Macular thickness measurements using Copernicus Spectral Domain Optical Coherence Tomography.

Laxmi Gella1, Rajiv Raman2, Tarun Sharma2.   

Abstract

PURPOSE: To provide normal macular thickness measurements using Spectral Domain Optical Coherence Tomography (SDOCT, Copernicus, Optopol Technologies, Zawierci, Poland).
METHODS: Fifty-eight eyes of 58 healthy subjects were included in this prospective study. All subjects had comprehensive ophthalmic examination including best-corrected visual acuity (BCVA). All the subjects underwent Copernicus SDOCT. Central foveal thickness (CFT) and photoreceptor layer (PRL) thickness were measured and expressed as mean and standard deviation. Mean retinal thickness for each of the 9 regions defined in the Early Treatment Diabetic Retinopathy Study was reported. The data were compared with published literature in Indians using Stratus and Spectralis OCTs to assess variation in instrument measurements.
RESULTS: The mean CFT in the study sample was 173.8 ± 18.16 microns (131-215 microns) and the mean PRL thickness was 65.48 ± 4.23 microns (56-74 microns). No significant difference (p = 0.148) was found between CFT measured automated (179.28 ± 22 microns) and manually (173.83 ± 18.1 microns). CFT was significantly lower in women (167.62 ± 16.36 microns) compared to men (180.03 ± 18 microns) (p = 0.008). Mean retinal thickness reported in this study was significantly different from published literature using Stratus OCT and Spectralis OCT.
CONCLUSION: We report the normal mean retinal thickness in central 1 mm area to be between 138 and 242 microns in Indian population using Copernicus SDOCT. We suggest that different OCT instruments cannot be used interchangeably for the measurement of macular thickness as they vary in segmentation algorithms.

Entities:  

Keywords:  Foveal thickness; Photoreceptor layer thickness; Retina; Retinal thickness; Spectral Domain Optical Coherence Tomography

Year:  2014        PMID: 25892930      PMCID: PMC4398812          DOI: 10.1016/j.sjopt.2014.10.003

Source DB:  PubMed          Journal:  Saudi J Ophthalmol        ISSN: 1319-4534


Introduction

The major development in ophthalmic imaging was the introduction of optical coherence tomography (OCT) in 1991 by Huang et al. Ophthalmoscopy, fundus photography and fluorescein angiography are the common tools to diagnose diabetic retinopathy (DR) and diabetic macular edema (DME). Due to the non-invasive nature of the OCT technique it might replace or add as a complementary to fluorescein angiography. Spectral Domain OCT (SDOCT) is an advanced modification of traditional time domain OCT. The main advantages of the SDOCT are speed and sensitivity, which have helped in conducting advanced clinical and research oriented studies. There is an increasing demand for high-resolution imaging of the ocular tissue to improve the diagnosis and management of various retinal diseases. Retinal thickness is defined as the distance between the vitreoretinal interface and the inner border of retinal pigment epithelium (RPE). Various SDOCTs are commercially available but the segmentation software of these instruments identifies different hyperreflective structures in each cross-sectional image. All instruments in common identify the vitreoretinal interface as inner retinal border. The segmentation of outer retinal border identified by different instruments varies significantly. Stratus OCT considers Photoreceptor-RPE complex as the outer retinal border. Copernicus SDOCT, Spectral OCT/SLO and RTVue-100 identify the inner border of RPE as the outer retinal border. Cirrus HD-OCT measures retinal thickness up to the outer band of the RPE, whereas the Spectralis OCT includes Bruch’s membrane in the retinal thickness measurement. Studies have reported normative data of macular thickness using various commercially available SDOCTs. The macular thickness measurement for diagnostic function may differ with the population which is used as a database as well as the instrument which is being used. So it is required that the retinal thickness measurements of normal subjects be as close as possible to the population for which the instrument is to be used. To the best of our knowledge there is no normative database available for Indians using Copernicus SDOCT. The present study provides retinal thickness measurements using Copernicus SDOCT in subjects without any known retinal diseases to establish a normative data for clinical use.

