Carla Alves de Oliveira1, Maria do Carmo Lopes1, Armanda Matos1.
Abstract
PURPOSE: To describe the tests that have been performed in order to commission the Brachytherapy module, version 3.2, service pack 3.0, of the Oncentra MasterPlan treatment plan system (OB), from Nucletron. The results were benchmarked against those obtained with the Plato system, v 14.3.7, also from Nucletron, used in the clinical routine.Entities:
Keywords: Oncentra Brachy; TPS commissioning; dose calculations
Year: 2010 PMID: 28050176 PMCID: PMC5086490
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1AOne catheter defined along the Y axis and centred with the XYZ coordinate system. The dwell position that was activated is located at the (0, 0, 0) position. A2 with coordinates (-0.5, 0, 0) corresponds to the normalization point. A1, A3, A4 e A5 have, respectively, the following coordinates: (-1.0, 0, 0), (-0.5, 0.5, 0), (-0.5, 1.0, 0) and (0, -0.5, 0); B. One catheter aligned through the Y axis and centred with the XYZ coordinate system. Two dwell positions were activated, respectively, at (0, -1.0, 0) and (0, 1.0, 0) coordinates. A2 with coordinates (-0.5, 0, 0) corresponds to the normalization point. A1, A3, and A4 have, respectively, the following coordinates (0, 0, 0), (-1.0, 1.0, 0) and (0, 1.5, 0); C. Two catheters aligned parallel to the Y axis and with a 1 cm separation between them. A3 with coordinates (-1.0, 0, 0) corresponds to the normalization point. A1, A2, A4 e A5 have, respectively, the following coordinates: (0, 0, 0), (0, 0.5, 0), (-1.0, 0.5, 0) and (-0.5, -1.0, 0)
A. Point dose calculations and percentage standard deviations for catheter geometry shown in Fig. 1A, using the Plato, OB TPS and manual calculations
| Points to Dose Calculation | PLATO | OB | Manual Calculations | δ (%) |
|---|---|---|---|---|
| D (Gy) | ||||
| A1 | 0.2577 | 0.2577 | 0.2553 | 0.53 |
| A2 | 1.000 | 1.000 | 0.9909 | 0.53 |
| A3 | 0.5085 | 0.5088 | 0.5074 | 0.15 |
| A4 | 0.1879 | 0.1880 | 0.1859 | 0.64 |
| A5 | 0.7285 | 0.7285 | 0.7223 | 0.49 |
Global and partial treatment times for an endovaginal gynaecological treatment, planned in Plato and OB. Within Plato these treatment plans are optimized on dose points and on distance, using a value of 0.5 for the DTGR parameter (results shown on the left). In order to reproduce in OB the same time pattern a slightly different value of OB had to be used (results shown on the right)
| DTGR = 0.5 | DTGR = 0.452 | ||||
|---|---|---|---|---|---|
| Plato | OB | ∆ (%) | OB | ∆ (%) | |
| 657.1 | 649.9 | -1.1 | 656.80 | -0.05 | |
|
|
| ||||
|
| 122.5 | 116.2 | -5.14 | 122.5 | 0.00 |
|
| 98.5 | 95.2 | -3.05 | 98.5 | 0.00 |
|
| 65.5 | 67.0 | 2.29 | 65.6 | 0.15 |
|
| 42.6 | 47.0 | 10.33 | 42.6 | 0.00 |
|
| 42.5 | 46.9 | 10.35 | 42.6 | 0.00 |
|
| 65.3 | 66.6 | 1.99 | 65.2 | -0.15 |
|
| 98.1 | 95.0 | -3.16 | 98.0 | -0.10 |
|
| 122.1 | 115.7 | -5.24 | 121.9 | -0.16 |
Fig. 2Dose on Axis Points, defined on applicator surface at every activated dwell position, calculated both with Plato, DTGR = 0.5, and OB both with DTGR = 0.5 and DTGR = 0.452
Fig. 3CTV calculated volumes in Plato and OB. The percentage differences, considering the values calculated by Plato as reference, are also shown
Fig. 4Coverage index, CI, calculated for each treatment plan planned in Plato and OB according to the local dosimetry protocol
Fig. 6Overdosage index, OI, calculated for each treatment plan planned in Plato and OB according to the local dosimetry protocol
Fig. 7Coverage index, CI, calculated for each treatment plan planned in Plato and OB, using the IPSA algorithm
Fig. 8Conformal index, COIN, calculated for each treatment plan planned in Plato and OB, using the IPSA algorithm
Fig. 9Overdosage index, OI, calculated for each treatment plan planned in Plato and OB, using the IPSA algorithm
B. Point dose calculation and percentage standard deviations for catheter geometry shown in Fig. 1B, using the Plato, OB TPS and manual calculations
| Points to Dose Calculation | PLATO | OB | Manual Calculations | δ (%) |
|---|---|---|---|---|
| D (Gy) | ||||
| A1 | 0.8717 | 0.8709 | 0.8587 | 0.84 |
| A2 | 1.0000 | 1.0000 | 1.0035 | 0.20 |
| A3 | 0.8112 | 0.8104 | 0.8099 | 0.08 |
| A4 | 2.1988 | 2.1967 | 2.1947 | 0.09 |
C. Point dose calculation and percentage standard deviations for catheter geometry shown in Fig. 1C, using the Plato, OB TPS and manual calculations
| Points to Dose Calculation | PLATO | OB | Manual Calculations | δ (%) |
|---|---|---|---|---|
| D (Gy) | ||||
| A1 | 1.7928 | 1.7928 | 1.7928 | 0.00 |
| A2 | 0.9117 | 0.9121 | 0.9121 | 0.03 |
| A3 | 1.000 | 1.000 | 1.000 | 0.00 |
| A4 | 0.5489 | 0.5492 | 0.5492 | 0.03 |
| A5 | 0.2542 | 0.2543 | 0.2543 | 0.02 |