| Literature DB >> 28153281 |
Hunho Jo1, Jin Her2, Heehyun Lee3, Yoon-Bo Shim4, Changill Ban5.
Abstract
In this study, we developed a sandwich aptamer-based screen-printed carbon electrode (SPCE) using chronoamperometry for the detection of cardiac troponin I (cTnI), one of the promising biomarkers for acute myocardial infarction (AMI). Disposable three-electrode SPCEs were manufactured using a screen printer, and various modifications such as electrodeposition of gold nanoparticles and electropolymerization of conductive polymers were performed. From the bare electrode to the aptamer-immobilized SPCE, all processes were monitored and analyzed via various techniques such as cyclic voltammetry, electrochemical impedance spectroscopy, and X-ray photoelectron spectroscopy. The quantification of cTnI was conducted based on amperometric signals from the catalytic reaction between hydrazine and H2O2. The fabricated aptasensor in a buffer, as well as in a serum-added solution, exhibited great analytical performance with a dynamic range of 1-100 pM (0.024-2.4ng/mL) and a detection limit of 1.0 pM (24pg/mL), which is lower than the existing cutoff values (40-700pg/mL). Furthermore, the developed sensor showed high sensitivity to cTnI over other proteins. It is anticipated that this potable SPCE aptasensor for cTnI will become an innovative diagnostic tool for AMI.Entities:
Keywords: Aptamer; Cardiac troponin I; Chronoamperometry; Diagnosis; Hydrazine; Screen-printed carbon electrode
Mesh:
Substances:
Year: 2016 PMID: 28153281 DOI: 10.1016/j.talanta.2016.12.091
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057