Monte Carlo comparison of superficial dose between flattening filter free and flattened beams.

Khosrow Javedan1, Vladimir Feygelman1, Ray R Zhang2, Eduardo G Moros1, Candace R Correa1, Andy Trotti1, Weiqi Li1, Geoffrey G Zhang3.   

Abstract

This study investigates the superficial dose from FFF beams in comparison with the conventional flattened ones using a Monte Carlo (MC) method. Published phase-space files which incorporated real geometry of a TrueBeam accelerator were used for the dose calculation in phantom and clinical cases. The photon fluence on the central axis is 3 times that of a flattened beam for a 6 MV FFF beam and 5 times for a 10 MV beam. The mean energy across the field in air at the phantom surface is 0.92-0.95 MeV for the 6 MV FFF beam and 1.18-1.30 MeV for the corresponding flattened beam. At 10 MV, the values are 1.52-1.72 and 2.15-2.87 MeV for the FFF and flattened beams, respectively. The phantom dose at the depth of 1 mm in the 6 MV FFF beam is 6% ± 2.5% (of the maximum dose) higher compared to the flattened beam for a 25 × 25 cm(2) field and 14.6% ± 1.9% for the 2 × 2 cm(2) field. For the 10 MV beam, the corresponding differences are 3.4% ± 1.5% and 10.7% ± 0.6%. The skin dose difference at selected points on the patient's surface between the plans using FFF and flattened beams in the head-and-neck case was 6.5% ± 2.3% (1SD), and for the breast case it was 6.4% ± 2.3%. The Monte Carlo simulations showed that due to the lower mean energy in the FFF beam, the clinical superficial dose is higher without the flattening filter compared to the flattened beam.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Flattening filter free; Monte Carlo; Skin dose

Mesh:

Year:  2014        PMID: 24662096     DOI: 10.1016/j.ejmp.2014.03.001

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  9 in total

1.  Skin Surface Dose for Whole Breast Radiotherapy Using Personalized Breast Holder: Comparison with Various Radiotherapy Techniques and Clinical Experiences.

Authors:  Chiu-Ping Chen; Chi-Yeh Lin; Chia-Chun Kuo; Tung-Ho Chen; Shao-Chen Lin; Kuo-Hsiung Tseng; Hao-Wen Cheng; Hsing-Lung Chao; Sang-Hue Yen; Ruo-Yu Lin; Chen-Ju Feng; Long-Sheng Lu; Jeng-Fong Chiou; Shih-Ming Hsu
Journal:  Cancers (Basel)       Date:  2022-06-30       Impact factor: 6.575

2.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

3.  Monte Carlo study of radiation dose enhancement by gadolinium in megavoltage and high dose rate radiotherapy.

Authors:  Daniel G Zhang; Vladimir Feygelman; Eduardo G Moros; Kujtim Latifi; Geoffrey G Zhang
Journal:  PLoS One       Date:  2014-10-02       Impact factor: 3.240

4.  Superficial and peripheral dose in compensator-based FFF beam IMRT.

Authors:  Daniel G Zhang; Vladimir Feygelman; Eduardo G Moros; Kujtim Latifi; Sarah Hoffe; Jessica Frakes; Geoffrey G Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-12-21       Impact factor: 2.102

5.  Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

Authors:  Madelaine K Tyler; Paul Z Y Liu; Christopher Lee; David R McKenzie; Natalka Suchowerska
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

6.  Clinically relevant investigation of flattening filter-free skin dose.

Authors:  Christopher L Guy; Kishor Karki; Manju Sharma; Siyong Kim
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

7.  Enhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode.

Authors:  Zeghari A; Saaidi R; Cherkaoui El Moursli R
Journal:  J Biomed Phys Eng       Date:  2019-08-01

8.  Surface dose measurements and comparison of unflattened and flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj; Girigesh Yadav; Ananda Giribabu; Karthikeyan Srinivasan; Venkadamanickam Gurusamy; Kothanda Raman; Kaviarasu Karunakaran; Rajesh Thiyagarajan
Journal:  J Med Phys       Date:  2016 Apr-Jun

9.  Skin recurrence in the radiation treatment of breast cancer.

Authors:  Leah M Katz; Carmen A Perez; Naamit K Gerber; Juhi Purswani; Allison McCarthy; Indra J Das
Journal:  Adv Radiat Oncol       Date:  2018-05-07
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.