BACKGROUND: The Labcor Dokimos Plus (LDP) is a stented externally mounted pericardial aortic bioprosthesis, which was recently introduced in Europe. Aims of the study are evaluation of operative and postoperative results as well as hemodynamic performance.Entities:
Keywords: Biological prosthesis; Echocardiography; Stented aortic valve replacement; Valve replacement
Mesh:
Year: 2016 PMID: 27899119 PMCID: PMC5129248 DOI: 10.1186/s13019-016-0561-5
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Baseline characteristics and risk stratification
| Characteristic | Number |
|---|---|
| Number of patients ( | 100 |
| • Age in years ± standard deviation | 65.9 ± 10.7 |
| Gender | |
| • Male ( | 77 |
| • Female ( | 23 |
| Body surface area ± standard deviation (m2) | 2.0 ± 0.2 |
| Aortic valve lesion | |
| • Stenosis ( | 55 |
| • Regurgitation ( | 25 |
| • Mixed lesion ( | 20 |
| • Active endocarditis ( | 9 |
| Predominant cardiac rhythm | |
| • Sinus rhythm ( | 73 |
| • Atrial fibrillation ( | 27 |
| Concomitant disease | |
| • Coronary artery disease ( | 38 |
| • Arterial hypertension ( | 67 |
| • Pulmonary hypertension ( | 7 |
| • Renal dysfunction ( | 32 |
| • Diabetes mellitus ( | 22 |
| • Obesity ( | 30 |
| NYHA class (Mean ± standard deviation) | 2.6 ± 0.7 |
| EuroSCORE II | 3.1 ± 3.9 |
| • Isolated aortic valve replacement | 2.0 ± 1.7 |
| • Complex procedures | 5.9 ± 3.8 |
Fig. 1Lateral view and front view of the Labcor Dokimos Plus
Operative characteristics
| Procedure | Number |
|---|---|
| Isolated aortic valve replacement ( | 66 |
| Combined procedures ( | 34 |
| • Coronary artery bypass grafting ( | 21 |
| • Mitral Valve Replacement ( | 5 |
| • Mitral Valve Repair ( | 3 |
| • Left atrial ablation ( | 3 |
| • Ascending Aorta Replacement ( | 2 |
| • Tricuspid Valve Reconstruction ( | 2 |
| • Aortic Annular Enlargement ( | 2 |
| Implanted valve sizes | |
| • 21 mm ( | 3 |
| • 23 mm ( | 20 |
| • 25 mm ( | 41 |
| • 27 mm ( | 36 |
| Technique of implantation | |
| • Supra-annularly ( | 40 |
| • Intra-annularly ( | 60 |
| Duration of procedure (min) | 212.4 ± 57.7 |
| • Isolated procedures (min) | 189.2 ± 36.3 |
| • Combined procedures (min) | 257.2 ± 65.4 |
| Cardiopulmonary bypass time (min) | 113.6 ± 40.6 |
| • Isolated procedures (min) | 96.6 ± 25.3 |
| • Combined procedures (min) | 140.2 ± 45.7 |
| Aortic cross clamp time (min) | 84.8 ± 30.0 |
| • Isolated procedures (min) | 74.5 ± 20.0 |
| • Combined procedures (min) | 103.7 ± 37.1 |
Echocardiographic results according to labeled valve sizes
| Valve size (mm) | 21 | 23 | 25 | 27 |
|---|---|---|---|---|
| Number | 3 | 18 | 39 | 33 |
| Mean Pressure | 17.3 ± 6.7 | 9.6 ± 3.0 | 8.5 ± 3.1 | 10.2 ± 3.6 |
| Maximum Pressure gradient in mmHg | 29.7 ± 12.1 | 19.3 ± 4.9 | 16.5 ± 5.8 | 19.1 ± 5.8 |
| Effective Orifice Area | 1.92 ± 0.44 | 1.79 ± 0.36 | 2.0 ± 0.6 | 2.16 ± 0.47 |
| Indexed Effective Orifice Area in cm2/m2 | 1.08 ± 0.33 | 0.95 ± 0.18 | 0.99 ± 0.29 | 1.01 ± 0.24 |