The measurement repeatability using different partition methods of intraretinal tomographic thickness maps in healthy human subjects.

Jia Tan1, Ye Yang2, Hong Jiang3, Che Liu4, Zhihong Deng1, Byron L Lam4, Liang Hu5, Jonathan Oakley6, Jianhua Wang4.   

Abstract

PURPOSE: To determine the repeatability and profiles with different partition methods in intraretinal thickness layers in healthy human subjects, using optical coherence tomography (OCT).
METHODS: A custom-built ultrahigh-resolution OCT was used to acquire three-dimensional volume of the macula in 20 healthy subjects. The dataset was acquired twice using the macular cube 512×128 protocol in an area of 6×6 mm2 centered on the fovea. Commercially available segmentation software (Orion™) was used to segment the dataset into thickness maps of six intraretinal layers. The coefficient of repeatability and intraclass coefficient of correlation (ICC) were analyzed using hemispheric zoning and sectors defined by the Early Treatment Diabetic Retinopathy Study (ETDRS).
RESULTS: All datasets were successfully segmented to create six thickness maps of individual intraretinal layers. Coefficients of repeatabilities of these layers in hemispheric zones ranged from 0.9 to 6.6 µm, with an average of 3.6 µm (standard deviation [SD] 1.4), which was not significantly different compared to ETDRS sectors (P>0.05). ICCs of these layers in hemispheric zones ranged from 0.68 to 0.99, with an average of 0.91 (SD 0.07). There were no significant differences in ICCs between two zoning methods (P>0.05). Significant variations of tomographic intraretinal thicknesses were found between the inner and outer annuli and among the quadrantal sectors within the inner and outer annuli (P<0.05). Significant variations of the quadrantal sectors including both inner and outer annuli were evident in intraretinal layers (P<0.05) except for the outer plexiform layer.
CONCLUSION: The measurement repeatabilities of tomographic thicknesses of intraretinal layers are comparable using both hemispheric and ETDRS partitions in volumetric data combined with the commercially available segmentation software. In keeping with known, normal anatomical variation, significant differences in tomographic thickness in various intraretinal layers were apparent in both hemispheric and ETDRS sectors.

Entities:  

Keywords:  healthy subject; intraretinal layer; thickness variation; tomography

Year:  2016        PMID: 27942202      PMCID: PMC5136365          DOI: 10.2147/OPTH.S117494

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Introduction

The retina is an extension of the brain, and the transparent ocular media facilitates direct inspection and measurement of its intraretinal layers. Optical coherence tomography (OCT) imaging has been widely used to study the neurodegeneration associated with ocular diseases such as age-related macular degeneration1 and in diseases of the central nervous system (CNS), such as Alzheimer’s disease,2 Parkinson’s disease,3 and multiple sclerosis.4–11 Atrophy of the peripapillary retinal nerve fiber layer (pRNFL) and/or the macular ganglion cell layer has been observed in these disorders. For CNS disorders, macular ganglion cell–inner plexiform layer (GCIPL) thickness has been shown to correlate better tomographically with central visual field defects, compared to pRNFL.12 Hence, GCIPL thickness may be a more important measure in studying CNS neurodegeneration and evaluating the efficacy of neuroprotective intervention. In addition to the GCIPL, other intraretinal layers such as inner nuclear layer (INL), outer nuclear layer (ONL), and outer plexiform layer (OPL) were found to be affected in conditions such as Parkinson’s disease13 and multiple sclerosis.6 Therefore, the imaging of intraretinal layers may provide additional information and facilitate a better understanding of the underlying disease mechanisms of these CNS pathologies that affect the anterior visual pathway. OCT has improved since the modality was first introduced in 1991.14 The advancements have focused primarily on imaging speed and axial resolution. Commercially available OCT systems are able to capture high-resolution, three-dimensional (3D) volumes of the retina and its sublayers.15,16 The propriety software automatically measures the thickness of pRNFL15,16 and GCIPL in sectors defined by the Early Treatment Diabetic Retinopathy Study (ETDRS).15,17 The diagnostic value of these core measurements is well documented;16 yet, other intraretinal layers provided additional and complimentary information regarding the health of the retina.18 However, the analysis of intraretinal layers is not widely used, mainly due to the limitations of the commercially available image processing software. Also, hemispherical partition may be well suitable for evaluating localized alterations in the retina due to retrogeniculate disorders.12,19 However, the hemispherical partition of the retinal thickness layers is not widely available in commonly used OCT systems. Alternative intraretinal layer segmentation software, such as Orion™ (Voxeleron LLC, Pleasanton, CA, USA), can be used in conjunction with both commercial OCT devices and custom-developed OCT systems for more advanced analyses. The software may be an important tool for quantitatively analyzing tomographic intraretinal layers, possibly resulting in a better understanding of pathogenic processing of ocular and cerebral diseases. The goal of the present study was to determine the repeatability using hemispherical and ETDRS partitions in intraretinal thickness layers in healthy human subjects using OCT. We additionally looked at the characteristic profiles of layer thicknesses within and across the measurement zones.

