (m/s) | 1.28±0.18 | 1.28±0.17 | 0.86±0.10 | 0.60±0.12 |
Performance of clinical gait assessments as a function of groups.
| Test | Control | Mild MS | Moderate MS | Severe MS |
|---|---|---|---|---|
| T25W (sec) | 4.22±0.54 | 5.10±0.80 | 6.81±1.04 | 10.6±4.41 |
| TUG (sec) | 5.86±1.01 | 7.28±1.19 | 11.0±3.24 | 16.5±5.72 |
| 6MW_Comfortable (m) | 424±127 | 443±102 | 256±67 | 135±54 |
| 6MW_Slow (m) | 339±100 | 354±83 | 206±68 | 105±43 |
| 6MW_Fast (m) | 504±139 | 536±115 | 303±101 | 168±58 |
Note: significantly (p<0.05) different from
*the controls;
†the mild MS;
δ the moderate MS
Gait parameters recorded by BioStampRC as a function of group and speed.
| Test | Gait parameter | Control | Mild MS | Moderate MS | Severe MS |
|---|---|---|---|---|---|
| Over-ground comfortable walking | Stride time (ms) | 1083±111 | 1035±64 | 1239±101 | 1594±419 |
| Swing time (ms) | 445±40 | 437±29 | 467±40 | 521±87 | |
| Step time (ms) | 543±53 | 522±36 | 619±51 | 800±208 | |
| T25W | Stride time (ms) | 874±108 | 915±50 | 1079±121 | 1369±368 |
| Swing time (ms) | 383±43 | 399±30 | 430±34 | 478±85 | |
| Step time (ms) | 422±52 | 463±26 | 542±60 | 687±18 | |
| TUG | Stride time (ms) | 856±122 | 929±85 | 1082±172 | 1468±408 |
| Swing time (ms) | 378±51 | 395±41 | 427±43 | 481±92 | |
| Step time (ms) | 429±58 | 471±43 | 549±93 | 744±211 | |
| 6MW_Comfortable | Step number | 636±184 | 692±60 | 484±150 | 392±94 |
| Stride time (ms) | 1076±92 | 1047±96 | 1385±193 | 1857±40 | |
| Swing time (ms) | 461±33 | 447±29 | 498±57 | 534±87 | |
| Step time (ms) | 538±46 | 523±46 | 693±97 | 929±202 | |
| 6MW_Slow | Step number | 562±168 | 614±68 | 466±74 | 340±92 |
| Stride time (ms) | 1214±128 | 1184±141 | 1587±285 | 2233±600 | |
| Swing time (ms) | 486±38 | 482±41 | 543±76 | 574±101 | |
| Step time (ms) | 607±64 | 592±705 | 793±142 | 1047±414 | |
| 6MW_Fast | Step number | 680±204 | 736±78 | 548±110 | 380±150 |
| Stride time (ms) | 1009±109 | 987±103 | 1342±267 | 1668±596 | |
| Swing time (ms) | 440±34 | 425±41 | 495±60 | 546±95 | |
| Step time (ms) | 505±54 | 494±51 | 671±133 | 833±297 |
Note: Significantly (p<0.05) difference from
*the controls;
†the mild MS group;
δ the moderate MS group
Accuracy and precision of stride/step number count per device.
| Test | Device | Gait parameter | Absolute accuracy | Relative accuracy | Precision | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Medianerror | IQR | % error | 5%<N | 10%<N | 15%<N | ABS | REL | |||
| 6MW_Comfortable | BioStampRC | Step N diff | 2 | 2–4 | 0.8% | 1 | 0 | 0 | 3 | 1.2% |
| MTx | Step N diff | 2 | 2–4 | 0.9% | 0 | 0 | 0 | 3 | 1.2% | |
| GT3X | Step N diff | 10 | 3–48 | 10.1% | 19 | 14 | 11 | 78 | 17.0% | |
| 6MW_Slow | BioStampRC | Step N diff | 2 | 2–6 | 1.0% | 1 | 0 | 0 | 2 | 1.3% |
| MTx | Step N diff | 2 | 2–4 | 1.0% | 1 | 0 | 0 | 2 | 1.2% | |
| GT3X | Step N diff | 13 | 2–56 | 9.4% | 23 | 17 | 10 | 53 | 14.0% | |
| 6MW_Fast | BioStampRC | Step N diff | 4 | 2–6 | 0.9% | 1 | 0 | 0 | 3 | 1.2% |
| MTx | Step N diff | 4 | 2–6 | 0.9% | 1 | 0 | 0 | 2 | 1.0% | |
| GT3X | Step N diff | 6 | 2–20 | 6.1% | 13 | 8 | 7 | 57 | 12.7% | |
Note: Total N = 60 for each test, ABS = absolute, REL = relative
Accuracy and precision of temporal gait parameters measured by BioStampRC.
