Lateral mass screw stimulation thresholds in posterior cervical instrumentation surgery: a predictor of medial deviation.

Bayard Wilson1, Erik Curtis2, Brian Hirshman2, Ahmet Oygar2, Karen Chen3, Brandon C Gabel2, Florin Vaida4, David W Allison5, Joseph D Ciacci2.   

Abstract

OBJECTIVE Normative data exists for stimulus-evoked pedicle screw electromyography (EMG) current thresholds in the lumbar spine, and is routinely referenced during spine surgeries to detect a screw breach, prevent injury of neural elements, and ensure the most biomechanically sound instrumentation construct. To date, similar normative data for cervical lateral mass screws is limited, thus the utility of lateral mass screw testing remains unclear. To address this disparity, in this study the authors describe cumulative lateral mass screw stimulation threshold data in patients undergoing posterior cervical instrumentation with lateral mass screws. These data are correlated with screw placement on postoperative imaging, and a novel correlation is discovered with direct clinical implications. METHODS Using a ball-tip probe, 154 lateral mass screws in 21 patients were electrically tested intraoperatively. In each case, for each screw, the lowest (or threshold) current at which the first polyphasic stimulus-evoked EMG response was reproducibly observed by a neurophysiologist was recorded. All patients underwent postoperative CT. Screw position within the lateral mass was first measured in the axial and sagittal planes for each lateral mass screw using the CT images. Screw placement was also evaluated by 2 independent physicians, blinded to current threshold data, on a binary scale of acceptability. The predictive capacity of screw EMG threshold data was evaluated via multivariable regression analyses and receiver operating characteristic (ROC) analyses. Predictive capacity was examined with respect to screw position within the lateral mass, as well as screw acceptability. RESULTS Lateral mass screw EMG thresholds did not appear to differ significantly for screws considered "acceptable" versus "unacceptable" according to the radiographic criteria. Accordingly, ROC analysis confirmed that EMG current threshold data were of minimal utility in predicting screw radiographic acceptability. However, EMG threshold was significantly predictive of screw medial distance from the spinal canal. A screw stimulating below 7.5 mA correctly identified a screw as being within 2 mm of the spinal canal with 75% sensitivity and 92% specificity (positive predictive value 20%, negative predictive value 99.3%), independent of its distance relative to other lateral mass landmarks. EMG current threshold was not significantly predictive of screw deviation in the superior or inferior directions, and was inversely predictive of screw deviations in the lateral direction. CONCLUSIONS In the context of uncertainty regarding the utility of cervical lateral mass EMG current threshold data, this study found that EMG current thresholds correspond significantly, and exclusively, with screw distance from the spinal canal. This association appears independent of other criteria for screw misplacement. As such, the authors recommend that EMG current thresholds be referenced in the case of a suspected medial breach as an effective means to rule out screw placement too medial to the spinal canal.

Entities:  

Keywords:  AUC = area under the curve; CI = confidence interval; EEG = electroencephalography; EMG = electromyography; ROC = receiver operating characteristic; TIVA = total intravenous anesthesia; TcMEP = transcranial motor-evoked potential; cervical; electromyography; intraoperative; lateral mass; medial breach

Mesh:

Year:  2016        PMID: 27935447      PMCID: PMC5485827          DOI: 10.3171/2016.8.SPINE16580

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  19 in total

1.  The anatomic relation of lateral mass screws to the spinal nerves. A comparison of the Magerl, Anderson, and An techniques.

Authors:  R Xu; S P Haman; N A Ebraheim; R A Yeasting
Journal:  Spine (Phila Pa 1976)       Date:  1999-10-01       Impact factor: 3.468

Review 2.  Principles and practical application of the receiver-operating characteristic analysis for diagnostic tests.

Authors:  M Greiner; D Pfeiffer; R D Smith
Journal:  Prev Vet Med       Date:  2000-05-30       Impact factor: 2.670

3.  Electrical stimulation for intraoperative evaluation of thoracic pedicle screw placement.

Authors:  Yong-bing Shi; Michael Binette; William H Martin; James M Pearson; Robert A Hart
Journal:  Spine (Phila Pa 1976)       Date:  2003-03-15       Impact factor: 3.468

4.  Correlation between low triggered electromyographic thresholds and lumbar pedicle screw malposition: analysis of 4857 screws.

Authors:  Barry L Raynor; Lawrence G Lenke; Keith H Bridwell; Brett A Taylor; Anne M Padberg
Journal:  Spine (Phila Pa 1976)       Date:  2007-11-15       Impact factor: 3.468

5.  Anatomic considerations for plate-screw fixation of the cervical spine.

Authors:  H S An; R Gordin; K Renner
Journal:  Spine (Phila Pa 1976)       Date:  1991-10       Impact factor: 3.468

6.  Ability of electromyographic monitoring to determine the presence of malpositioned pedicle screws in the lumbosacral spine: analysis of 2450 consecutively placed screws.

Authors:  Scott L Parker; Anubhav G Amin; S Harrison Farber; Matthew J McGirt; Daniel M Sciubba; Jean-Paul Wolinsky; Ali Bydon; Ziya L Gokaslan; Timothy F Witham
Journal:  J Neurosurg Spine       Date:  2011-08

7.  Cervical pedicle screws versus lateral mass screws. Anatomic feasibility and biomechanical comparison.

Authors:  E L Jones; J G Heller; D H Silcox; W C Hutton
Journal:  Spine (Phila Pa 1976)       Date:  1997-05-01       Impact factor: 3.468

8.  Utility of evoked EMG monitoring to improve bone screw placements in the cervical spine.

Authors:  Robert N Holdefer; Daniel S Heffez; Bernard A Cohen
Journal:  J Spinal Disord Tech       Date:  2013-07

9.  Anesthesia considerations for monitoring TCMEPs in adults diagnosed with poliomyelitis as children: a case report.

Authors:  David W Allison; Jeffrey H Gertsch; Mark A Mahan; Geoffrey L Sheean; Justin M Brown
Journal:  Neurodiagn J       Date:  2014-03

10.  Internal fixation of the lumbar spine with pedicle screw plating.

Authors:  R Roy-Camille; G Saillant; C Mazel
Journal:  Clin Orthop Relat Res       Date:  1986-02       Impact factor: 4.176

View more
  1 in total

Review 1.  Complications associated with subaxial placement of pedicle screws versus lateral mass screws in the cervical spine: systematic review and meta-analysis comprising 1768 patients and 8636 screws.

Authors:  Mohamed A R Soliman; Slah Khan; Nicco Ruggiero; Brandon L Mariotti; Alexander O Aguirre; Cathleen C Kuo; Alexander G Fritz; Siddharth Sharma; Anxhela Nezha; Bennett R Levy; Asham Khan; Amany A Salem; Patrick K Jowdy; Qazi Zeeshan; Moleca M Ghannam; Robert V Starling; John Pollina; Jeffrey P Mullin
Journal:  Neurosurg Rev       Date:  2022-02-09       Impact factor: 3.042

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.