| Literature DB >> 27853474 |
Vicente Carmona1, Jose Perez-Calatayud2, Françoise Lliso1, Jose Richart3, Facundo Ballester4, Mª Carmen Pujades-Claumarchirant1, Jack L M Venselaar5.
Abstract
PURPOSE: In this work a spreadsheet based program is presented that to a large extent independently verifies the calculations of individual plans of brachytherapy treatment planning systems for low dose rate, high dose rate and pulsed dose rate techniques.Entities:
Keywords: brachytherapy; physics dosimetry; planning system; verification
Year: 2010 PMID: 27853474 PMCID: PMC5104832 DOI: 10.5114/jcb.2010.16924
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Treatment planning systems and sources/seeds used at our departments for which the verification spreadsheets are used
| Treatment planning system | Application | Source/Seed |
|---|---|---|
| PLATO | HDR and PDR intracavitary and interstitial implants | mHDR-v2 |
| Oncentra MasterPlan Brachy | HDR and PDR intracavitary and interstitial implants | mHDR-v2 |
| Oncentra Prostate | HDR, Prostate intraoperative implants | mHDR-v2 |
| SPOT-PRO | LDR seeds, Prostate intraoperative implants | IsoCord |
| TPS Plaque Simulator X | LDR seeds, Ophtalmic implants | IsoSeed I-125 |
Fig. 1Diagram showing the verification process for each point, in HDR and PDR calculations. Ḋ is the dose rate per unit air-kerma strength, G is the geometry function (linesource model) and D is the absorbed dose at point i
Fig. 2Screen shots of the result of a verification. On the left, the data for the implant are shown along with two of the three main views. On the right, the third main view as well as the comparison for the first eleven points
Mean deviation, standard deviation (1sd), and maximum difference observed between the TPS calculations and the verification program
| Application | N° cases | Difference (%) | ||
|---|---|---|---|---|
| Mean | Standard deviation | Maximum difference | ||
| HDR-PDR | 95 | –0.22 | 0.48 | –1.93 |
| LDR | 155 | 0.17 | 0.07 | 0.42 |