Materials and methods

Subjects

Twenty normal healthy subjects (11 females and 9 males) were recruited. Their mean age was 36.2±10.0 years (range 25–55 years). One eye of each subject was imaged. No subjects had ocular or systemic diseases, or underwent intraocular surgery or laser treatment. None of them had refractive error exceeding −6.00 D or +3.00 D. The study was approved by the University of Miami Institutional Review Board. All of the subjects were treated in accordance with the tenets of the Declaration of Helsinki, and written informed consent was obtained from each subject.

Custom OCT (ultrahigh-resolution OCT)

A custom-built ultrahigh-resolution OCT (UHR-OCT) device with an axial resolution of ~3 µm was used to acquire the images of macula. This system has been described elsewhere in detail.20 Briefly, UHR-OCT was based on a spectrometer with a line scan camera running at 24,000 A-scans per second. A superluminescent diode with a center wavelength of 840 nm and a bandwidth of 100 nm was used as the light source. A light delivery system with an ocular lens (60 D; Volk Optical, Mentor, OH, USA) was adapted to a slit-lamp system to facilitate retinal imaging. A green dot generated by a miniature liquid crystal display screen provided an internal fixation target. OCT real-time imaging in both the horizontal and vertical medians was displayed to facilitate positioning the scan center on the fovea during prealignment period. The 3D volume of the macula was acquired twice using the macular cube 512×128 protocol (128 consecutive B-scans) in an area of 6×6 mm2 centered on the fovea. Each B-scan consisted of 1,365 (depth) × 512 (width) pixels. The scan depth was calibrated as 2.078 mm in air.

Macular cube acquisition and image analysis

All subjects were asked to sit in front of the slit lamp, on which the OCT probe was mounted, and look at the fixation light. After X- and Y-axes alignment view showed the fovea was centered, the macular cube volumes were acquired by raster scanning. The entire scan was acquired in approximately 2.7 seconds. Automatic software (Orion, ver. 2.0.6023; Voxeleron LLC) was used to segment the thickness maps of retinal layers (Figure 1) and process the thickness maps using hemispheric and the ETDRS partition definitions (Figure 2). The segmented layers of the macular cube included the total retinal thickness (TRT), retinal nerve fiber layer (RNFL), GCIPL, INL, OPL, ONL, and photoreceptor (PR) complex. Manual review of all segmentation results was performed by one analyzer (YY) to ensure the summary measurements were not affected by noticeable errors. No manual correction was performed. The coefficients of repeatability (CoRs) and intraclass coefficients of correlation (ICCs) were analyzed using the hemispheric and ETDRS partitions. The CoR was calculated as 1.96× the standard deviation (SD) of the difference between repeated measurements. Thickness variations among zones were analyzed using analysis of variance.
Figure 1

Cross-sectional retina and segmented tomographic thickness maps of intraretinal layers.

Notes: The cross-sectional retina (A) of a healthy subject was acquired using Zeiss Cirrus HD-OCT with 6 mm scan width as a reference image. The image was screen-captured from a digital report (PDF file) of macular cube 512×128. The cross-sectional retina (B) of another healthy subject was acquired using UHR-OCT and averaged three scans centered at the fovea, illustrating the intraretinal layers, and the retina (C) directly exported from the Orion software, shows seven segmented boundaries defining six intraretinal layers, which correspond to six tomographic thickness maps (D, diameter of 6 mm). Note the image acquired with UHR-OCT appears to show higher definition of these intraretinal layers. Scale bars in (C) =250 µm.

