| Literature DB >> 28092855 |
Lyndsey White1, Jing Ma1, Su Liang1, Beatriz Sanchez-Espiridion2, Dong Liang3.
Abstract
Several metabolites in human serum have been identified as potential cancer biomarkers for early detection. This study focuses on the LC-MS/MS method development and validation of d-mannose in human serum. Surrogate blank serum, coupled with stable isotope d-mannose-13C6, as internal standard, was used for generating standard curves ranging from 1 to 50μg/mL. Separation was achieved by an Agilent 1200 series HPLC equipped with a SUPELCOGELTM Pb, 6% Crosslinked column with HPLC water as a mobile phase at flow rate of 0.5mL/min at 80°C. Mass detection was performed under negative ionization electrospray. Inter- and intra-day accuracy and precision were <2%. The extraction recovery and matrix effect were 104.1%-105.5% and 97.0%-100.0%, respectively. This method was successfully applied for the quantification of d-mannose in the serum samples of 320 esophageal cancer patients and 323 healthy volunteers. We report a simple, specific and reproducible LC-MS/MS method for the quantification of d-mannose in human serum as a potential cancer biomarker.Entities:
Keywords: Biomarker; Esophageal cancer; LC–MS/MS; Mannose; Serum
Mesh:
Substances:
Year: 2016 PMID: 28092855 PMCID: PMC5924413 DOI: 10.1016/j.jpba.2016.12.017
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935