Materials and methods

Fifty-eight eyes of 58 healthy volunteers were included in this prospective study. One eye of each subject was selected randomly by generating random numbers in excel. The study was approved by the organization’s Institutional Review Board, and informed consent was obtained from the subjects in accordance with the Helsinki Declaration. Demographic data and detailed medical and ocular history were obtained from the subjects. All subjects underwent comprehensive ophthalmic examination which included refraction, assessment of intraocular pressure, anterior and posterior segment evaluation including cup disk ratio measurement. Subjects with any ocular pathology or with glaucomatous changes or glaucoma suspects and those with history of any ocular surgeries were excluded. Normal eyes were defined as those that had best correct visual acuity of 20/20 or better. Refractive errors within ±3.0D sphere and less than or equal to ±1.0D cylinder were included. And subjects with significant media opacities which could lead to poor quality scans on SDOCT were excluded. All included subjects underwent Spectral Domain Optical Coherence Tomography (Copernicus, SDOCT, Optopol Technologies, Zawierci, Poland) by a single operator. SDOCT scans were performed through a dilated pupil while monitoring the reconstructed video image of the central retina. The programs used for the present study were asterisk scan and 3D scan protocols. We can vary the scan length from 4 to 10 mm and also the number of A-scans and B-scans. For the purpose of this study we used a scan length of 7 mm with 6 B-scans and 3000 A-scans per B-scans through the center of the fovea for the asterisk scan protocol. The scan acquisition time was 0.8 s. 3D scan protocol was used with 7 mm scan length with 50 B-scans and 1000 A-scans per B-scan with the time acquisition of 2.4 s. All the measurements were calculated in microns. The following parameters were noted on SDOCT: Central foveal thickness (CFT) was defined as the distance between the vitreoretinal interface and the anterior surface of the RPE and this was measured manually and also automated using measurement software in SDOCT and this was measured in the B-scan where a hyperreflective echo was noted which represents the center of the fovea. This was acquired from asterisk scan protocol. Mean retinal thickness (MRT) was noted at the central 1 mm, middle 3 mm and the outer 6 mm ring, in the superior, inferior, nasal and temporal quadrants. These measurements are given by the automated software. This measurement was acquired from 3D scan protocol. Photoreceptor layer (PRL) thickness was measured at the central fovea (defined above) from asterisk scan, which was defined as the distance between the external limiting membrane which appears as a thin hyperreflective line on SDOCT and the anterior surface of the retinal pigment epithelium (RPE). Intra-observer and inter-observer repeatability was determined for the manual measurement of OCT thickness parameters from a pilot study. Intra-observer repeatability was found to be good in measuring the SDOCT outcomes with intraclass correlation of 0.99 for CFT, 0.63 for PRL thickness. For the inter-observer repeatability the mean difference in the CFT measurements was 6.7 μm with limits of agreement ranging from −19.4 to 32.8 μm. The mean difference in the PRL thickness was 8 μm with 95% limits of agreement between −7.4 and 23.4 μm. Statistical analyses were performed using SPSS (Statistical Package for Social Sciences, version 15.0, Chicago, IL, USA). The results were expressed as mean ± SD. Independent t-test was used for comparing the parameters between the groups. The level of statistical significance for P-value was <0.05.

Results

Mean age of the subjects was 36 ± 12 years (range: 21–76 years). There was no significant difference in age between men and women (p = 0.28). The mean central foveal thickness in the study sample was 173.8 ± 18.16 microns (131–215 microns) and the mean photoreceptor layer thickness was 65.48 ± 4.23 microns (56–74 microns). No significant difference (p = 0.148) was found between CFT measured automated (179.28 ± 22 microns) and manually (173.83 ± 18.1 microns). Fig. 1 shows the mean retinal thickness in 9 ETDRS subfields. Temporal quadrant of 6 mm radius was the thinnest in relation to nasal, superior and inferior quadrants.
Figure 1

Macular thickness (μm) of study sample in 9 subfields as defined by ETDRS (mean ± SD).

Table 1 shows the comparison of thickness parameters between men and women. CFT was significantly lower in women (167.62 ± 16.36 microns) compared to men (180.03 ± 18 microns) (p = 0.008). Women showed significantly decreased retinal thickness in all ETDRS subfields except for the nasal quadrant in 3 mm radius, superior and temporal quadrants of 6 mm radius which did not show any significant difference.
Table 1

Comparison of thickness parameters among men and women.