Abbreviations: GCIPL, ganglion cell–inner plexiform layer; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; UHR-OCT, ultrahigh-resolution optical coherence tomography; HD-OCT, High Definition Optical Coherence Tomography.

Figure 2

Hemispheric and ETDRS quadrants and annuli.

Notes: Two zoning systems were used to define sectors of intraretinal layers for calculating thickness. Three concentric rings including 1, 3, and 6 mm diameters were used to divide the map into nine zones. The central 1 mm zone of the fovea was removed. The inner and outer annuli were then divided into four quadrants for each annulus. In the hemispheric definition, the division was done using vertical and horizontal medians (A). In the ETDRS definition, the quadrantal division was done using 45° and 135° medians (B).

Abbreviations: ETDRS, Early Treatment Diabetic Retinopathy Study. In the hemispheric zones: I_N_I, inner nasal inferior area; I_N_S, inner nasal superior area; I_T_I, inner temporal inferior area; I_T_S, inner temporal superior area; O_N_I, outer nasal inferior area; O_N_S, outer nasal superior area; O_T_I, outer temporal inferior area; O_T_S, outer temporal superior area. In the ETDRS zones: I_I, inner inferior area; I_N, inner nasal area; I_S, inner superior area; I_T, inner temporal area; O_I, outer inferior area; O_N, outer nasal area; O_S, outer superior area; O_T, outer temporal area.

Results

All datasets were successfully segmented to create six thickness maps of individual intraretinal layers in addition to the map of TRT (Figure 3). The CoRs of these layers in hemispheric zones ranged from 0.9 to 6.6 µm, with an average of 3.6 µm (SD 1.4), which was not significantly different compared to those of ETDRS sectors (average 3.6 µm, SD 1.4, range 1.1–8.0 µm, P>0.05) (Figure 4). The CoRs showed excellent repeatabilities in RNFL, INL, OPL, ONL, and PR (CoR of each sector <6 µm) and fair repeatability in GCIPL and TRT (the worst sector ≥6 µm) between two measurements. ICCs of these layers in hemispheric zones ranged from 0.68 to 0.99, with an average of 0.91 (SD 0.07). There were no significant differences in ICCs between two different zoning methods (P>0.05) (Figure 5).
Figure 3

Tomographic thickness (µm) of the total retina and six intraretinal layers.

Notes: The mean thicknesses (±SD) of these layers are displayed in each sector using hemispheric (A) and ETDRS (B) definitions. Different thicknesses were apparent between the sectors in each layer in both zoning definitions.

Abbreviations: ETDRS, Early Treatment Diabetic Retinopathy Study; GCIPL, ganglion cell–inner plexiform layer; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; SD, standard deviation; TRT, total retinal thickness.

Figure 4

Repeatability of measuring total retinal thickness and six intraretinal layers in hemispheric and ETDRS sectors.

Notes: CoRs of hemispheric sectors (A) ranged from 0.9 to 6.6 µm, with an average of 3.6 µm (SD 1.4), which was not significantly different compared to those of ETDRS sectors (B, average 3.6 µm, SD 1.4, range 1.1–8.0 µm).

Abbreviations: CoRs, coefficients of repeatability; ETDRS, Early Treatment Diabetic Retinopathy Study; GCIPL, ganglion cell–inner plexiform layer; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor layer; RNFL, retinal nerve fiber layer; SD, standard deviation; TRT, total retinal thickness.

Figure 5

ICC values.

Notes: ICC of measuring tomographic thicknesses of total retina and six intraretinal layers in hemispheric zones ranged from 0.68 to 0.99, with an average of 0.91 (SD 0.07). There were no significant differences in ICCs between hemispheric (A) and ETDRS (B) definitions (average 0.91, SD 0.08, range 0.64–0.99, P>0.05). The ICC values were classified as fair (≤0.8), good (0.8–0.9), and excellent (≥0.9).

