A novel cuffless device for self-measurement of blood pressure: concept, performance and clinical validation.

N Boubouchairopoulou1, A Kollias1, B Chiu2, B Chen2, S Lagou1, P Anestis1, G S Stergiou1.   

Abstract

A pocket-size cuffless electronic device for self-measurement of blood pressure (BP) has been developed (Freescan, Maisense Inc., Zhubei, Taiwan). The device estimates BP within 10 s using three embedded electrodes and one force sensor that is applied over the radial pulse to evaluate the pulse wave. Before use, basic anthropometric characteristics are recorded on the device, and individualized initial calibration is required based on a standard BP measurement performed using an upper-arm BP monitor. The device performance in providing valid BP readings was evaluated in 313 normotensive and hypertensive adults in three study phases during which the device sensor was upgraded. A formal validation study of a prototype device against mercury sphygmomanometer was performed according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 2013 protocol. The test device succeeded in obtaining a valid BP measurement (three successful readings within up to five attempts) in 55-72% of the participants, which reached 87% with device sensor upgrade. For the validation study, 125 adults were recruited and 85 met the protocol requirements for inclusion. The mean device-observers BP difference was 3.2±6.7 (s.d.) mm Hg for systolic and 2.6±4.6 mm Hg for diastolic BP (criterion 1). The estimated s.d. (inter-subject variability) were 5.83 and 4.17 mm Hg respectively (criterion 2). These data suggest that this prototype cuffless BP monitor provides valid self-measurements in the vast majority of adults, and satisfies the BP measurement accuracy criteria of the ANSI/AAMI/ISO 2013 validation protocol.

Entities:  

Mesh:

Year:  2017        PMID: 28124684     DOI: 10.1038/jhh.2016.101

Source DB:  PubMed          Journal:  J Hum Hypertens        ISSN: 0950-9240            Impact factor:   3.012


  5 in total

Review 1.  Self blood pressure monitoring at home by wrist devices: a reliable approach?

Authors:  Gianfranco Parati; Roland Asmar; George S Stergiou
Journal:  J Hypertens       Date:  2002-04       Impact factor: 4.844

2.  An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices.

Authors:  E O'Brien; J Petrie; W Littler; M de Swiet; P L Padfield; D G Altman; M Bland; A Coats; N Atkins
Journal:  J Hypertens       Date:  1993-06       Impact factor: 4.844

3.  European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring.

Authors:  Gianfranco Parati; George S Stergiou; Roland Asmar; Grzegorz Bilo; Peter de Leeuw; Yutaka Imai; Kazuomi Kario; Empar Lurbe; Athanasios Manolis; Thomas Mengden; Eoin O'Brien; Takayoshi Ohkubo; Paul Padfield; Paolo Palatini; Thomas Pickering; Josep Redon; Miriam Revera; Luis M Ruilope; Andrew Shennan; Jan A Staessen; Andras Tisler; Bernard Waeber; Alberto Zanchetti; Giuseppe Mancia
Journal:  J Hypertens       Date:  2008-08       Impact factor: 4.844

4.  Can validated wrist devices with position sensors replace arm devices for self-home blood pressure monitoring? A randomized crossover trial using ambulatory monitoring as reference.

Authors:  George S Stergiou; George R Christodoulakis; Efthimia G Nasothimiou; Periklis P Giovas; Petros G Kalogeropoulos
Journal:  Am J Hypertens       Date:  2008-04-24       Impact factor: 2.689

5.  Call to action on use and reimbursement for home blood pressure monitoring: executive summary: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association.

Authors:  Thomas G Pickering; Nancy Houston Miller; Gbenga Ogedegbe; Lawrence R Krakoff; Nancy T Artinian; David Goff
Journal:  Hypertension       Date:  2008-05-22       Impact factor: 10.190

  5 in total
  10 in total

Review 1.  Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals.

Authors:  Ramakrishna Mukkamala; Mohammad Yavarimanesh; Keerthana Natarajan; Jin-Oh Hahn; Konstantinos G Kyriakoulis; Alberto P Avolio; George S Stergiou
Journal:  Hypertension       Date:  2021-09-13       Impact factor: 10.190

Review 2.  Cuffless Blood Pressure Devices.

Authors:  Corey K Bradley; Daichi Shimbo; David Alexander Colburn; Daniel N Pugliese; Raj Padwal; Samuel K Sia; D Edmund Anstey
Journal:  Am J Hypertens       Date:  2022-05-10       Impact factor: 3.080

3.  Diagnostic accuracy of a novel cuffless self-blood pressure monitor for atrial fibrillation screening in the elderly.

Authors:  Konstantinos G Kyriakoulis; Anastasios Kollias; Ioannis Anagnostopoulos; Areti Gravvani; Petros Kalogeropoulos; Antonios Destounis; George S Stergiou
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-11-19       Impact factor: 3.738

4.  The first study comparing a wearable watch-type blood pressure monitor with a conventional ambulatory blood pressure monitor on in-office and out-of-office settings.

Authors:  Kazuomi Kario; Daichi Shimbo; Naoko Tomitani; Hiroshi Kanegae; Joseph E Schwartz; Bryan Williams
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-01-19       Impact factor: 3.738

5.  Improvement of Diurnal Blood Pressure Variation by Azilsartan.

Authors:  Keisuke Okamura; Kazuyuki Shirai; Tetsu Okuda; Hidenori Urata
Journal:  J Clin Med Res       Date:  2017-12-01

6.  Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device.

Authors:  Naoki Watanabe; Yasuko K Bando; Taiji Kawachi; Hiroshi Yamakita; Kouki Futatsuyama; Yoshikazu Honda; Hisae Yasui; Kazuyuki Nishimura; Takahiro Kamihara; Takahiro Okumura; Hideki Ishii; Takahisa Kondo; Toyoaki Murohara
Journal:  JACC Basic Transl Sci       Date:  2017-12-25

7.  Feasibility of a New Cuffless Device for Ambulatory Blood Pressure Measurement in Patients With Hypertension: Mixed Methods Study.

Authors:  Paula Am Ogink; Jelske M de Jong; Mats Koeneman; Mariska Weenk; Lucien Jlpg Engelen; Harry van Goor; Tom H van de Belt; Sebastian Jh Bredie
Journal:  J Med Internet Res       Date:  2019-06-19       Impact factor: 5.428

8.  An Arterial Compliance Sensor for Cuffless Blood Pressure Estimation Based on Piezoelectric and Optical Signals.

Authors:  Cheng-Yan Guo; Hao-Ching Chang; Kuan-Jen Wang; Tung-Li Hsieh
Journal:  Micromachines (Basel)       Date:  2022-08-16       Impact factor: 3.523

9.  A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices.

Authors:  Sarah Melville; Robert Teskey; Shona Philip; Jeremy A Simpson; Sohrab Lutchmedial; Keith R Brunt
Journal:  J Med Internet Res       Date:  2018-04-25       Impact factor: 5.428

10.  Blood Pressure Measurement Validation Off the Cuff? Comment on "A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study".

Authors:  Noud van Helmond; Timothy B Plante; Jeffrey I Joseph
Journal:  J Med Internet Res       Date:  2018-10-22       Impact factor: 5.428

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.