VariablesMean ± SD
p
Men (n = 29)Women (n = 29)
Age37.9 ± 1334.5 ± 10.30.28
Central foveal thickness180.03 ± 18167.62 ± 16.360.008
Photoreceptor layer thickness65.96 ± 4.4765 ± 40.39
Mean retinal thickness (1 mm)202.31 ± 21.2178.1 ± 26<0.001



3 mm radius
 Superior277.07 ± 24261.8 ± 250.02
 Inferior291 ± 20.48279.3 ± 24.030.05
 Temporal279 ± 19.40260 ± 240.002
 Nasal283 ± 21270 ± 300.06



6 mm radius
 Superior296 ± 14.59289.03 ± 18.110.114
 Inferior293.48 ± 17281.10 ± 16.350.006
 Temporal282 ± 13.33276.14 ± 17.140.161
 Nasal310 ± 13.25297 ± 16.580.002

p < 0.05 is statistically significant.

Table 2 shows comparison of mean retinal thickness assessed using OCT in healthy individuals given by various studies. It shows that there was a significant variation in mean retinal thickness measured by different SDOCT instruments in different population.
Table 2

Comparison of mean retinal thickness in other studies and present study.

S No.StudyYearSampleOCT instrumentMean retinal thickness (1 mm)Scan protocolA-scan/B-scanRefractive error
1Tewari HK (India)42004170Stratus181.15 ± 18.42Radial scanNA(−8 to +5.8D)
2Appukuttan B (India)52013105Spectralis260.1 ± 18Cube scan49 B scans, 40,000 A/BWithin ±5D
3Choovuthayakorn J (Thailand)62012368Spectralis259.18 ± 19.08Raster scanNANA
4Adhi M (Pakistan)720122203D OCT, Topcon229 ± 20.46Raster scan256 × 256Within −5D
5Mitkova-Hristova VT (Bulgaria)82011163Optovue248.9 ± 17.9Retinal mapNAWithin ±5D
6Legarreta JE (Florida)9200850Cirrus HD-OCT258.2 ± 23.5Macular cube200 × 200NA
7Grover S (Florida)10200950Spectralis270.2 ± 22.5Radial scansNAWithin ±6D
8Sabates FN (Missouri)112011169Spectral OCT/SLO254 ± 27NA200 × 2003 to −10
9Wolf-Schnurrbusch UE18200920Stratus OCT213 ± 19Fast macula6 B scans, 128 A/B1 to −2
Spectralis288 ± 16Volume scan512 × 49
Spectral OCT/SLO243 ± 253D retinal topography512 × 64
Cirrus HD-OCT276 ± 17Macular cube512 × 128
SOCT Copernicus246 ± 233D scan673 × 50
RTVue-100245 ± 28Macular map512 × 101
10Kakinoki M (Japan)19200850Stratus OCT197.2 ± 17.8Fast macula6 B scans, 128 A/BNA
Cirrus HD-OCT257.6 ± 19.6Macular cube200 × 200
11Leung CK (Hong Kong)20200835Stratus OCT195.6 ± 17.2Fast macula6 B scans, 128 A/BFrom +3 to −6D
3D OCT, Topcon216.4 ± 18Raster scan256 × 256
12Present study201458SOCT Copernicus1903D scan50 B scans, 1000A/BWithin ±3D
Normative data of this study using Copernicus SDOCT were compared to the normative data of Indian population published using Stratus OCT and Spectralis OCT in Table 3. There was a significant difference in mean retinal thickness measurement in all ETDRS subfields among the three studies.
Table 3

Comparison with Indian normative data of other studies.

Thickness parametersOCT considering IS/OS RPE junction as outer retinal border (1)OCT considering outer RPE surface as outer retinal border (2)OCT considering inner RPE surface as outer retinal border (3)p1 vs 21 vs 32 vs 3
Stratus4Spectralis5Copernicus (current study)
Central foveal thickness149.16 ± 21.15220.96 ± 13.76173.83 ± 18.16<0.0001<0.0001<0.0001<0.0001
Mean retinal thickness (1 mm)181.15 ± 18.42260.10 ± 18.19190.21 ± 26.31<0.0001<0.0001<0.0001<0.0001



