An equation to QA check the total treatment time for single-catheter HDR brachytherapy.

R D Rogus1, M J Smith, H D Kubo.   

Abstract

PURPOSE: To develop a simple equation that can be used to quickly QA check the total treatment time needed for straight or moderately bending, single-catheter implants from the treatment length and dose-prescription depth. A method to quickly check the total treatment time is useful for high dose rate (HDR) treatments, but may be most useful for intravascular patients that are catheterized while planning the treatment. METHODS AND MATERIALS: Nucletron-Oldelft treatment planning software was used to plan treatments with treatment lengths from 50 to 200 mm, and dose-prescription depths from 2 to 14 mm. For each plan, dose points were defined at a fixed depth away from the catheter. Doses were calculated, normalized, and optimized using the dose points. From these plans, a single equation was developed for the total treatment time as a function of treatment length and dose-prescription depth.
RESULTS: Total treatment time increases linearly with treatment length and supra-linearly with dose-prescription depth. A difference of 3% or more between the treatment time determined by the QA check equation and by the treatment planning software will alert the clinician and physicist to check for potential errors in dose prescription, prescription depth, source activity, treatment length, or inadvertent modifications to physics parameters used in the treatment planning software.
CONCLUSION: The simple equation that is developed can be used to quickly and accurately QA check the total treatment time needed for straight or moderately bending, single-catheter implants.

Entities:  

Mesh:

Year:  1998        PMID: 9422582     DOI: 10.1016/s0360-3016(97)00487-2

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  High dose rate (HDR) brachytherapy quality assurance: a practical guide.

Authors:  D A Wilkinson
Journal:  Biomed Imaging Interv J       Date:  2006-04-01

2.  A program for the independent verification of brachytherapy planning system calculations.

Authors:  Vicente Carmona; Jose Perez-Calatayud; Françoise Lliso; Jose Richart; Facundo Ballester; Mª Carmen Pujades-Claumarchirant; Jack L M Venselaar
Journal:  J Contemp Brachytherapy       Date:  2010-10-13

3.  A method for checking high dose rate treatment times for vaginal applicators.

Authors:  C S Mayo; K Ulin
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

4.  The usefulness of an independent patient-specific treatment planning verification method using a benchmark plan in high-dose-rate intracavitary brachytherapy for carcinoma of the uterine cervix.

Authors:  Yutaka Takahashi; Masahiko Koizumi; Iori Sumida; Fumiaki Isohashi; Toshiyuki Ogata; Yuichi Akino; Yasuo Yoshioka; Shintaro Maruoka; Shinichi Inoue; Koji Konishi; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2012-07-17       Impact factor: 2.724

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.