Abbreviations: ETDRS, Early Treatment Diabetic Retinopathy Study; GCIPL, ganglion cell–inner plexiform layer; ICC, intraclass coefficient of variation; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; SD, standard deviation; TRT, total retinal thickness.

Significant variations of tomographic intraretinal thicknesses were found between inner and outer annuli and among quadrantal sectors within the inner and outer annuli (P<0.05) (Figure 6). The average thickness of the total retina and the intraretinal layers, except for RNFL and PR, in the inner annulus was greater than in the outer annulus (P<0.01) (Figure 7). Significant variations among quadrantal sectors including both inner and outer annuli were evident in intraretinal layers (P<0.05) except for OPL and PR (P>0.05) (Figure 8). Post hoc tests of the sectors summed from inner and outer annuli showed that thicknesses in nasal side including nasal, nasal superior, and nasal inferior sectors were significantly more than in temporal side including temporal, temporal superior, and temporal inferior sectors in total retina and intraretinal layers (P<0.05) except for OPL (P>0.05).
Figure 6

Variations of tomographic thickness of total retina and six intraretinal layers.

Notes: Significant variations of tomographic intraretinal thicknesses were found between inner and outer annuli (A) and among quadrantal sectors within the inner (B) and outer annuli (C) (P<0.05). Significant variations of the quadrantal sectors (D) combined from both inner and outer annuli of each intraretinal layer were evident in intraretinal layers (P<0.05) except for OPL (P>0.05). Post hoc tests of the sectors summed from inner and outer annuli (D) showed that thicknesses in nasal side including nasal, nasal superior, and nasal inferior sectors were significantly higher than the temporal side including temporal, temporal superior, and temporal inferior sectors, and in thicknesses in total retina and intraretinal layers (P<0.05) except for OPL (P>0.05). Bars denote standard deviation.

Abbreviations: GCIPL, ganglion cell–inner plexiform layer; I, inferior; INL, inner nuclear layer; N, nasal; NI, nasal inferior; NS, nasal superior; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; S, superior; T, temporal; TI, temporal inferior; TRT, total retinal thickness; TS, temporal superior.

Figure 7

Tomographic thickness variations between annuli of the total retina and six segmented intraretinal layers.

Notes: The average thickness of the total retina and intraretinal layers, except for RNFL and PR, in the inner annuli was greater than the outer annuli (P<0.01). In those layers showing significant differences, the significantly thicker annuli are marked in green and the thinner annuli are marked in red. There was significant difference between inner and outer annuli for OPL. Data presented as thickness (µm) ± SD.

Abbreviations: GCIPL, ganglion cell–inner plexiform layer; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; TRT, total retinal thickness.

Figure 8

Tomographic thickness variations among quadrants of the total retina and six segmented intraretinal layers.

Notes: The thickness maps were divided into quadrants using both hemispheric (A) and ETDRS (B) definitions. Significant variations of quadrantal thicknesses were evident in the total retina and intraretinal layers (P<0.01) except for OPL. The quadrants with the greatest thickness (marked in green) were mostly located in the nasal sides including nasal, nasal superior, and nasal inferior quadrants, except for PR. In contrast, the thinnest quadrants (marked in red) were mostly located in the temporal side including temporal and temporal inferior and the inferior quadrant (PR). Post hoc tests showed the difference between the thickest and thinnest quadrants to be significant (P<0.01) in the total retina and six segmented intraretinal layers, except for OPL.

Abbreviations: ETDRS, Early Treatment Diabetic Retinopathy Study; GCIPL, ganglion cell–inner plexiform layer; INL, inner nuclear layer; ONL, outer nuclear layer; OPL, outer plexiform layer; PR, retinal photoreceptor; RNFL, retinal nerve fiber layer; TRT, total retinal thickness.

