Evaluation of artifacts generated by zirconium implants in cone-beam computed tomography images.

Taruska Ventorini Vasconcelos1, Boulos B Bechara2, Clyde Alex McMahan3, Deborah Queiroz Freitas4, Marcel Noujeim2.   

Abstract

OBJECTIVE: To evaluate zirconium implant artifact production in cone beam computed tomography images obtained with different protocols. STUDY
DESIGN: One zirconium implant was inserted in an edentulous mandible. Twenty scans were acquired with a ProMax 3D unit (Planmeca Oy, Helsinki, Finland), with acquisition settings ranging from 70 to 90 peak kilovoltage (kVp) and voxel sizes of 0.32 and 0.16 mm. A metal artifact reduction (MAR) tool was activated in half of the scans. An axial slice through the middle region of the implant was selected for each dataset. Gray values (mean ± standard deviation) were measured in two regions of interest, one close to and the other distant from the implant (control area). The contrast-to-noise ratio was also calculated.
RESULTS: Standard deviation decreased with greater kVp and when the MAR tool was used. The contrast-to-noise ratio was significantly higher when the MAR tool was turned off, except for low resolution with kVp values above 80. Selection of the MAR tool and greater kVp resulted in an overall reduction of artifacts in images acquired with low resolution.
CONCLUSIONS: Although zirconium implants do produce image artifacts in cone-bean computed tomography scans, the setting that best controlled artifact generation by zirconium implants was 90 kVp at low resolution and with the MAR tool turned on.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 28086998     DOI: 10.1016/j.oooo.2016.10.021

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  16 in total

1.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

2.  Effect of anatomical region on the formation of metal artefacts produced by dental implants in cone beam computed tomographic images.

Authors:  Alessiana Helena Machado; Karolina Aparecida Castilho Fardim; Camila Furtado de Souza; Bruno Salles Sotto-Maior; Neuza Maria Souza Picorelli Assis; Karina Lopes Devito
Journal:  Dentomaxillofac Radiol       Date:  2018-01-10       Impact factor: 2.419

3.  Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.

Authors:  Rocharles C Fontenele; Eduarda Hl Nascimento; Taruska V Vasconcelos; Marcel Noujeim; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2018-05-10       Impact factor: 2.419

4.  Are metal artefact reduction algorithms effective to correct cone beam CT artefacts arising from the exomass?

Authors:  Amanda Pelegrin Candemil; Benjamin Salmon; Deborah Queiroz Freitas; Glaucia Maria Bovi Ambrosano; Francisco Haiter-Neto; Matheus Lima Oliveira
Journal:  Dentomaxillofac Radiol       Date:  2019-01-28       Impact factor: 2.419

5.  Metal artifact reduction using common dental materials.

Authors:  Nicole V Hinchy; Nina K Anderson; Mina Mahdian
Journal:  Dentomaxillofac Radiol       Date:  2021-08-18       Impact factor: 2.419

6.  Difference in the artefacts production and the performance of the metal artefact reduction (MAR) tool between the buccal and lingual cortical plates adjacent to zirconium dental implant.

Authors:  Eduarda Helena Leandro Nascimento; Rocharles Cavalcante Fontenele; Gustavo Machado Santaella; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2019-07-10       Impact factor: 2.419

7.  Evaluation of soft tissues simulant materials in cone beam computed tomography.

Authors:  Priscila A Lopes; Gustavo M Santaella; Carlos Augusto S Lima; Karla de Faria Vasconcelos; Francisco C Groppo
Journal:  Dentomaxillofac Radiol       Date:  2018-07-20       Impact factor: 2.419

8.  Quantitative assessment of artefacts and identification of gaps in prosthetic crowns: a comparative in vitro study between periapical radiography and CBCT images.

Authors:  Letícia Queiroz Mauad; Paulo Victor Teixeira Doriguêtto; Daniela de Almeida; Karolina Aparecida Castilho Fardim; Alessiana Helena Machado; Karina Lopes Devito
Journal:  Dentomaxillofac Radiol       Date:  2020-09-17       Impact factor: 2.419

9.  CBCT image artefacts generated by implants located inside the field of view or in the exomass.

Authors:  Husniye Demirturk Kocasarac; Lisa J Koenig; Gulbahar Ustaoglu; Matheus Lima Oliveira; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2021-07-29       Impact factor: 2.419

10.  Influence of windowing and metal artefact reduction algorithms on the volumetric dimensions of five different high-density materials: a cone-beam CT study.

Authors:  Fernanda Coelho-Silva; Luciano Augusto Cano Martins; Daniela Azeredo Braga; Eliana Zandonade; Francisco Haiter-Neto; Sergio Lins de-Azevedo-Vaz
Journal:  Dentomaxillofac Radiol       Date:  2020-06-25       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.