A multicentre survey of the condition of ultrasound probes.

Nicholas J Dudley1, Darren J Woolley2.   

Abstract

BACKGROUND: The implementation of quality assurance for ultrasound scanners in the United Kingdom is patchy, but government appointed bodies require quality assurance and there are regulatory requirements for maintenance and inspection of equipment. Previous studies have shown high fault rates in ultrasound probes; some of these studies used electronic probe testers, but there is good evidence that over 90% of faults may be detected using simple methods. We aimed to conduct a multicentre survey of the condition of probes, using visual inspection and assessing the in-air reverberation.
METHODS: Visitors to the stand run by Multi-Medix Ltd at the BMUS Annual Scientific Meeting in 2014 were invited to participate in the study. One or both of the authors visited participants, performing a visual inspection of probes for evidence of damage or wear and inspecting the in-air reverberation pattern for uniformity. Probes were classified using a risk-based traffic light system: green-no fault found; amber-fault found; further action required but the probe need not be removed from use; red-the probe should be removed from use due to physical or clinical diagnostic risks.
RESULTS: Twelve sites and 219 probes were included in the survey. Sixty-three percent of probes were classified as green; 25% as amber and 13% as red.
CONCLUSION: More than one in three probes were faulty. Simple tests, at minimal cost, have the potential to demonstrate over 90% of probe faults, making it possible for employers to comply with their duties defined by regulations, national standards and professional guidelines.

Entities:  

Keywords:  Ultrasound; quality assurance; transducer

Year:  2016        PMID: 27847533      PMCID: PMC5098704          DOI: 10.1177/1742271X16662301

Source DB:  PubMed          Journal:  Ultrasound        ISSN: 1742-271X


  8 in total

1.  Ultrasound transducer function: annual testing is not sufficient.

Authors:  Mattias Mårtensson; Mats Olsson; Lars-Åke Brodin
Journal:  Eur J Echocardiogr       Date:  2010-06-08

2.  Quality assurance in diagnostic ultrasound.

Authors:  Outi Sipilä; Vilma Mannila; Eija Vartiainen
Journal:  Eur J Radiol       Date:  2010-12-08       Impact factor: 3.528

3.  Linear array test tool.

Authors:  A Goldstein; D Ranney; R D McLeary
Journal:  J Ultrasound Med       Date:  1989-07       Impact factor: 2.153

4.  High incidence of defective ultrasound transducers in use in routine clinical practice.

Authors:  Mattias Mårtensson; Mats Olsson; Björn Segall; Alan G Fraser; Reidar Winter; Lars-Ake Brodin
Journal:  Eur J Echocardiogr       Date:  2008-10-22

5.  The effect of dead elements on the accuracy of Doppler ultrasound measurements.

Authors:  Jaromir Vachutka; Ladislav Dolezal; Christian Kollmann; Jakob Klein
Journal:  Ultrason Imaging       Date:  2014-01       Impact factor: 1.578

6.  Four-year experience with a clinical ultrasound quality control program.

Authors:  Nicholas J Hangiandreou; Scott F Stekel; Donald J Tradup; Krzysztof R Gorny; Deirdre M King
Journal:  Ultrasound Med Biol       Date:  2011-06-16       Impact factor: 2.998

7.  BMUS guidelines for the regular quality assurance testing of ultrasound scanners by sonographers.

Authors:  Nick Dudley; Stephen Russell; Barry Ward; Peter Hoskins
Journal:  Ultrasound       Date:  2013-11-29

8.  The BMUS guidelines for regular quality assurance testing of ultrasound scanners.

Authors:  Nick Dudley; Stephen Russell; Barry Ward; Peter Hoskins
Journal:  Ultrasound       Date:  2014-02-06
  8 in total
  4 in total

1.  A precise, reproducible method for measuring ultrasound probe slice thickness using a Gammex 403 phantom.

Authors:  Steven J Jackson; Stephen Russell
Journal:  Ultrasound       Date:  2019-02-14

2.  Modeling and Experimental Characterization of Bonding Delaminations in Single-Element Ultrasonic Transducer.

Authors:  Wenxiang Ding; Maxime Bavencoffe; Marc Lethiecq
Journal:  Materials (Basel)       Date:  2021-04-27       Impact factor: 3.623

3.  Method for automatic detection of defective ultrasound linear array transducers based on uniformity assessment of clinical images - A case study.

Authors:  Robert Lorentsson; Nasser Hosseini; Jan-Olof Johansson; Wiebke Rosenberg; Benny Stenborg; Lars Gunnar Månsson; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2018-01-11       Impact factor: 2.102

4.  Failure mode effect and criticality analysis of ultrasound device by classification tracking.

Authors:  Longchen Wang; Bin Li; Bing Hu; Guofeng Shen; Yunxin Zheng; Yuanyi Zheng
Journal:  BMC Health Serv Res       Date:  2022-04-01       Impact factor: 2.655

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.