Discussion

OCT is widely used as a noninvasive diagnostic tool in the field of ophthalmology. With fine resolution, fast data acquisition, and good penetration (depending on wavelength), the layered structure of the retina can be resolved. Most commercial devices such as Cirrus HD-OCT™ (Carl Zeiss Meditec, Inc, Dublin, CA, USA), Heidelberg Spectralis OCT system (Heidelberg Engineering, Heidelberg, Germany), and RTVue OCT (Optovue, Inc, Fremont, CA, USA) share similar functionalities for acquiring volumetric data in a very short period of time.21,22 In addition, custom laboratory OCT devices, such as the one used in the present study and in other studies,23,24 can resolve equivalent or better intraretinal layers, given the better axial resolution. The details of multiple intraretinal layers can be clearly visualized in the cross-sectional images and attempts have been made to segment these layers in both two dimensions24–26 and three dimensions.27–30 Yet, these commercially available OCT devices generate automated thickness maps for only limited layers such as TRT and GCIPL thicknesses.21,31,32 The advances in automated and semiautomated segmentation have facilitated delineation of multiple intraretinal layers in three dimensions.27,29,30,33,34 However, with the exception of Orion, a software that automates the thickness mappings of six to eleven intraretinal layers is not commercially available. Orion has been previously applied in processing volumetric data obtained from a commercial device (Cirrus HD-OCT), segmenting in this case, the thickness maps of six retinal layers.29 The software was adapted to our custom OCT, prompting our need for testing the repeatability of the intraretinal thicknesses ahead of performing further clinical studies. This is especially important when the software was developed entirely independently of the acquisition hardware, as is the case with Orion. Testing the repeatability and characteristic profiles of these segmented layers will lay the necessary foundation for further study design, including, potentially, the creation of normative ranges. Our tomographic thickness data obtained from our custom-made OCT with the implementation of Orion software are in agreement with the published data.31,32,34,35 Sull et al compared TRT in ETDRS partitions obtained with several commercial devices and found the results obtained with SD-OCT to be similar.21 Our results of TRT were within the ranges, compared to the results reported by others.21,25 The thickness maps for six intraretinal layers obtained in the present study in EDTRS sections were also similar to those reported in a previous study,22 with some mild disparities. This could, in part, be due to the definition of the boundaries where the thickness of a certain layer is calculated. For example, TRT can be defined as from the ILM to the inner/outer segment junction, retinal pigment epithelium (RPE), or to Bruch’s membrane.36 Given the lack of standardization and consensus among the devices, we may interpret our findings in the context of the existing literature. Compared with previous studies,35,37,38 the average thickness value of ONL was higher in the present study, while the thickness values of ONL and the inner segment of photoreceptors (IS) were similar to those in previous studies.37,39 Furthermore, the average thickness of PR was consistent with that of the outer segment of photoreceptors.22,38,39 Hu et al37 reported that the normal average thickness of ONL was 54 µm and that of IS was 26 µm. Demirkaya et al39 measured the thickness of ONL + IS, and the result (85±8 µm) was consistent with the values obtained in a previous study by Hu et al37 and in the present study (81±5 µm). These findings may possibly be because the ONL in the present study included ONL and IS, and PR might include the outer segment of photoreceptors. The thickness values of OPL layer in ETDRS sectors were lower than those in other studies,22,25,37,39 but similar to the value reported by Liu et al.35 Since the OCT used in the present study and by Liu et al are both custom-built systems with similar hardware configuration, similar results from healthy subjects may be expected. The remaining differences between the present study and other previous reports may also be simply due to the variation across normal subjects, imaging resolution, and algorithm differences. Further studies with larger sample size will establish more robust normative database for thickness maps of the intraretinal layers, and a common consensus of what needs to be measured will also be helpful.21,36 While the present study focused on the comparability of two partition methods in segmentation repeatability, analyzing the repeatability of each segmented intraretinal layer using different area partitions provides characteristic information about the measurement, which will be useful for determining the study sample size for future clinical studies. We reported the details of segmentation repeatability in annuli and sectors using two partition methods and found that they were comparable. In addition, the repeatabilities in annuli and section were comparable to those reported in previous studies.24,26,40,41 Oberwahrenbrock et al reviewed the reliability of intraretinal thickness measurements and estimated the ICCs and CoRs from previously published data.40 With semiautomated segmentation of 2D OCT images, very good ICCs and CoRs were evident for all intraretinal layers, except for OPL.40 Similar reliability of segmenting 2D intraretinal layers was documented in previous studies,24,41 in which manual segmentation or semiautomated segmentation was used on 2D OCT images. It may not, however, be feasible to use manual correction in volumetric data segmentation when a group of subjects is tested. Fully automated segmentation without supervision and manual correction was intentionally used in the present study, where we found by reviewing that the layers obtained with UHR-OCT can be measured reliably. The repeatability of the average thickness values of total macula and each intraretinal layer appeared to be much better than the results obtained with the time-domain OCT.26 The results we report in this study are similar to the published repeatability data using manual correction in 2D segmentation.24,40,41 Our results show that good repeatabilities are achievable for automated segmentation of most of intraretinal layers. However, segmentation validation against the ground truth may need to be further investigated. Also, further studies will be needed to test whether OCT with higher resolution, such as the one we used in the present study, yields higher repeatability than conventional OCT. Because two boundaries are needed to measure a given layer, we may also hypothesize segmentation accuracy for a single boundary. Current reports using laboratory segmentation approaches revealed that segmentation accuracy of 3–4 µm could be achieved in segmenting one boundary of these intraretinal layers.27,30,42 It may be reasonable to expect, therefore, that the thickness accuracy for a single layer could be larger; toward 8 µm. We found the averaged CoR in the represented study was ~4 µm for a layer with two boundaries, indicating excellent overall repeatability. Ahead of future clinical studies, it is also important to establish the intravisit variance for determining the application of OCT with difference partition methods. More detailed information can be obtained after subfield partition. Different software employ different methods for reporting, with the ETDRS grid being the most common43 and the hemispheric sectors being more appropriate for the loss resulting from CNS neurodegeneration.12,19 We tested both of these partitions in an attempt to better understand and characterize the thickness variation in the tomographic mappings. In the parafoveal area with hemispheric segmentation, thickness maps of the RNFL showed disparity in different quadrants. The RNFL was the thickest in the nasal quadrant of outer annuli and the thinnest in the temporal quadrant of outer annuli, which might be expected due to the known normal distribution of nerve fibers between the optic nerve head and macula. The GCIPL thickness was the greatest in the inferior sector of the inner annulus in our study. When counting the sector of both inner and outer annuli, the thickest GCIPL was located in the nasal side. This is in agreement with a previous report that the GCIPL was thicker in the nasal sector–based EDTRS division of the parafoveal retinal area.22 Our data may provide an alternative quantitative measurement which might reflect the intraretinal structural characterization. The different results might be explained by the difference in sample size, sex, and age. The thickness values of INL, OPL, and photoreceptor + retinal pigment epithelium were relatively uniform in all of the sectors of macula, while the thickness maps of the ONL showed a maximum value in the center of the fovea and a relatively high value in inner annulus subfield. This again might be due to the elongated cone photoreceptors in the normal fovea.22

