Literature DB >> 28168706

Flow-Field Simulations and Hemolysis Estimates for the Food and Drug Administration Critical Path Initiative Centrifugal Blood Pump.

Margaret L Heck1, Allen Yen1, Trevor A Snyder1,2, Edgar A O'Rear1,3, Dimitrios V Papavassiliou1.   

Abstract

The design of blood pumps for use in ventricular assist devices, which provide life-saving circulatory support in patients with heart failure, require remarkable precision and attention to detail to replicate the functionality of the native heart. The United States Food and Drug Administration (FDA) initiated a Critical Path Initiative to standardize and facilitate the use of computational fluid dynamics in the study and development of these devices. As a part of the study, a simplified centrifugal blood pump model generated by computer-aided design was released to universities and laboratories nationwide. The effects of changes in fluid rheology due to temperature, hematocrit, and turbulent flow on key metrics of the FDA pump were examined in depth using results from a finite volume-based commercial computational fluid dynamics code. Differences in blood damage indices obtained using Eulerian and Lagrangian formulations were considered. These results are presented and discussed awaiting future validation using experimental results, which will be released by the FDA at a future date.
© 2017 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Blood pump; Computational fluid dynamics; Damage index estimates; Hemolysis; Ventricular assist device

Mesh:

Year:  2017        PMID: 28168706     DOI: 10.1111/aor.12837

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  7 in total

1.  Computational modeling of the Food and Drug Administration's benchmark centrifugal blood pump.

Authors:  Bryan C Good; Keefe B Manning
Journal:  Artif Organs       Date:  2020-02-16       Impact factor: 3.094

2.  Hemodynamic evaluation and in vitro hemolysis evaluation of a novel centrifugal pump for extracorporeal membrane oxygenation.

Authors:  Minrui Fu; Gang Liu; Weining Wang; Bin Gao; Bingyang Ji; Yu Chang; Youjun Liu
Journal:  Ann Transl Med       Date:  2021-04

3.  Simulation, Fabrication and Analysis of Silver Based Ascending Sinusoidal Microchannel (ASMC) for Implant of Varicose Veins.

Authors:  Muhammad Javaid Afzal; Shahzadi Tayyaba; Muhammad Waseem Ashraf; M Khalid Hossain; M Jalal Uddin; Nitin Afzulpurkar
Journal:  Micromachines (Basel)       Date:  2017-09-14       Impact factor: 2.891

4.  Sinusoidal Microchannel with Descending Curves for Varicose Veins Implantation.

Authors:  Muhammad Javaid Afzal; Muhammad Waseem Ashraf; Shahzadi Tayyaba; M Khalid Hossain; Nitin Afzulpurkar
Journal:  Micromachines (Basel)       Date:  2018-01-31       Impact factor: 2.891

5.  Haemolysis induced by mechanical circulatory support devices: unsolved problems.

Authors:  Inge Köhne
Journal:  Perfusion       Date:  2020-06-23       Impact factor: 1.972

6.  Distribution and history of extensional stresses on vWF surrogate molecules in turbulent flow.

Authors:  Oanh L Pham; Samuel E Feher; Quoc T Nguyen; Dimitrios V Papavassiliou
Journal:  Sci Rep       Date:  2022-01-07       Impact factor: 4.996

7.  Computations of the shear stresses distribution experienced by passive particles as they circulate in turbulent flow: A case study for vWF protein molecules.

Authors:  Oanh L Pham; Samuel E Feher; Quoc T Nguyen; Dimitrios V Papavassiliou
Journal:  PLoS One       Date:  2022-08-29       Impact factor: 3.752

  7 in total