| Test | Gait parameter | Absolute accuracy | Relative accuracy | Precision | |||||
|---|---|---|---|---|---|---|---|---|---|
| Meanerror | 95% CI | Mean error | 5%<N | 10%<N | 15%<N | ABS | REL | ||
| Over-ground comfortable walking | Stride time diff (ms) | 8.9 | 5.2–12.6 | 0.6% | 0 | 0 | 0 | 12.0 | 0.9% |
| Swing time diff (ms) | 20.6 | 17.3–23.9 | 4.4% | 18 | 5 | 1 | 17.8 | 3.6% | |
| Step time diff (ms) | 7.1 | 4.4–9.7 | 1.2% | 1 | 0 | 0 | 9.1 | 1.5% | |
| T25W | Stride time diff (ms) | 8.3 | 6.0–10.5 | 0.8% | 0 | 0 | 0 | 11.8 | 1.3% |
| Swing time diff (ms) | 21 | 17.8–24.2 | 5.0% | 19 | 5 | 1 | 16.9 | 4.1% | |
| Step time diff (ms) | 7.4 | 5.7–9.2 | 1.4% | 1 | 0 | 0 | 9.4 | 1.4% | |
| TUG | Stride time diff (ms) | 12.6 | 9.7–15.5 | 1.2% | 1 | 0 | 0 | 14.9 | 1.2% |
| Swing time diff (ms) | 26.9 | 23.2–30.6 | 6.5% | 32 | 18 | 7 | 23.2 | 5.0% | |
| Step time diff (ms) | 14.1 | 11.4–16.8 | 2.6% | 7 | 0 | 0 | 12.5 | 2.1% | |
| 6MW_Comfortable | Stride time diff (ms) | 0.7 | 0.3–1.1 | 0.0% | 0 | 0 | 0 | 1.5 | 0.1% |
| Swing time diff (ms) | 29.8 | 24.7–35.0 | 6.3% | 31 | 9 | 1 | 19.0 | 3.9% | |
| Step time diff (ms) | 0.4 | 0.2–0.6 | 0.0% | 0 | 0 | 0 | 0.6 | 0.1% | |
| 6MW_Slow | Stride time diff (ms) | 1.6 | 0.4–2.8 | 0.0% | 0 | 0 | 0 | 4.5 | 0.2% |
| Swing time diff (ms) | 27.5 | 22.3–32.8 | 5.3% | 24 | 6 | 1 | 19.7 | 3.7% | |
| Step time diff (ms) | 0.9 | 0.4–1.3 | 0.0% | 0 | 0 | 0 | 1.8 | 0.2% | |
| 6MW_Fast | Stride time diff (ms) | 1.2 | 0.4–2.1 | 0.1% | 0 | 0 | 0 | 3.2 | 0.2% |
| Swing time diff (ms) | 29.7 | 24.3–35.0 | 6.1% | 30 | 9 | 0 | 20.0 | 3.8% | |
| Step time diff (ms) | 0.6 | 0.3–0.9 | 0.1% | 0 | 0 | 0 | 1.1 | 0.1% | |
Note: Total N = 60 for each test, ABS = absolute, REL = relative
Fig 4Bland-Altman plot of (a) Stride time (b) Swing time (c) Step time of BioStampRC compared with that of reference system (MTx).
Limits of agreement are specified as average difference (solid line) and ±1.96 standard deviation of the error (dotted line).