Limitations

There were some limitations in this study. First, we only investigated normal subjects to demonstrate the imaging and segmentation capability and repeatability in order to establish a comparative reference for eyes with diseases. The repeatability and thickness variation in eyes with diseases will need to be further investigated. Second, the speed of UHR-OCT is not fast enough to reduce motion artifacts and we do not have eye tracking system, common to most commercial instrumentations, which may explain the relatively poor repeatability in some of the outer annulus or quadrants. We examined all volumetric datasets and found that the 6 mm diameter circle centered on the fovea would occasionally go outside the bounds of the acquired image data. It could, therefore, be speculated that the repeatability of measuring tomographic thickness of intraretinal layers can be improved by using eye tracking or high-speed OCT systems. Third, our sample size is small, although we were able to find significant thickness variations among the segmented intraretinal layers, which, to our understanding, likely characterize the normal structural variation. Future studies with larger sample sizes may further confirm these variations and perhaps facilitate the creation of a normative database. Fourth, we compared our results of TRT and tomographic thicknesses of intraretinal layers with published data and found that they were compatible. Future studies will be needed to directly compare the results obtained from our UHR-OCT with those obtained from commercial devices. Due to large data volume, it took about 7.5 seconds to process one volumetric dataset, which is slower compared to the processing of data from the Cirrus instrument, a direct result of the larger number of voxels in our data set (typically, the processing time is exponential with data size) and the increased number of layers that are reported. However, the processing speed remains faster than other custom software for segmenting multiple intraretinal layers and, therefore, more suitable to clinical applications.27,30,44 Fifth, the accuracy will need to be tested by comparing the segmented results to the ground truth using manual segmentation. Lastly, the system used in the present study is a custom-made OCT and there is no normative database available. This study may serve as a start point to generate normative thickness maps of intraretinal layers.

