| Literature DB >> 27929489 |
Christopher L Guy1, Kishor Karki, Manju Sharma, Siyong Kim.
Abstract
As flattening filter-free (FFF) photon beams become readily available for treat-ment delivery in techniques such as SBRT, thorough investigation of skin dose from FFF photon beams is necessary under clinically relevant conditions. Using a parallel-plate PTW Markus chamber placed in a custom water-equivalent phantom, surface-dose measurements were taken at 2 × 2, 3 × 3, 4 × 4, 6 × 6, 8 × 8, 10 × 10, 20 × 20, and 30 × 30 cm2 field sizes, at 80, 90, and 100 cm source-to-surface distances (SSDs), and with fields defined by jaws and multileaf collimator (MLC) using multiple beam energies (6X, 6XFFF, 10X, and 10XFFF). The same set of measurements was repeated with the chamber at a reference depth of 10cm. Each surface measurement was normalized by its corresponding reference depth measurement for analysis. The FFF surface doses at 100 cm SSD were higher than flattened surface doses by 45% at 2 × 2 cm2 to 13% at 20 × 20 cm2 for 6MV energy. These surface dose differences varied to a greater degree as energy increased, ranging from +63% at 2 × 2 cm2 to -2% at 20 × 20 cm2 for 10 MV. At small field sizes, higher energy increased FFF surface dose relative to flattened surface dose; while at larger field sizes, relative FFF surface dose was higher for lower energies. At both energies investigated, decreasing SSD caused a decrease in the ratios of FFF-to-flattened surface dose. Variability with SSD of FFF-to-flattened surface dose differences increased with field size and ranged from 0% to 6%. The field size at which FFF and flattened beams gave the same skin dose increased with decreasing beam energy. Surface dose was higher with MLC fields compared to jaw fields under most conditions, with the difference reaching its maximum at a field size between 4 × 4 cm2 and 6 × 6 cm2 for a given energy and SSD. This study conveyed the magnitude of surface dose in a clinically meaning-ful manner by reporting results normalized to 10 cm depth dose instead of depth of dose maximum.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27929489 PMCID: PMC5690509 DOI: 10.1120/jacmp.v17i6.6307
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experimental measurement setup: (a) experimental setup is shown for surface measurements using the water‐equivalent phantom; (b) close‐up view of the Markus chamber set in the custom‐machined phantom for surface measurements.
Figure 2Diagrams of measurement geometries are shown for (a) surface measurements and (b) reference (10 cm) depth measurements. The appropriate SSD was maintained when switching between the two geometries.
Figure 3Ratios of FFF‐to‐flattened surface dose are shown as a function of field size in water‐equivalent phantom for jaw fields (left) and MLC fields (right). For MLC fields, jaws were set to . Each surface measurement was first normalized by the reference measurement of the same field size. The same legend applies to both subfigures.
Figure 4Ratios of MLC‐to‐jaw field surface dose in water‐equivalent phantom at each energy as a function of field size at 100 cm SSD. For MLC fields, jaws were set to . Each surface measurement was first normalized by the reference measurement of the same field size.
Figure 5FFF‐to‐flattened depth‐dose differences. Percentage differences between FFF and flattened doses at a reference depth of 10 cm.