3 mm radius
 Superior254.65 ± 20.99337.95 ± 17.46269.47 ± 25.22<0.0001<0.0001<0.0001<0.0001
 Inferior256.19 ± 18.83335.53 ± 17.87285.03 ± 22.85<0.0001<0.0001<0.0001<0.0001
 Temporal223.69 ± 25.05324.90 ± 16.35269.19 ± 23.65<0.0001<0.0001<0.0001<0.0001
 Nasal257.90 ± 20.54338.88 ± 18.17276.29 ± 26.07<0.0001<0.0001<0.0001<0.0001



6 mm radius
 Superior228.08 ± 15.72295.62 ± 14.71292.50 ± 16.67<0.0001<0.0001<0.00010.192
 Inferior217.86 ± 15.19283.46 ± 15.25287.29 ± 17.53<0.0001<0.0001<0.00010.094
 Temporal209.48 ± 16.88281.60 ± 14.21279.00 ± 15.49<0.0001<0.0001<0.00010.16
 Nasal245.09 ± 16.75312.23 ± 17.08303.29 ± 16.29<0.0001<0.0001<0.00010.001

p < 0.05 is statistically significant.

Discussion

Nasal quadrant of 6 mm radius was significantly thicker compared to other quadrants’ mean retinal thickness which could be explained by the anatomical relationship of the converging retinal nerve fibers with optic disk. The observed macular thickness was thinnest at the fovea and increased towards parafoveal area which is consistent with the normal anatomical contour. Using the criteria of mean ± 2 SDs, which includes 95% of the population, we suggest the normal mean retinal thickness in central 1 mm area to be between 138 and 242 microns in Indian population using Copernicus SDOCT. This range of central retinal thickness was significantly more compared to the study done by Tewari et al. and less compared to that suggested by the Appukuttan et al. which could be due to the different instruments used in the present study. Women had a significantly lower CFT and also reduced thickness in all ETDRS quadrants except for the nasal quadrant in 3 mm radius and superior and temporal quadrants of 6 mm radius. These findings are in support with those in previous reports, confirming the impact of sex on central retinal thickness measurements. Reduced central foveal thickness is compatible with the observation that women have higher risk of developing macular holes. Table 2 shows that there was a significant variation in mean retinal thickness among various studies using different OCT instruments. This could be explained by the difference in the study methodologies and the sample population. Mean retinal thickness values in Indian population were significantly different in all ETDRS subfields when our study results were compared with the normative data published by other studies in India using Stratus and Spectralis OCT. It is attributed by the fact that the retinal layer segmentation was significantly varied between the instruments. Stratus OCT considers the inner segment outer segment junction of photoreceptors as the outer retinal border for the retinal thickness measurement and Spectralis OCT considers outer border of retinal pigment epithelium as the outer retinal border. Thus the decreased mean retinal thickness in the current study could be due to the exclusion of retinal pigment epithelial thickness which was included in Spectralis OCT. In summary we provide the normative values of retinal thickness in Indian sample using Copernicus SDOCT. We found that the mean retinal thickness given by the three different OCT instruments in Indian sample were significantly different. These discrepancies were probably due to the segmentation algorithms used by different OCT instruments. Furthermore, as the segmentation algorithms are different among OCT devices, the outer boundary used for the thickness measurement may strongly depend on the OCT system used. These data imply that different OCT instruments cannot be used interchangeably for the measurement of macular thickness.

Conflict of interest

The authors declared that there is no conflict of interest.
  18 in total

1.  Macular thickness measurements in healthy eyes using spectral optical coherence tomography.

Authors:  Vesela T Mitkova-Hristova; Marieta I Konareva-Kostyaneva
Journal:  Folia Med (Plovdiv)       Date:  2011 Oct-Dec

2.  Macular thickness evaluation using the optical coherence tomography in normal Indian eyes.

Authors:  Hem K Tewari; Vijay B Wagh; Parul Sony; Pradeep Venkatesh; Rajbeer Singh
Journal:  Indian J Ophthalmol       Date:  2004-09       Impact factor: 1.848

3.  Macular thickness variations with sex, age, and axial length in healthy subjects: a spectral domain-optical coherence tomography study.