Conclusion

In summary, the measurement repeatabilities of tomographic thicknesses of intraretinal layers using both hemispheric and ETDRS partitions were comparable. Significant variations were apparent in tomographic thickness of intraretinal layers in both hemispheric and ETDRS sectors. Such characterization of tomographic thickness of intraretinal layers using OCT and automated segmentation in normal eyes is an important requisite step ahead of utilizing such techniques in diseased eyes and clinical studies.
  44 in total

1.  Slit-lamp-adapted ultra-high resolution OCT for imaging the posterior segment of the eye.

Authors:  Hong Jiang; Fawzi Abukhalil; Meixiao Shen; Giovanni Gregori; Byron L Lam; Yuhong Wang; Jianhua Wang
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2012 Jan-Feb

2.  Quantitative analysis of the intraretinal layers and optic nerve head using ultra-high resolution optical coherence tomography.

Authors:  Yuhong Wang; Hong Jiang; Meixiao Shen; Byron L Lam; Delia Cabrera DeBuc; Yufeng Ye; Ming Li; Aizhu Tao; Yilei Shao; Jianhua Wang
Journal:  J Biomed Opt       Date:  2012-06       Impact factor: 3.170

3.  Effects of age, sex, and axial length on the three-dimensional profile of normal macular layer structures.

Authors:  Sotaro Ooto; Masanori Hangai; Atsuo Tomidokoro; Hitomi Saito; Makoto Araie; Tomohiro Otani; Shoji Kishi; Kenji Matsushita; Naoyuki Maeda; Motohiro Shirakashi; Haruki Abe; Shinji Ohkubo; Kazuhisa Sugiyama; Aiko Iwase; Nagahisa Yoshimura
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-11-11       Impact factor: 4.799

4.  Early Treatment Diabetic Retinopathy Study design and baseline patient characteristics. ETDRS report number 7.

Authors: 
Journal:  Ophthalmology       Date:  1991-05       Impact factor: 12.079

5.  Changes in macular layers in the early course of non-arteritic ischaemic optic neuropathy.

Authors:  Johannes Keller; Jonathan D Oakley; Daniel B Russakoff; Magí Andorrà-Inglés; Pablo Villoslada; Bernardo F Sánchez-Dalmau
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-05-28       Impact factor: 3.117

6.  Ganglion cell-inner plexiform layer thickness in patients with Parkinson disease and association with disease severity and duration.

Authors:  Esin S Sari; Rabia Koc; Alper Yazici; Gözde Sahin; Sitki S Ermis
Journal:  J Neuroophthalmol       Date:  2015-06       Impact factor: 3.042

7.  Optical coherence tomography and disease subtype in multiple sclerosis.

Authors:  M Pulicken; E Gordon-Lipkin; L J Balcer; E Frohman; G Cutter; P A Calabresi
Journal:  Neurology       Date:  2007-11-27       Impact factor: 9.910

8.  Lesions in the posterior visual pathway promote trans-synaptic degeneration of retinal ganglion cells.

Authors:  Johannes Keller; Bernardo F Sánchez-Dalmau; Pablo Villoslada
Journal:  PLoS One       Date:  2014-05-23       Impact factor: 3.240

9.  Comparative diagnostic accuracy of ganglion cell-inner plexiform and retinal nerve fiber layer thickness measures by Cirrus and Spectralis optical coherence tomography in relapsing-remitting multiple sclerosis.

