Prospective comparison of dual-energy CT aortography using 70% reduced iodine dose versus single-energy CT aortography using standard iodine dose in the same patient.

William P Shuman1, Ryan B O'Malley2, Janet M Busey2, Mario M Ramos2, Kent M Koprowicz2.   

Abstract

PURPOSE: To compare dual-energy computed tomography (DECT) aortography using a 70% reduced iodine dose to single-energy CT (SECT) aortography using a standard iodine dose in the same patient.
METHODS: Twenty-one patients with a prior SECT aortography using standard iodine dose had DECT aortography using 70% reduced iodine dose. Section 120 kVp images were compared to DECT images reconstructed at both 50 and 77 keV. Reviewers measured image noise and attenuation in the aorta at eight locations from proximal to distal and subjectively scored vascular enhancement on a four-point scale. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The volume CT dose index (CTDIvol) for each exam was recorded.
RESULTS: Mean iodine dose was 50 g for SECT and 15 g for DECT (70% reduction). Mean aortic attenuation was similar for section 120 kVp (350 ± 67 HU) and DECT 50 keV (338 ± 57 HU, p = 0.547) but was lower at 77 keV (152 ± 23 HU). Measured image noise was greatest at 50 keV (12 ± 5 HU) and was lowest at 77 keV (7 ± 2 HU, p > 0.001). There was no difference in SNR or CNR between 120 kVp and 50 keV (p > 0.05). Mean subjective vascular enhancement scores for SECT were between good and excellent (3.33-3.69), and for DECT at 50 keV were between moderate and good (2.54-2.93, p < 0.0001). CTDIvol was 13.6 mGy for SECT and 13.1 mGy for DECT (p = 0.637).
CONCLUSION: 70% Reduced iodine DECT aortography may result in similar aortic attenuation, CNR, SNR, and lower although acceptable subjective image scores when compared to standard iodine SECT aortography in the same patient.

Entities:  

Keywords:  Aortography; CT angiography; Computed tomography; Contrast media; Dual energy; Iodine dose; Renal dysfunction

Mesh:

Substances:

Year:  2017        PMID: 28084544     DOI: 10.1007/s00261-016-1041-z

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  10 in total

Review 1.  Acute aortic syndromes and aortic emergencies.

Authors:  Vinit Baliyan; Anushri Parakh; Anand M Prabhakar; Sandeep Hedgire
Journal:  Cardiovasc Diagn Ther       Date:  2018-04

2.  Aortic CT angiography using the double region of interest timing bolus technique: feasibility of 80 kVp scanning in lean patients.

Authors:  Nobuo Tomizawa; Shingo Ito; Tastuya Nakao; Hiroaki Arakawa; Kodai Yamamoto; Shinichi Inoh; Takeshi Nojo; Sunao Nakamura
Journal:  Int J Cardiovasc Imaging       Date:  2019-07-02       Impact factor: 2.357

3.  Contrast agent concentration optimization in CTA using low tube voltage and dual-energy CT in multiple vendors: a phantom study.

Authors:  Robbert W van Hamersvelt; Nienke G Eijsvoogel; Casper Mihl; Pim A de Jong; Arnold M R Schilham; Nico Buls; Marco Das; Tim Leiner; Martin J Willemink
Journal:  Int J Cardiovasc Imaging       Date:  2018-03-08       Impact factor: 2.357

4.  Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom.

Authors:  Domenico De Santis; Damiano Caruso; U Joseph Schoepf; Marwen Eid; Moritz H Albrecht; Taylor M Duguay; Akos Varga-Szemes; Andrea Laghi; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2018-02-27       Impact factor: 5.315

5.  Iodine quantification and detectability thresholds among major dual-energy CT platforms.

Authors:  Ross Edward Taylor; Pamela Mager; Nam C Yu; David P Katz; Jett R Brady; Nakul Gupta
Journal:  Br J Radiol       Date:  2019-10-07       Impact factor: 3.039

6.  Optimizing low contrast volume thoracic CT angiography: From the basics to the advanced.

Authors:  Katherine Kaproth-Joslin; Susan Hobbs; Prabhakar Rajiah; Apeksha Chaturvedi; Abhishek Chaturvedi
Journal:  J Clin Imaging Sci       Date:  2022-07-29

7.  The optimal monoenergetic spectral image level of coronary computed tomography (CT) angiography on a dual-layer spectral detector CT with half-dose contrast media.

Authors:  Xin Huang; Sizhe Gao; Yue Ma; Xiaomei Lu; Zheng Jia; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2020-03

8.  Advantages and disadvantages of single-source dual-energy whole-body CT angiography with 50% reduced iodine dose at 40 keV reconstruction.

Authors:  Yoshifumi Noda; Fumihiko Nakamura; Noriyuki Yasuda; Toshiharu Miyoshi; Nobuyuki Kawai; Hiroshi Kawada; Fuminori Hyodo; Masayuki Matsuo
Journal:  Br J Radiol       Date:  2021-02-22       Impact factor: 3.039

9.  Two Small Intravenous Catheters for High-Rate Contrast Medium Injection for Computed Tomography in Patients Lacking Superficial Veins to Accommodate a Large Catheter.

Authors:  Bum Gu Son; Min Jung Kim; Myeung Hwa Park; Kyoungsook Kim; Jiyu Kim; Se-Young Kim; Kyung Jin Lee; Sang Hyun Choi; Ah Young Kim; Seong Ho Park
Journal:  Korean J Radiol       Date:  2018-04-06       Impact factor: 3.500

10.  Evaluating the Image Quality of Monoenergetic Images From Dual-Energy Computed Tomography With Low-Concentration and Low-Flow-Rate Contrast Media for the Arterials Supply to the Nipple-Areola Complex in Breast Cancer Compared With Conventional Computed Tomography Angiography.

Authors:  Xiangfei Zeng; Xiaoxia Wang; Huifang Chen; Daihong Liu; Jinfang Shi; Jing Zhang; Xiaoqin Li; Liyin Zhang; Yi Yang; Jiuquan Zhang
Journal:  J Comput Assist Tomogr       Date:  2020 Nov/Dec       Impact factor: 2.081

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.