Authors:  Won Kyung Song; Sung Chul Lee; Eun Suk Lee; Chan Yun Kim; Sung Soo Kim
Journal:  Invest Ophthalmol Vis Sci       Date:  2010-03-31       Impact factor: 4.799

4.  Comparison of macular thickness measurements between time domain and spectral domain optical coherence tomography.

Authors:  Christopher Kai-shun Leung; Carol Yim-lui Cheung; Robert N Weinreb; Gary Lee; Dusheng Lin; Chi Pui Pang; Dennis S C Lam
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-04-30       Impact factor: 4.799

5.  Optical coherence tomography.

Authors:  D Huang; E A Swanson; C P Lin; J S Schuman; W G Stinson; W Chang; M R Hee; T Flotte; K Gregory; C A Puliafito
Journal:  Science       Date:  1991-11-22       Impact factor: 47.728

6.  Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments.

Authors:  Ute E K Wolf-Schnurrbusch; Lala Ceklic; Christian K Brinkmann; Milko E Iliev; Manuel Frey; Simon P Rothenbuehler; Volker Enzmann; Sebastian Wolf
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-02-21       Impact factor: 4.799

7.  Comparison of retinal thickness in normal eyes using Stratus and Spectralis optical coherence tomography.

Authors:  Sandeep Grover; Ravi K Murthy; Vikram S Brar; Kakarla V Chalam
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-12-10       Impact factor: 4.799

8.  Risk factors for idiopathic macular holes. The Eye Disease Case-Control Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1994-12-15       Impact factor: 5.258

9.  Macular thickness measurements in normal eyes using spectral domain optical coherence tomography.

Authors:  John E Legarreta; Giovanni Gregori; Omar S Punjabi; Robert W Knighton; Geeta A Lalwani; Carmen A Puliafito
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2008 Jul-Aug

10.  Macular thickness by age and gender in healthy eyes using spectral domain optical coherence tomography.

Authors:  Mehreen Adhi; Sumbul Aziz; Kashif Muhammad; Mohammad I Adhi
Journal:  PLoS One       Date:  2012-05-21       Impact factor: 3.240

View more
  6 in total

1.  Masticatory Muscle Thickness and Activity Correlates to Eyeball Length, Intraocular Pressure, Retinal and Choroidal Thickness in Healthy Women versus Women with Myopia.

Authors:  Grzegorz Zieliński; Marcin Wójcicki; Maria Rapa; Anna Matysik-Woźniak; Michał Baszczowski; Michał Ginszt; Monika Litko-Rola; Jacek Szkutnik; Ingrid Różyło-Kalinowska; Robert Rejdak; Piotr Gawda
Journal:  J Pers Med       Date:  2022-04-13

2.  Comments on Retinal Sensitivity Over Hard Exudates in Diabetic Retinopathy.

Authors:  Brijesh Takkar; Shorya Azad
Journal:  J Ophthalmic Vis Res       Date:  2016 Oct-Dec

3.  Macular thickness in healthy Saudi adults. A spectral-domain optical coherence tomography study.

Authors:  Waseem M Al-Zamil; Fahad M Al-Zwaidi; Sanaa A Yassin
Journal:  Saudi Med J       Date:  2017-01       Impact factor: 1.484

4.  Thickness of individual layers at the macula and associated factors: the Beijing Eye Study 2011.

Authors:  Qian Wang; Wen Bin Wei; Ya Xing Wang; Yan Ni Yan; Jing Yan Yang; Wen Jia Zhou; Szy Yann Chan; Liang Xu; Jost B Jonas
Journal:  BMC Ophthalmol       Date:  2020-02-12       Impact factor: 2.209

5.  The measurement repeatability using different partition methods of intraretinal tomographic thickness maps in healthy human subjects.

Authors:  Jia Tan; Ye Yang; Hong Jiang; Che Liu; Zhihong Deng; Byron L Lam; Liang Hu; Jonathan Oakley; Jianhua Wang
Journal:  Clin Ophthalmol       Date:  2016-11-29

6.  Macular thickness measurements of healthy, naïve cynomolgus monkeys assessed with spectral-domain optical coherence tomography (SD-OCT).

Authors:  Nora Denk; Peter Maloca; Guido Steiner; Christian Freichel; Simon Bassett; Tobias K Schnitzer; Pascal W Hasler
Journal:  PLoS One       Date:  2019-10-07       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.