Authors:  Julio J González-López; Gema Rebolleda; Marina Leal; Noelia Oblanca; Francisco J Muñoz-Negrete; Lucienne Costa-Frossard; José C Alvarez-Cermeño
Journal:  Biomed Res Int       Date:  2014-09-18       Impact factor: 3.411

10.  Macular thickness measurements using Copernicus Spectral Domain Optical Coherence Tomography.

Authors:  Laxmi Gella; Rajiv Raman; Tarun Sharma
Journal:  Saudi J Ophthalmol       Date:  2014-10-18
View more
  16 in total

1.  Retinal Tissue Perfusion in Patients with Multiple Sclerosis.

Authors:  Yi Liu; Silvia Delgado; Hong Jiang; Ying Lin; Jeffrey Hernandez; Yuqing Deng; Giovana Rosa Gameiro; Jianhua Wang
Journal:  Curr Eye Res       Date:  2019-05-15       Impact factor: 2.424

2.  The effect of hemodialysis on individual retinal layer thickness.

Authors:  Idit Maharshak; Idan Hecht; Lihi Mankuta; Asaf Achiron; Oriel Spierer; Zvia Burgansky; Asaf Bar; Aliza Grossbard; Anna Katkov; Relu Cernes; Leonid Feldman
Journal:  Int Ophthalmol       Date:  2021-01-03       Impact factor: 2.031

3.  Retinal layers thickness changes following epiretinal membrane surgery.

Authors:  I Hecht; I Yeshurun; E Bartov; A Bar; Z Burgansky-Eliash; A Achiron
Journal:  Eye (Lond)       Date:  2017-11-10       Impact factor: 3.775

4.  Changes in volume of various retinal layers over time in early and intermediate age-related macular degeneration.

Authors:  Ali Lamin; Jonathan D Oakley; Adam M Dubis; Daniel B Russakoff; Sobha Sivaprasad
Journal:  Eye (Lond)       Date:  2018-10-11       Impact factor: 3.775

5.  Age-related focal thinning of the ganglion cell-inner plexiform layer in a healthy population.

Authors:  Yuqing Deng; Huijuan Wang; Ava-Gaye Simms; Huiling Hu; Juan Zhang; Giovana Rosa Gameiro; Tatjana Rundek; Joseph F Signorile; Bonnie E Levin; Jin Yuan; Jianhua Wang; Hong Jiang
Journal:  Quant Imaging Med Surg       Date:  2022-06

6.  Focal alteration of the intraretinal layers in neurodegenerative disorders.

Authors:  Shriya Airen; Ce Shi; Zhiping Liu; Bonnie E Levin; Joseph F Signorile; Jianhua Wang; Hong Jiang
Journal:  Ann Eye Sci       Date:  2020-03

Review 7.  Advances in ophthalmic structural and functional measures in multiple sclerosis: do the potential ocular biomarkers meet the unmet needs?

Authors:  Hong Jiang; Silvia Delgado; Jianhua Wang
Journal:  Curr Opin Neurol       Date:  2021-02-01       Impact factor: 6.283

8.  Age-Related Alterations in the Retinal Microvasculature, Microcirculation, and Microstructure.

Authors:  Yantao Wei; Hong Jiang; Yingying Shi; Dongyi Qu; Giovanni Gregori; Fang Zheng; Tatjana Rundek; Jianhua Wang
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-07-01       Impact factor: 4.799

9.  Longitudinal Study of Retinal Structure, Vascular, and Neuronal Function in Patients With Relapsing-Remitting Multiple Sclerosis: 1-Year Follow-Up.

Authors:  Qi Chen; Hong Jiang; Silvia Delgado; Jeffrey Hernandez; Diego Eduardo Alba; Giovanni Gregori; Kottil W Rammohan; Vittorio Porciatti; Jianhua Wang
Journal:  Transl Vis Sci Technol       Date:  2021-05-03       Impact factor: 3.283

10.  Retinal tissue hypoperfusion in patients with clinical Alzheimer's disease.

Authors:  Giovana Rosa Gameiro; Hong Jiang; Yi Liu; Yuqing Deng; Xiaoyan Sun; Bernardo Nascentes; Bernard Baumel; Tatjana Rundek; Jianhua Wang
Journal:  Eye Vis (Lond)       Date:  2018-08-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.