Literature DB >> 27994689

Random Assay in Radioimmunoassay: Feasibility and Application Compared with Batch Assay.

Jung Min Lee1, Hwan Hee Lee1, Sohyun Park1, Tae Sung Kim2, Seok-Ki Kim2.   

Abstract

PURPOSE: The batch assay has been conventionally used for radioimmunoassay (RIA) because of its technical robustness and practical convenience. However, it has limitations in terms of the relative lag of report time due to the necessity of multiple assays in a small number of samples compared with the random assay technique. In this study, we aimed to verify whether the random assay technique can be applied in RIA and is feasible in daily practice.
METHODS: The coefficients of variation (CVs) of eight standard curves within a single kit were calculated in a CA-125 immunoradiometric assay (IRMA) for the reference of the practically ideal CV of the CA-125 kit. Ten standard curves of 10 kits from 2 prospectively collected lots (pLot) and 85 standard curves of 85 kits from 3 retrospectively collected lots (Lot) were obtained. Additionally, the raw measurement data of both 170 control references and 1123 patients' sera were collected retrospectively between December 2015 and January 2016. A standard curve of the first kit of each lot was used as a master standard curve for a random assay. The CVs of inter-kits were analyzed in each lot, respectively. All raw measurements were normalized by decay and radioactivity. The CA-125 values from control samples and patients' sera were compared using the original batch assay and random assay.
RESULTS: In standard curve analysis, the CVs of inter-kits in pLots and Lots were comparable to those within a single kit. The CVs from the random assay with normalization were similar to those from the batch assay in the control samples (CVs % of low/high concentration; Lot1 2.71/1.91, Lot2 2.35/1.83, Lot3 2.83/2.08 vs. Lot1 2.05/1.21, Lot2 1.66/1.48, Lot3 2.41/2.14). The ICCs between the batch assay and random assay using patients' sera were satisfactory (Lot1 1.00, Lot2 0.999, Lot3 1.00).
CONCLUSION: The random assay technique could be successfully applied to the conventional CA-125 IRMA kits. The random assay showed strong agreement with the batch assay. The random assay procedure could increase the flexibility and decrease the turnaround time of the radioimmunoassay technique.

Entities:  

Keywords:  Batch assay; CA-125; Immunoradiometric assay; Radioimmunoassay; Random assay; Standard curve

Year:  2016        PMID: 27994689      PMCID: PMC5135698          DOI: 10.1007/s13139-016-0436-7

Source DB:  PubMed          Journal:  Nucl Med Mol Imaging        ISSN: 1869-3474


  11 in total

Review 1.  Quality requirements and control: EGTM recommendations. European Group on Tumour Markers.

Authors: 
Journal:  Anticancer Res       Date:  1999 Jul-Aug       Impact factor: 2.480

2.  Radioimmunoassay: review of basic principles.

Authors:  S J Goldsmith
Journal:  Semin Nucl Med       Date:  1975-04       Impact factor: 4.446

3.  Radioimmunoassay and saturation analysis. The immunoradiometric assay and related techniques.

Authors:  J S Woodhead; G M Addison; C N Hales
Journal:  Br Med Bull       Date:  1974-01       Impact factor: 4.291

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Determination of prostate-specific antigen in serum by immunoradiometric assay.

Authors:  G Lindstedt; A Jacobsson; P A Lundberg; H Hedelin; S Pettersson; B Unsgaard
Journal:  Clin Chem       Date:  1990-01       Impact factor: 8.327

6.  Evaluation of the utility of a radioimmunoassay for serum CA 19-9 levels in patients before and after treatment of carcinoma of the pancreas.

Authors:  J Glenn; W M Steinberg; S H Kurtzman; S M Steinberg; W F Sindelar
Journal:  J Clin Oncol       Date:  1988-03       Impact factor: 44.544

7.  Predictive value of CA 125 antigen levels in second-look procedures for ovarian cancer.

Authors:  J M Niloff; R C Bast; E M Schaetzl; R C Knapp
Journal:  Am J Obstet Gynecol       Date:  1985-04-01       Impact factor: 8.661

8.  Immunoradiometric assay of carcinoembryonic antigen with use of avidin-biotin labeling.

Authors:  J Peters; H Schmidt-Gayk; B Peters; F P Armbruster; A Quentmeier; D Mathias
Journal:  Clin Chem       Date:  1989-04       Impact factor: 8.327

9.  Comparison of radioimmunoassay and immunoradiometric assay for serum prostatic acid phosphatase.

Authors:  C M Sturgeon; A J Hussey; L Beynon; A Ritchie; G D Chisholm; L Micklem; K James; J Seth
Journal:  Clin Chim Acta       Date:  1986-11-30       Impact factor: 3.786

Review 10.  Statistical methods used to test for agreement of medical instruments measuring continuous variables in method comparison studies: a systematic review.

Authors:  Rafdzah Zaki; Awang Bulgiba; Roshidi Ismail; Noor Azina Ismail
Journal:  PLoS One       Date:  2012-05-25       Impact factor: 3.240

View more
  1 in total

1.  Survey Result of the Counting Efficiency of Gamma Counter by Certified Reference Materials.

Authors:  Ji-In Bang; Ho-Young Lee; Geyoung Woon Noh; Jeong Mi Park; June Kee Yoon
Journal:  Nucl Med Mol Imaging       Date:  2018-03-14
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.